"Accessible surface " means surface of equipment or of an equipment part that can be easily or accidentally touched by persons without the use of a tool.
Added filtration" means any filtration which is in addition to the inherent filtration.
"Beam-limiting device" means a field defining collimator, integral to the therapeutic radiation machine, which provides a means to restrict the dimensions of the useful beam.
"Beam-scattering foil" means a thin piece of material (usually metallic) placed in the beam to scatter a beam of electrons in order to provide a more uniform electron distribution in the useful beam.
"Bent beam linear accelerator" means a linear accelerator geometry in which the accelerated electron beam must change direction by passing through a bending magnet.
"Contact therapy system " means a therapeutic radiation machine with a short target-to-skin distance (TSD), usually less than 5 centimeters.
"Dose monitor unit (DMU) " means a unit response from the beam monitoring system from which the absorbed dose can be calculated.
"External beam radiation therapy" means therapeutic irradiation in which the source of radiation is at a distance from the body.
"Field flattening filter " means a filter used to homogenize the absorbed dose rate over the radiation field.
"Filter " means material placed in the useful beam to change beam quality or its intensity profile in therapeutic radiation machines.
"Gantry " means that part of a radiation therapy system supporting and allowing movements of the radiation head around a center of rotation.
"Interruption of irradiation" means the stopping of irradiation with the possibility of continuing irradiation without resetting of operating conditions at the control panel.
"Isocenter" means the center of the sphere through which the useful beam axis passes while the gantry moves through its full range of motions.
"Megavolt (MV) (mega electron volt (MeV)) " means the energy equal to that acquired by a particle with one electron charge in passing through a potential difference of 1 million volts in a vacuum. (Note: Current convention is to use MV for photons and MeV for electrons.)
"Monitor unit (MU). " See "Dose monitor unit."
"Moving beam radiation therapy " means radiation therapy with continuous displacement of one or more mechanical axes relative to the patient during irradiation. It includes arc therapy, skip therapy, conformal therapy, intensity modulation, and rotational therapy.
"Nominal treatment distance" means:
"Periodic quality assurance check" means a procedure which is performed to ensure that a previous calibration continues to be valid.
"Practical range of electrons" corresponds to classical electron range where the only remaining contribution to dose is from bremsstrahlung X-rays. A further explanation may be found in "Clinical Electron Beam Dosimetry: Report of AAPM Radiation Therapy Committee Task Group 25" (Medical Physics 18(1): 73-109, Jan/Feb 1991) and ICRU Report 35, "Radiation Dosimetry: Electron Beams with Energies Between 1 and 50 MeV," International Agency on Radiation Units and Measurements, September 15, 1984.
"Radiation field. " See "Useful beam."
"Radiation head" means the structure from which the useful beam emerges.
"Radiation therapy physicist" means an individual qualified in accordance with 41.3(6).
"Redundant beam monitoring system " means a combination of two dose monitoring systems in which each system is designed to terminate irradiation in accordance with a preselected number of dose monitor units.
"Shadow tray" means a device attached to the radiation head to support auxiliary beam blocking material.
"Stationary beam radiation therapy " means radiation therapy without displacement of one or more mechanical axes relative to the patient during irradiation.
"Target" means that part of an X-ray tube or accelerator onto which is directed a beam of accelerated particles to produce ionizing radiation or other particles.
"Tenth-value layer (TVL) " means the thickness of a specified material which attenuates X-radiation or gamma radiation to an extent such that the air kerma rate, exposure rate or absorbed dose rate is reduced to one-tenth of the value measured without the material at the same point.
"Therapeutic radiation machine " means X-ray or electron-producing equipment designed and used for external beam radiation therapy.
"Virtual source " means a point from which radiation appears to originate.
* Whenever quality assurance check measurements indicate that the radiation output differs by more than 5 percent from the value obtained at the last full calibration and the difference cannot be reconciled; and
* Following any component replacement, major repair, or modification of components that could significantly affect the characteristics of the radiation beam.
