Masimo CorporationDownload PDFPatent Trials and Appeals BoardFeb 23, 2022IPR2020-01539 (P.T.A.B. Feb. 23, 2022) Copy Citation Trials@uspto.gov Paper 41 571-272-7822 Entered: February 23, 2022 UNITED STATES PATENT AND TRADEMARK OFFICE ____________ BEFORE THE PATENT TRIAL AND APPEAL BOARD APPLE INC., Petitioner, v. MASIMO CORPORATION, Patent Owner. IPR2020-01539 Patent 10,588,554 B2 Before GEORGE R. HOSKINS, ROBERT L. KINDER, and AMANDA F. WIEKER, Administrative Patent Judges. WIEKER, Administrative Patent Judge. JUDGMENT Final Written Decision Determining Some Challenged Claims Unpatentable 35 U.S.C. § 318(a) IPR2020-01539 Patent 10,588,554 B2 2 I. INTRODUCTION A. Background Apple Inc. (“Petitioner”) filed a Petition requesting an inter partes review of claims 1-28 (“challenged claims”) of U.S. Patent No. 10,588,554 B2 (Ex. 1001, “the ’554 patent”). Paper 3 (“Pet.”). Masimo Corporation (“Patent Owner”) waived filing a Preliminary Response. Paper 7. We instituted an inter partes review of all challenged claims 1-28 on all asserted grounds of unpatentability, pursuant to 35 U.S.C. § 314. Paper 8 (“Inst. Dec.”). After institution, Patent Owner filed a Response (Paper 23, “PO Resp.”) to the Petition, Petitioner filed a Reply (Paper 26, “Pet. Reply”), and Patent Owner filed a Sur-reply (Paper 30, “PO Sur-reply”).1 An oral hearing was held on December 7, 2021, and a transcript of the hearing is included in the record. Paper 40 (“Tr.”). We issue this Final Written Decision pursuant to 35 U.S.C. § 318(a) and 37 C.F.R. § 42.73. For the reasons set forth below, Petitioner has met its burden of showing, by a preponderance of the evidence, that challenged claims 1-12 and 14-28 of the ’554 patent are unpatentable. However, Petitioner has not met its burden of showing, by a preponderance of the evidence, that challenged claim 13 is unpatentable. 1 After the Sur-reply was filed, we authorized Petitioner to file an Identification of Testimony. Paper 36. IPR2020-01539 Patent 10,588,554 B2 3 B. Related Proceedings The parties identify the following matters related to the ’554 patent: Masimo Corporation v. Apple Inc., Civil Action No. 8:20-cv-00048 (C.D. Cal.) (filed Jan. 9, 2020); Apple Inc. v. Masimo Corporation, IPR2020-01538 (PTAB Sept. 2, 2020) (also challenging claims 1-7 and 20-28 of the ’554 patent); Apple Inc. v. Masimo Corporation, IPR2020-01520 (PTAB Aug. 31, 2020) (challenging claims of U.S. Patent No. 10,258,265 B1); Apple Inc. v. Masimo Corporation, IPR2020-01521 (PTAB Sept. 2, 2020) (challenging claims of U.S. Patent No. 10,292,628 B1); Apple Inc. v. Masimo Corporation, IPR2020-01523 (PTAB Sept. 9, 2020) (challenging claims of U.S. Patent No. 8,457,703 B2); Apple Inc. v. Masimo Corporation, IPR2020-01524 (PTAB Aug. 31, 2020) (challenging claims of U.S. Patent No. 10,433,776 B2); Apple Inc. v. Masimo Corporation, IPR2020-01526 (PTAB Aug. 31, 2020) (challenging claims of U.S. Patent No. 6,771,994 B2); Apple Inc. v. Masimo Corporation, IPR2020-01536 (PTAB Aug. 31, 2020) (challenging claims of U.S. Patent No. 10,588,553 B2); and Apple Inc. v. Masimo Corporation, IPR2020-01537 (PTAB Aug. 31, 2020) (challenging claims of U.S. Patent No. 10,588,553 B2). Pet. 3-4; Paper 5, 1-3. Patent Owner further identifies the following pending patent applications, among other issued and abandoned applications, that claim priority to, or share a priority claim with, the ’554 patent: U.S. Patent Application No. 16/834,538; U.S. Patent Application No. 16/449,143; and IPR2020-01539 Patent 10,588,554 B2 4 U.S. Patent Application No. 16/805,605. Paper 5, 1-2. C. The ’554 Patent The ’554 patent is titled “Multi-Stream Data Collection System for Noninvasive Measurement of Blood Constituents,” and issued on March 17, 2020, from U.S. Patent Application No. 16/544,713, filed August 19, 2019. Ex. 1001, codes (21), (22), (45), (54). The ’554 patent claims priority through a series of continuation and continuation-in-part applications to Provisional Application Nos. 61/078,228 and 61/078,207, both filed July 3, 2008. Id. at codes (60), (63). The ’554 patent discloses a two-part data collection system including a noninvasive sensor that communicates with a patient monitor. Id. at 2:38- 40. The sensor includes a sensor housing, an optical source, and several photodetectors, and is used to measure a blood constituent or analyte, e.g., oxygen or glucose. Id. at 2:29-35, 64-65. The patient monitor includes a display and a network interface for