p = (w/M)(RT/v),
where R is the gas constant (8.31 Pa m2 mol-1 K-1) when the pressure is in pascals (Pa) and the volume is in cubic meters. As noted by Spencer and Cliath (1970) under paragraph (d)(3) of this section, direct vapor pressure measurements by gas saturation techniques are more directly related to the volatilization of chemicals than are other techniques.
FIGURE 1-SCHEMATIC DIAGRAM OF VAPOR SATURATION APPARATUS
The insulated box, containing sample holders, may be of any suitable size and shape. The sketch in Figure 1 shows a box containing three solid sample holders and three liquid sample holders, which allows for the triplicate analysis of either a solid or liquid sample. The temperature within the box is controlled to ±0.5° or better. Nitrogen gas, split into six streams and controlled by fine needle valves (approximately 0.79 mm orifice), flows into the box via 3.8 mm (0.125 in.) i.d. copper tubing. After temperature equilibration, the gas flows through the sample and the sorbent trap and exits from the box. The flow rate of the effluent carrier gas is measured at room temperature with a bubble flow meter or other suitable device. The flow rate is checked frequently during the experiment to assure that there is an accurate value for the total volume of carrier gas. The flow rate is used to calculate the total volume (at room temperature) of gas that has passed through the sample and sorbent [(vol/time) * time = volume]. The vapor pressure of the test substance can be calculated from the total gas volume and the mass of sample vaporized. If v is the volume of gas that transported mass w of the vaporized test material having a molecular weight M, and if p is the equilibrium vapor pressure of the sample at temperature T, then p is calculated by the equation
p = (w/M)(RT/v).
In this equation, R is the gas constant (8.31 Pa m3mol-1 K-1). The pressure is expressed in pascals (Pa), the volume in cubic meters (m3), mass in grams and T in kelvins (K). T = 273.15 + t, if t is measured in degrees Celsius (°C).
FIGURE 2-SOLID COMPOUND SAMPLING SYSTEM
FIGURE 3-LIQUID COMPOUND SAMPLING SYSTEM
The most reproducible method for measuring the vapor pressure of liquids is to coat the liquid on glass beads and to pack the holder in the designated place with these beads.
40 C.F.R. §796.1950