* Full calibration of therapeutic radiation machines with multienergy capabilities is required only for those modes or energies that are not within their acceptable range; and
* If the repair, replacement or modification does not affect all energies, full calibration shall be performed on the affected energy that is in most frequent clinical use at the facility. The remaining energies may be validated with quality assurance check procedures against the criteria in 41.3(17)"b "(3).
* A maximum of 2 percent and average of 0.5 percent of the absorbed dose on the central axis of the useful beam at the nominal treatment distance. This limit shall apply beyond a line seven centimeters outside the periphery of the useful beam; and
* A maximum of 10 percent of the absorbed dose on the central axis of the useful beam at the nominal treatment distance. This limit shall apply beyond a line two centimeters outside the periphery of the useful beam.
* Irradiation shall not be possible until a selection of a filter or a positive selection to use "no filter" has been made at the treatment control panel, either manually or automatically;
* An interlock system shall be provided to prevent irradiation if the filter selected is not in the correct position;
* A display shall be provided at the treatment control panel showing the wedge filter(s), interchangeable field-flattening filter(s), and interchangeable beam-scattering foil(s) in use; and
* An interlock shall be provided to prevent irradiation if any filter or be am-scattering foil selection operation carried out in the treatment room does not agree with the filter or beam-scattering foil selection operation carried out at the treatment control panel.
* Each detector shall be removable only with tools and, if movable, shall be interlocked to prevent incorrect positioning;
* Each detector shall form part of a beam monitoring system from whose readings in dose monitor units the absorbed dose at a reference point can be calculated;
* Each beam monitoring system shall be capable of independently monitoring, interrupting, and terminating irradiation; and
* Malfunctioning of one system shall not affect the correct functioning of the other system(s); and
* Failure of any element common to both systems which could affect the correct function of both systems shall terminate irradiation or prevent the initiation of radiation.
* Maintain a reading until intentionally reset;
* Have only one scale and no electrical or mechanical scale multiplying factors;
* Utilize a design such that increasing dose is displayed by increasing numbers; and
* In the event of power failure, the beam monitoring information required in 41.3(18) "a"(6)"4" displayed at the control panel at the time of failure shall be retrievable in at least one system for a 20-minute period of time.
* An interlock system shall be provided to terminate irradiation if the number of dose monitor units delivered in any ten degrees of rotation or one centimeter of linear motion differs by more than 20 percent from the selected value;
* Where angle terminates the irradiation in moving beam radiation therapy, the dose monitor units delivered shall differ by less than 5 percent from the dose monitor unit value selected;
* An interlock shall be provided to prevent motion of more than five degrees or one centimeter beyond the selected limits during moving beam radiation therapy;
* An interlock shall be provided to require that a selection of direction be made at the treatment control panel in all units which are capable of both clockwise and counterclockwise moving beam radiation therapy.
* Moving beam radiation therapy shall be controlled with both primary position sensors and secondary position sensors to obtain the selected relationships between incremental dose monitor units and incremental movement.
* Occurs during stationary beam radiation therapy; or
* Does not start or stops during moving beam radiation therapy unless such stoppage is a preplanned function.
* Whenever quality assurance check measurements indicate that the radiation output differs by more than 5 percent from the value obtained at the last full calibration and the difference cannot be reconciled. Therapeutic radiation machines with multienergy or multimode capabilities or both shall only require measurements for those modes or energies that are not within their acceptable range; and
* Following any component replacement, major repair, or modification of components that could significantly affect the characteristics of the radiation beam. If the repair, replacement or modification does not affect all modes or energies, measurements shall be performed on the affected mode/energy that is in most frequent clinical use at the facility. The remaining energies/modes may be validated with quality assurance check procedures against the criteria in 41.3(18)"e "(1)"3."
Iowa Admin. Code r. 641-41.3