communicating with a handheld computing device. Id. at 2:45-48. IPR2020-01539 Patent 10,588,554 B2 5 Figure 1 of the ’554 patent is reproduced below. Figure 1 illustrates a block diagram of data collection system 100 including sensor 101 and monitor 109. Id. at 11:47-58. Sensor 101 includes optical emitter 104 and detectors 106. Id. at 11:59-63. Emitters 104 emit light that is attenuated or reflected by the patient’s tissue at measurement site 102. Id. at 14:3-7. Detectors 106 capture and measure the light attenuated or reflected from the tissue. Id. In response to the measured light, detectors 106 output detector signals 107 to monitor 109 through front-end interface 108. Id. at 14:7-10, 26-32. Sensor 101 also may include tissue shaper 105, which may be in the form of a convex surface that: (1) reduces the thickness of the patient’s measurement site; and (2) provides more surface area from which light can be detected. Id. at 11:2-14. Monitor 109 includes signal processor 110 and user interface 112. Id. at 15:16-18. “[S]ignal processor 110 includes processing logic that determines measurements for desired analytes . . . based on the signals received from the detectors.” Id. at 15:21-24. User interface 112 presents IPR2020-01539 Patent 10,588,554 B2 6 the measurements to a user on a display, e.g., a touch-screen display. Id. at 15:46-56. The monitor may be connected to storage device 114 and network interface 116. Id. at 15:60-16:11. The ’554 patent describes various examples of sensor devices. Figures 14D and 14F, reproduced below, illustrate detector portions of sensor devices. Figure 14D illustrates portions of a detector submount and Figure 14F illustrates portions of a detector shell. Id. at 6:44-47. As shown in Figure 14D, multiple detectors 1410c are located within housing 1430 and under transparent cover 1432, on which protrusion 605b (or partially cylindrical protrusion 605) is disposed. Id. at 35:36-39, 36:30-37. Figure 14F illustrates a detector shell 306f including detectors 1410c on substrate 1400c. Id. at 37:9-17. Substrate 1400c is enclosed by shielding enclosure 1490 and noise shield 1403, which include window 1492a and window 1492b, respectively, placed above detectors 1410c. Id. Alternatively, cylindrical housing 1430 may be disposed under noise shield 1403 and may enclose detectors 1410c. Id. at 37:47-48. IPR2020-01539 Patent 10,588,554 B2 7 Figures 4A and 4B, reproduced below, illustrate an alternative example of a tissue contact area of a sensor device. Figures 4A and 4B illustrate arrangements of protrusion 405 including measurement contact area 470. Id. at 23:18-24. “[M]easurement site contact area 470 can include a surface that molds body tissue of a measurement site.” Id. “For example, . . . measurement site contact area 470 can be generally curved and/or convex with respect to the measurement site.” Id. at 23:39-43. The measurement site contact area may include windows 420-423 that “mimic or approximately mimic a configuration of, or even house, a plurality of detectors.” Id. at 23:49-63. D. Illustrative Claim Of the challenged claims, claims 1 and 20 are independent. Claim 1 is illustrative and is reproduced below. 1. A physiological measurement system comprising: [a] a physiological sensor device comprising: [b] a plurality of emitters configured to emit light into tissue of a user; [c] at least four detectors, wherein each of the at least four detectors has a corresponding window that allows light to pass through to the detector; IPR2020-01539 Patent 10,588,554 B2 8 [d] a wall that surrounds at least the at least four detectors; and [e] a cover that operably connects to the wall and that is configured to be located between tissue of the user and the at least four detectors when the physiological sensor device is worn by the user, wherein: [f] the cover comprises a single protruding convex surface, and [g] at least a portion of the cover is sufficiently rigid to cause tissue of the user to conform to at least a portion of a shape of the single protruding convex surface when the physiological sensor device is worn by the user; and [h] a handheld computing device in wireless communication with the physiological sensor device, wherein the handheld computing device comprises: [i] one or more processors configured to wirelessly receive one or more signals from the physiological sensor device, the one or more signals responsive to at least a physiological parameter of the user; [j] a