S.C. Code Regs. § § 61-64.IV.RHB 4.8

Current through Register Vol. 48, No. 10, October 25, 2024
Section 61-64.IV.RHB 4.8 - Mobile and Portable Radiographic Systems

In addition to the applicable provisions of this regulation, the requirements of this Part apply to all mobile and portable radiographic systems.

4.8.1 All provisions of RHB 4.7.4 through 4.7.14 apply, except 4.7.12 and 4.7.4.2.2.
4.8.2 Means shall be provided for independent stepless adjustment of at least two (2) dimensions of the x-ray field.
4.8.3 Means shall be provided to indicate when the axis of the x-ray beam is perpendicular to the plane of the image receptor.
4.8.4 Means shall be provided for visually defining the perimeter of the x-ray field. The total misalignment of the edges of the visually defined field with the respective edges of the x-ray field along either the length or width of the visually defined field shall not exceed two percent (2%) of the distance from the source to the center of the visually defined field when the surface upon which it appears is perpendicular to the axis of the x-ray beam.
4.8.5 If provided, the beam-limiting device shall indicate numerically the field size in the plane of the image receptor to which it is adjusted.
4.8.6 If collimator indications are provided, the indications shall be such that aperture adjustments result in x-ray field dimensions in the plane of the image receptor which correspond to those of the image receptor to within two percent (2%) of the SID when the beam axis is perpendicular to the plane of the image receptor.
4.8.7 Means shall be provided to measure SIDs, and shall be accurate to within two percent (2%).
4.8.8 Mobile and portable x-ray systems which are used in a single location for a period of five (5) or more consecutive days shall be considered a stationary radiographic system and shall meet the requirements of RHB 4.4.
4.8.9 Mobile and portable x-ray systems which are used at multiple locations shall be provided with an adequate protective barrier or protective apron for the operator and with a method of control which will permit the operator to be at least six feet (6 ft) from the tube head and the nearest edge of the useful beam during exposures.
4.8.10 Tube stands. A tube stand or other mechanical support shall be used for portable x-ray systems so that the x-ray tube housing assembly need not be hand-held during exposures.
4.8.11 All mobile or portable radiographic systems shall be provided with means to limit the source-to-skin distance to equal to or greater than thirty centimeters (30 cm).

S.C. Code Regs. § 61-64.IV.RHB 4.8

Replaced and amended by State Register Volume 40, Issue No. 06, eff. 6/24/2016; State Register Volume 47, Issue No. 05, eff. 5/26/2023.