In addition to other requirements of this chapter, diagnostic X-ray systems shall meet the following requirements:
WAC 246-225-040 TABLE 1
Design operating range (kilovolts peak) | Measured potential (kilovolts peak) | Half-value layer (milli- meters of aluminum equivalent) | Half-value layer (milli- meter of aluminum equivalent for dental units) |
Below 51-- | 30 | 0.3 | N/A |
40 | 0.4 | N/A | |
50 | 0.5 | 1.5 | |
51 to 70--- | 51 | 1.2 | 1.5 |
60 | 1.3 | 1.5 | |
70 | 1.5 | 1.5 | |
Above 70-- | 71 | 2.1 | 2.1 |
80 | 2.3 | 2.3 | |
90 | 2.5 | 2.5 | |
100 | 2.7 | 2.7 | |
110 | 3.0 | 3.0 | |
120 | 3.2 | 3.2 | |
130 | 3.5 | 3.5 | |
140 | 3.8 | 3.8 | |
150 | 4.1 | 4.1 |
X1 - X2 <= 0.10 (X1 + X2) |
Where X1 and X2 are the ratios (mR/mAs) for each mA or mAs station.
The test shall be performed at any selections of mA or mAs without regard to focal spot size, provided neither focal spot size is less than 0.45 millimeter.
Wash. Admin. Code § 246-225-040
Statutory Authority: RCW 70.98.050. 94-01-073, § 246-225-040, filed 12/9/93, effective 1/9/94. Statutory Authority: RCW 70.98.050 and 70.98.080. 91-15-083 (Order 183), § 246-225-040, filed 7/23/91, effective 8/23/91. Statutory Authority: RCW 43.70.040. 91-02-049 (Order 121), recodified as § 246-225-040, filed 12/27/90, effective 1/31/91. Statutory Authority: RCW 70.98.080. 87-01-031 (Order 2450), § 402-28-035, filed 12/11/86; 83-19-050 (Order 2026), § 402-28-035, filed 9/16/83. Statutory Authority: RCW 70.98.050. 81-01-011 (Order 1570), § 402-28-035, filed 12/8/80; Order 1084, § 402-28-035, filed 1/14/76. Formerly WAC 402-28-030 (part).