touch-screen display configured to provide a user interface, wherein: [k] the user interface is configured to display indicia responsive to measurements of the physiological parameter, and [l] an orientation of the user interface is configurable responsive to a user input; and [m] a storage device configured to at least temporarily store at least the measurements of the physiological parameter. IPR2020-01539 Patent 10,588,554 B2 9 Ex. 1001, 44:51-45:21 (bracketed identifiers a-m added). Independent claim 20 includes limitations substantially similar to limitations [a]-[h] of claim 1. Id. at 46:31-52. E. Applied References Petitioner relies upon the following references: Aizawa, U.S. Patent Application Publication No. 2002/0188210 A1, filed May 23, 2002, published December 12, 2002 (Ex. 1006, “Aizawa”); Inokawa et al., Japanese Patent Application Publication No. 2006-296564 A, filed April 18, 2005, published November 2, 2006 (Ex. 1007, “Inokawa”);2 Ohsaki et al., U.S. Patent Application Publication No. 2001/0056243 A1, filed May 11, 2001, published December 27, 2001 (Ex. 1009, “Ohsaki”); Y. Mendelson et al., “A Wearable Reflectance Pulse Oximeter for Remote Physiological Monitoring,” Proceedings of the 28th IEEE EMBS Annual International Conference, 912-915 (2006) (Ex. 1010, “Mendelson-2006”); and Bergey, U.S. Patent No. 3,789,601, filed July 15, 1971, issued February 5, 1974 (Ex. 1016, “Bergey”). Pet. 12. Petitioner also submits, inter alia, a Declaration of Dr. Thomas W. Kenny, Ph.D. (Ex. 1003) and a Second Declaration of Dr. Kenny (Ex. 1047). Patent Owner submits, inter alia, the Declaration of Dr. Vijay K. Madisetti (Ex. 2004). The parties also provide deposition testimony from Dr. Kenny and Dr. Madisetti, including from this proceeding and others. Exs. 1041-1043, 2006-2009, 2027. 2 Petitioner relies on a certified English translation of Inokawa (Ex. 1008). Ex. 1008, 24. In this Decision, we also refer to the translation. IPR2020-01539 Patent 10,588,554 B2 10 F. Asserted Grounds of Unpatentability We instituted an inter partes review based on the following grounds. Inst. Dec. 9, 32. Claims Challenged 35 U.S.C. § References/Basis 1-7, 20-28 103 Aizawa, Inokawa, Ohsaki, Mendelson-2006 8-19 103 Aizawa, Inokawa, Ohsaki, Mendelson-2006, Bergey II. DISCUSSION A. Claim Construction For petitions filed on or after November 13, 2018, a claim shall be construed using the same claim construction standard that would be used to construe the claim in a civil action under 35 U.S.C. § 282(b). 37 C.F.R. § 42.100(b) (2019). Petitioner submits that no claim term requires express construction. Pet. 11. Patent Owner submits that claim terms should be given their ordinary and customary meaning, consistent with the Specification. PO Resp. 9. We agree that no claim terms require express construction. Nidec Motor Corp. v. Zhongshan Broad Ocean Motor Co. Ltd., 868 F.3d 1013, 1017 (Fed. Cir. 2017). B. Principles of Law A claim is unpatentable under 35 U.S.C. § 103(a) if “the differences between the subject matter sought to be patented and the prior art are such that the subject matter as a whole would have been obvious at the time the invention was made to a person having ordinary skill in the art to which said subject matter pertains.” KSR Int’l Co. v. Teleflex Inc., 550 U.S. 398, 406 IPR2020-01539 Patent 10,588,554 B2 11 (2007). The question of obviousness is resolved on the basis of underlying factual determinations, including (1) the scope and content of the prior art; (2) any differences between the claimed subject matter and the prior art; (3) the level of skill in the art; and (4) objective evidence of nonobviousness.3 Graham v. John Deere Co., 383 U.S. 1, 17-18 (1966). When evaluating a combination of teachings, we must also “determine whether there was an apparent reason to combine the known elements in the fashion claimed by the patent at issue.” KSR, 550 U.S. at 418 (citing In re Kahn, 441 F.3d 977, 988 (Fed. Cir. 2006)). Whether a combination of elements would have produced a predictable result weighs in the ultimate determination of obviousness. Id. at 416-417. In an inter partes review, the petitioner must show with particularity why each challenged claim is unpatentable. Harmonic Inc. v. Avid Tech., Inc., 815 F.3d 1356, 1363 (Fed. Cir. 2016); 37 C.F.R. § 42.104(b). The burden of persuasion never shifts to Patent Owner. Dynamic Drinkware, LLC v. Nat’l Graphics, Inc., 800 F.3d 1375, 1378 (Fed. Cir. 2015). To prevail, Petitioner must support its challenge by a preponderance of the evidence. 