Rule 5. A procedure that is used in conjunction with evidential breath alcohol analysis must be approved by the department and comply with the following requirements:
Table 1
Allowable variation of second breath result range from the first sample result.
First Sample | Allowable variation |
0.00 - 0.14 | +/- 0.01 |
0.15 - 0.24 | +/- 0.02 |
0.25 - 0.34 | +/- 0.03 |
0.35 - or more | +/- 0.04 |
For instruments reporting 3 digits after the decimal, the second breath sample result must not vary from the first sample result by more than the allowed variation listed in Table 2.
Table 2
Allowable variation of second breath sample result from the first breath sample result.
First Sample | Allowable Variation |
0.000 - 0.149 | +/- 0.010 |
0.150 - 0.249 | +/- 0.020 |
0.250 - 0.349 | +/- 0.030 |
0.350 or more | +/- 0.040 |
Mich. Admin. Code R. 325.2655