Section 12VAC5-481-1653 - Hand-held radiographic unit In addition to the applicable provisions found elsewhere in this chapter, the following provisions apply to a hand-held radiographic unit.
1. A hand-held radiograph unit shall be:a. Certified by the manufacturer pursuant to 21 CFR Part 803, Medical Device Reporting of the Federal Food and Drug Administration Modernization Act of 1997; 21 USC Chapter 9, Subchapter V, Part C - Electronic Product Radiation Control (EPRC) (§ 360hh et seq.) of the Federal Food, Drug and Cosmetic Act; and 21 CFR 1020.30, Diagnostic x-ray systems and their major components.b. Registered with the agency in accordance with applicable parts of this chapter.c. Maintained and operated in accordance with the manufacturer's specifications.2. For all uses: a. Operators of a hand-held radiographic unit shall be specifically trained to operate such equipment.b. When operating a hand-held radiographic unit, operators shall wear dosimetry unless otherwise authorized by the agency.c. A hand-held radiographic unit shall have the backscatter radiation shield in place to protect the operator during operation.d. The operator shall ensure there are no bystanders within a radius of at least six feet from the patient being examined with a hand-held radiograph unit.e. A hand-held radiographic unit shall not be used in hallways or waiting rooms.12 Va. Admin. Code § 5-481-1653