35 U.S.C. § 316(e); 37 C.F.R. § 42.1(d). We analyze the challenges presented in the Petition in accordance with the above-stated principles. C. Level of Ordinary Skill in the Art Petitioner identifies the appropriate level of skill in the art as that possessed by a person having “a Bachelor of Science degree in an academic discipline emphasizing the design of electrical, computer, or software 3 Patent Owner has not presented objective evidence of non-obviousness. IPR2020-01539 Patent 10,588,554 B2 12 technologies, in combination with training or at least one to two years of related work experience with capture and processing of data or information.” Pet. 10 (citing Ex. 1003 ¶¶ 1-18, 20-21). “Additional education in a relevant field or industry experience may compensate for one of the other aspects of the . . . characteristics stated above.” Id. at 11. Patent Owner makes several observations regarding Petitioner’s identified level of skill in the art but, “[f]or this proceeding, [Patent Owner] nonetheless applies Petitioner’s asserted level of skill.” PO Resp. 10. We adopt Petitioner’s assessment as set forth above, which appears consistent with the level of skill reflected in the Specification and prior art. D. Obviousness over the Combined Teachings of Aizawa, Inokawa, Ohsaki, and Mendelson-2006 Petitioner contends that claims 1-7 and 20-28 of the ’554 patent would have been obvious over the combined teachings of Aizawa, Inokawa, Ohsaki, and Mendelson-2006. Pet. 41-87; see also Pet. Reply 1-36. Patent Owner disagrees. PO Resp. 11-67; see also PO Sur-reply 1-27. Based on our review of the parties’ arguments and the cited evidence of record, we determine that Petitioner has met its burden of showing by a preponderance of the evidence that claims 1-7 and 20-28 are unpatentable. Overview of Aizawa (Ex. 1006) Aizawa is a U.S. patent application publication titled “Pulse Wave Sensor and Pulse Rate Detector,” and discloses a pulse wave sensor that detects light output from a light emitting diode and reflected from a patient’s artery. Ex. 1006, codes (54), (57). IPR2020-01539 Patent 10,588,554 B2 13 Figure 1(a) of Aizawa is reproduced below. Figure 1(a) is a plan view of a pulse wave sensor. Id. ¶ 23. As shown in Figure 1(a), pulse wave sensor 2 includes light emitting diode (“LED”) 21, four photodetectors 22 symmetrically disposed around LED 21, and holder 23 for storing LED 21 and photodetectors 22. Id. Aizawa discloses that, “to further improve detection efficiency, . . . the number of the photodetectors 22 may be increased.” Id. ¶ 32, Fig. 4(a). “The same effect can be obtained when the number of photodetectors 22 is 1 and a plurality of light emitting diodes 21 are disposed around the photodetector 22.” Id. ¶ 33. Figure 1(b) of Aizawa is reproduced below. IPR2020-01539 Patent 10,588,554 B2 14 Figure 1(b) is a sectional view of the pulse wave sensor. Id. ¶ 23. As shown in Figure 1(b), pulse wave sensor 2 includes drive detection circuit 24 for detecting a pulse wave by amplifying the outputs of photodetectors 22. Id. ¶ 23. Arithmetic circuit 3 computes a pulse rate from the detected pulse wave and transmitter 4 transmits the pulse rate data to an “unshown display.” Id. The pulse rate detector further includes outer casing 5 for storing pulse wave sensor 2, acrylic transparent plate 6 mounted to detection face 23a of holder 23, and attachment belt 7. Id. Aizawa discloses that LED 21 and photodetectors 22 “are stored in cavities 23b and 23c formed in the detection face 23a” of the pulse wave sensor. Id. ¶ 24. Detection face 23a “is a contact side between the holder 23 and a wrist 10, respectively, at positions where the light emitting face 21s of the light emitting diode 21 and the light receiving faces 22s of the photodetectors 22 are set back from the above detection face 23a.” Id. ¶ 24. Aizawa discloses that “a subject carries the above pulse rate detector 1 on the inner side of his/her wrist 10 . . . in such a manner that the light emitting face 21s of the light emitting diode 21 faces down (on the wrist 10 side).” Id. ¶ 26. Furthermore, “the above belt 7 is fastened such that the acrylic transparent plate 6 becomes close to the artery 11 of the wrist 10. Thereby, adhesion between the wrist 10 and the pulse rate detector 1 is improved.” Id. ¶¶ 26, 34. Overview of Inokawa (Ex. 1007) Inokawa is a Japanese published patent