Section 333-106-0230 - Fluoroscopic Timer(1) Means shall be provided to present the cumulative on-time of the fluoroscopic tube.(2) The maximum cumulative time of the timing device shall not exceed five minutes without resetting.(3) A signal audible to the fluoroscopist, or the appropriate operator, shall indicate the completion of any preset cumulative on-time; or if no audible signal is provided, the exposure shall terminate.Or. Admin. Code § 333-106-0230
HD 4-1985, f. & ef. 3-20-85; PH 14-2008, f. & cert. ef. 9-15-08Stat. Auth.: ORS 453.752 - 453.775