177 Neb. Admin. Code, ch. 1, § 011

Current through September 17, 2024
Section 177-1-011 - REPAIR OF CLASS B BREATH TESTING DEVICES
011.01 When inspection of a testing device reveals the need for repair which may affect the validity of the test and requires the attention of a manufacturer's representative or an individual trained for repair, it shall be repaired by the appropriate repairman and a record of each repair will be retained at the testing site.
011.02 Repair of a testing device, as opposed to maintenance (009.01D), includes the removal of the malfunctioning part(s) and the installation of the repair part(s). The removal or installation of all parts or electronic boards shall be recorded.
011.03 Calibration verification procedures shall be performed on a testing device following its repair, before it is returned to service.
011.04 The records to be maintained for repair activities shall include the type of malfunction of a testing device, the nature of the repair, the date of the repair, and shall show the name of the person performing these activities or the name of the person's agency or business organization.
011.05 The repair records, or copies of the repair records, for a testing device shall be made available to the Department upon request.

177 Neb. Admin. Code, ch. 1, § 011