application titled “Optical Vital Sensor, Base Device, Vital Sign Information Gathering System, and Sensor Communication Method,” and discloses a pulse sensor device. Ex. 1008 code (57), ¶ 6. IPR2020-01539 Patent 10,588,554 B2 15 Figure 1 of Inokawa is reproduced below. Figure 1 illustrates a schematic view of a pulse sensor. Id. ¶ 56. Pulse sensor 1 includes box-shaped sensor unit 3 and flexible annular wristband 5. Id. ¶ 57. Sensor unit 3 includes a top surface with display 7 and control switch 9, and a rear surface (sensor-side) with optical device component 11 for optically sensing a user’s pulse. Id. Figure 2 of Inokawa is reproduced below. Figure 2 illustrates a schematic view of the rear surface of the pulse sensor. Id. ¶ 58. The rear-side (sensor-side) of pulse sensor 1 includes a pair of IPR2020-01539 Patent 10,588,554 B2 16 light-emitting elements, i.e., green LED 21 and infrared LED 23, as well as photodiode 25 and lens 27. Id. In various embodiments, Inokawa discloses that the sensor-side lens is convex. See id. ¶¶ 99, 107. Green LED 21 is used to sense “the pulse from the light reflected off of the body (i.e.[,] change in the amount of hemoglobin in the capillary artery),” and infrared LED 23 is used to sense body motion from the change in reflected light. Id. ¶ 59. The pulse sensor stores this information in memory. Id. ¶ 68. To read and store information, the pulse sensor includes a CPU that “performs the processing to sense pulse, body motion, etc. from the signal . . . and temporarily stores the analysis data in the memory.” Id. ¶ 69. Figure 3 of Inokawa is reproduced below. Figure 3 illustrates a schematic view of a pulse sensor mounted to a base device. Id. ¶ 60. Pulse sensor 1 is depicted as mounted to base device 17, which “is a charger with communication functionality.” Id. When so mounted, sensor optical device component 11 and base optical device IPR2020-01539 Patent 10,588,554 B2 17 component 41 face each other in close proximity. Id. ¶ 66. In this position, pulse sensor 1 can output information to the base device through the coupled optical device components. Id. ¶ 67. Specifically, the pulse sensor CPU performs the controls necessary to transmit pulse information using infrared LED 23 to photodetector 45 of base device 17. Id. ¶¶ 67, 70, 76. In an alternative embodiment, additional sensor LEDs and base photodetectors can be used to efficiently transmit data and improve accuracy. Id. ¶ 111. Overview of Ohsaki (Ex. 1009) Ohsaki is a U.S. patent application publication titled “Wristwatch-type Human Pulse Wave Sensor Attached on Back Side of User’s Wrist,” and discloses an optical sensor for detecting a pulse wave of a human body. Ex. 1009, code (54), ¶ 3. Figure 1 of Ohsaki is reproduced below. Figure 1 illustrates a cross-sectional view of pulse wave sensor 1 attached on the back side of user’s wrist 4. Id. ¶¶ 12, 16. Pulse wave sensor 1 includes detecting element 2 and sensor body 3. Id. ¶ 16. IPR2020-01539 Patent 10,588,554 B2 18 Figure 2 of Ohsaki, reproduced below, illustrates further detail of detecting element 2. Figure 2 illustrates a mechanism for detecting a pulse wave. Id. ¶ 13. Detecting element 2 includes package 5, light emitting element 6, light receiving element 7, and translucent board 8. Id. ¶ 17. Light emitting element 6 and light receiving element 7 are arranged on circuit board 9 inside package 5. Id. ¶¶ 17, 19. “[T]ranslucent board 8 is a glass board which is transparent to light, and attached to the opening of the package 5. A convex surface is formed on the top of the translucent board 8.” Id. ¶ 17. “[T]he convex surface of the translucent board 8 is in intimate contact with the surface of the user’s skin,” preventing detecting element 2 from slipping off the detecting position of the user’s wrist. Id. ¶ 25. By preventing the detecting element from moving, the convex surface suppresses “variation of the amount of the reflected light which is emitted from the light emitting element 6 and reaches the light receiving element 7 by being reflected by the surface of the IPR2020-01539 Patent 10,588,554 B2 19 user’s skin.” Id. Additionally, the convex surface prevents penetration by “noise such as disturbance light from the outside.” Id. Sensor body 3 is connected to detecting element 2 by signal line 13. Id. ¶ 20. Signal line 13 connects detecting element 2 to drive circuit 11, microcomputer 12, and a monitor display (not shown). Id. Drive circuit 11 drives light emitting element 6 to emit light toward wrist 4. Id. Detecting element 2 receives reflected light which is used by microcomputer 12 to calculate pulse rate. Id. “The monitor display shows the calculated pulse rate.” Id. Mendelson-2006 (Ex. 1010) Mendelson-2006 is a journal article titled “A Wearable Reflectance Pulse Oximeter for Remote Physiological Monitoring,” and discloses a wireless wearable pulse oximeter connected to a personal digital assistant (“PDA”). Ex. 1010, 1.4 Figure 1 of Mendelson-2006 is reproduced below. 4 Petitioner cites to the page numbers added to Exhibit 1010, rather than the native page numbering that accompanies the article. See, e.g., Pet. 20-22. We follow Petitioner’s numbering scheme. IPR2020-01539 Patent 10,588,554 B2 20 Figure 1 illustrates a sensor module attached to the skin (top), and a photograph of a disassembled sensor module and receiver module (bottom). The sensor module includes an optical transducer, a stack of round printed circuit boards, and a coin cell battery. Id. at 2. Figure 2 of Mendelson-2006 is reproduced below. Figure 2 depicts a system block diagram of the wearable, wireless, pulse oximeter including the sensor module (top) and the receiver module (bottom). Id. The sensor module includes at least one light-emitting diode (“LED”), a photodetector, signal processing circuitry, an embedded microcontroller, and an RF transceiver. Id. at 1-2. Mendelson-2006 discloses that a concentric array of discrete photodetectors could be used to increase the amount of backscattered light detected by a reflectance type pulse oximeter sensor. Id. at 4. The receiver module includes an embedded IPR2020-01539 Patent 10,588,554 B2 21 microcontroller, an RF transceiver for communicating with the sensor module, and a wireless module for communicating with the PDA. Id. at 2. As a PDA for use with the system, Mendelson-2006 discloses “the HP iPAQ h4150 PDA because it can support both 802.11b and Bluetooth™ wireless communication” and “has sufficient computational resources.” Id. at 3. Mendelson-2006 further discloses that [t]he use of a PDA as a local terminal also provides a low-cost touch screen interface. The user-friendly touch screen of the PDA offers additional flexibility. It enables multiple controls to occupy the same physical space and the controls appear only when needed. Additionally, a touch screen reduces development cost and time, because no external hardware is required. . . . The PDA can also serve to temporarily store vital medical information received from the wearable unit. Id. The PDA is shown in Figure 3 of Mendelson-2006, reproduced below. Figure 3 illustrates a sample PDA and its graphical user interface (“GUI”). Id. Mendelson-2006 explains that the GUI allows the user to interact with the wearable system. Id. “The GUI was configured to present the input and IPR2020-01539 Patent 10,588,554 B2 22 output information to the user and allows easy activation of various functions.” Id. “The GUI also displays the subject’s vital signs, activity level, body orientation, and a scrollable PPG waveform that is transmitted by the wearable device.” Id. For example, the GUI displays numerical oxygen saturation (“SpO2”) and heart rate (“HR”) values. Id. Independent Claim 1 Petitioner contends that claim 1 would have been obvious over the combined teachings of Aizawa, Inokawa, Ohsaki, and Mendelson-2006. Pet. 41-65. Below, we set forth how the combination of prior art references teaches or suggests the claim limitations that are not disputed by the parties. For those limitations and reasons for combining the references that are disputed, we examine each of the parties’ contentions and then provide our analysis. i. “A physiological measurement system comprising” The cited evidence supports Petitioner’s undisputed contention that Aizawa satisfies the subject matter of the preamble.5 Pet. 41; see, e.g., Ex. 1006 ¶ 2 (“The present invention relates to a pulse wave sensor for detecting the pulse wave of a subject.”). ii. “[a] a physiological sensor device comprising” The cited evidence supports Petitioner’s undisputed contention that Aizawa discloses a physiological sensor device including a pulse rate detector. Pet. 41-43; see, e.g., Ex. 1006 ¶ 23 (pulse wave sensor 2), Figs. 1(a)-(b). 5 Whether the preamble is limiting need not be resolved because Petitioner shows sufficiently that the preamble’s subject matter is satisfied by the art. IPR2020-01539 Patent 10,588,554 B2 23 iii. “[b] a plurality of emitters configured to emit light into tissue of a user” Petitioner’s Undisputed Contentions Petitioner contends that Aizawa discloses an emitter-LED 21-and also states that, in certain embodiments, multiple LEDs may be employed. Pet. 13, 22. Patent Owner does not dispute this contention, and we agree with Petitioner. See Ex. 1006 ¶¶ 23 (“LED 21”), 32 (“The arrangement of the light emitting diode 21 and the photodetectors 22 is not limited to this.”). For example, Aizawa explains that “[t]he same effect can be obtained when the number of photodetectors 22 is 1 and a plurality of light emitting diodes 21 are disposed around the photodetector.” Id. ¶ 33. Petitioner also contends that Inokawa teaches a sensor with two LEDs-a green LED to sense pulse and an infrared LED to sense body motion. Pet. 16, 23. Petitioner also contends that when Inokawa’s sensor is mounted on a base device, the infrared LED also is used to wirelessly transmit vital information to the base device. Id. at 17-18, 23-24. Patent Owner does not dispute these contentions, and we agree. Inokawa teaches a pair of LEDs 21, 23, where “the basic function of the S-side green LED 21 is to sense the pulse from the light reflected off of the body . . ., while the S- side infrared LED 23 serves to sense body motion from the change in this reflected light.” Ex. 1008 ¶¶ 58-59. Inokawa also explains that “vital sign information stored in the memory 63 [of the sensor], such as pulse and body motion, is transmitted to the base device 17 using the S-side infrared LED 23 of the pulse sensor 1 and the B-side PD 45 of the base device 17,” such that “there is no need to use a special wireless communication circuit or a communication cable.” Id. ¶¶ 76-77. IPR2020-01539 Patent 10,588,554 B2 24 Petitioner’s Disputed Contentions Moreover, Petitioner contends that a person of ordinary skill in the art would have been motivated to modify Aizawa “to include an additional LED as taught by Inokawa to improve the detected pulse wave by distinguishing between blood flow detection and body movement.” Pet. 23-24, 26, 43-45. According to Dr. Kenny, “one of ordinary skill would have recognized that this would improve Aizawa’s sensor by enabling it to account for motion load through use of the second LED, by detecting and recording body motion in addition to blood flow.” Ex. 1003 ¶ 120 (cited at Pet. 43). As a second and independent motivation, Petitioner also contends that such a modification also would have provided “additional functionality, including that of a wireless communication method,” which would have “eliminate[d] problems associated with a physical cable, and, as taught by Inokawa, without requiring a separate RF circuit.” Pet. 23-24. Petitioner contends that although Aizawa discloses data transmission, Aizawa “is silent about how such transmission would be implemented.” Id. at 24. According to Petitioner, a skilled artisan “would have recognized that Aizawa’s LED could have been used for wireless data communication with a personal computer to eliminate problems associated with a physical cable, and, as taught by Inokawa, without requiring a separate RF,” which “would result in enhanced accuracy of the transmitted information.” Id. According to Dr. Kenny, “as one of ordinary skill would have recognized, the LEDs provided on the sensor can be used not only to detect pulse rate, but also to ‘accurately, easily, and without malfunction’ transmit sensed data to a base station.” Ex. 1003 ¶ 122. IPR2020-01539 Patent 10,588,554 B2 25 To illustrate its proposed modification, Petitioner includes annotated and modified views of Aizawa’s Figures 1(a) and 1(b), reproduced below. Pet. 25; see also id. at 44 (similar figures); Ex. 1003 ¶ 76. Petitioner’s annotated and modified figures depict the sensor of Aizawa with an added “LED B” (illustrated in light purple), as Petitioner contends would have been rendered obvious by Inokawa. Id. at 26-27, 44-45; see also Ex. 1003 ¶¶ 71-79, 110-129. Patent Owner’s Arguments Patent Owner disputes Petitioner’s contentions regarding the obviousness of modifying Aizawa to include two emitters. See PO Resp. 51-58; Sur-reply 23-25. IPR2020-01539 Patent 10,588,554 B2 26 First, Patent Owner argues that neither Aizawa nor Inokawa discloses a device with both multiple detectors and multiple emitters in the same sensor, because Aizawa’s embodiments have either a single emitter and multiple detectors (e.g., Ex. 1006, Fig. 1(a)) or multiple emitters and a single detector (e.g., id. ¶ 33), and Inokawa discloses multiple emitters and a single detector (e.g., Ex. 1008, Fig. 2). See PO Resp. 51-52 (citing, e.g., Ex. 1006 ¶ 33, Figs. 1, 2, 4, 5; Ex. 1008 ¶ 58, Fig. 2; Ex. 2004 ¶¶ 100-102). Patent Owner concludes, therefore, that a person of ordinary skill in the art would not have added a second emitter to Aizawa, when Aizawa already discloses an embodiment with multiple LEDs, i.e., an embodiment with only a single detector. PO Resp. 52 (citing, e.g., Ex. 2004 ¶ 103). Patent Owner argues that Dr. Kenny’s testimony in this regard “conflicts with the references themselves” and so “is not credible.” PO Resp. 35-36 (citing Ex. 1003 ¶¶ 55, 77). Second, Patent Owner argues that the evidence does not support either of Petitioner’s two proffered motivations for modifying Aizawa to include two emitters. As to the first motivation (to measure body movement using a second emitter), Patent Owner asserts that Dr. Kenny erroneously testifies that Aizawa cannot do this with its single emitter. PO Resp. 53 (citing, e.g., Ex. 1006 ¶ 15; Ex. 2007, 400:7-401:10; Ex. 2004 ¶ 104). Patent Owner argues that “Petitioner admits that Aizawa’s sensor ‘already records and accounts for’ motion load.” PO Resp. 53 (citing, e.g., Pet. 26; Ex. 1006 ¶ 15; PO Resp. 36; Ex. 2004 ¶ 84). Thus, Patent Owner contends that the proposed motivation would not realize an improvement over Aizawa alone. Id. IPR2020-01539 Patent 10,588,554 B2 27 As to Petitioner’s second motivation (to enable transmission of data to a base device using an optical communication link), Patent Owner argues that “Aizawa already includes a wireless transmitter . . . so Aizawa does not need to incorporate Inokawa’s base device [optical] data transmission arrangement.” PO Resp. 54 (citing, e.g., Ex. 1006 ¶¶ 23, 28, 35; Ex. 2004 ¶¶ 105-106). Indeed, Patent Owner argues “Dr. Kenny acknowledged Aizawa identifies no problems with Aizawa’s form of data transmission.” Id. at 55 (citing Ex. 2007, 409:13-410:2). Patent Owner further argues that “Aizawa’s goal is ‘real-time measuring’ with the transmitter ‘transmitting the measured pulse rate data to a display’” but that “Inokawa’s data transfer approach does not allow real-time display of measurements.” Id. at 55 (citing, e.g., Ex. 1006 ¶¶ 4, 15; Ex. 1008 ¶¶ 70, 74; Ex. 2004 ¶ 107). Patent Owner insists Inokawa does not aid Petitioner’s case, because Inokawa discloses the benefits of using a second emitter in only two situations, i.e., first, to improve over a “cable” communication and, second, to avoid use of a “dedicated wireless communication circuit,” whereas “Aizawa already incorporates a transmitter into its design.” Id. at 56 (citing, e.g., Ex. 1008 ¶ 4; Ex. 1006 ¶¶ 23, 28; Ex. 2004 ¶ 108). Third, Patent Owner accuses Petitioner and Dr. Kenny of overlooking further complications that would ensue from modifying Aizawa to have two emitters. Patent Owner argues that Dr. Kenny overlooked how placing “two LEDs in close proximity may cause thermal interference that could create significant issues for sensor performance,” and would require “structural changes” to Aizawa’s configuration. PO Resp. 57 (citing, e.g., Ex. 2004 ¶¶ 109-110; Ex. 1019, 59-60). Patent Owner also argues that “Petitioner widened Aizawa’s emitter cavity to accommodate the extra LED with no IPR2020-01539 Patent 10,588,554 B2 28 explanation or recognition of this change,” which could impact optical performance of the device. Id. at 57-58 (citing, e.g., Ex. 2004 ¶¶ 109-111). Petitioner’s Reply Concerning Petitioner’s first motivation, Petitioner asserts that Aizawa “is silent on whether it uses the computed motion load to improve the detection signal” and thereby provide a “more reliable” pulse reading, which is Petitioner’s asserted improvement to Aizawa. Pet. Reply 28-29 (citing, e.g., Ex. 1003 ¶¶ 61-63, 71-72; Ex. 2007, 401:11-402:4; Ex. 1047 ¶ 51). Moreover, Petitioner contends that by using multiple LEDs at different wavelengths, “two separate signals [can] be collected[, which] allows noise arising from body motion to be better isolated and accounted for.” Id. at 29 (citing Ex. 1047 ¶ 51). Concerning Petitioner’s second motivation, Petitioner maintains that Inokawa’s use of two emitters having different wavelengths to upload data to a base device using optical communication advantageously improves the accuracy of the transmission by providing checksum information. Pet. Reply 22-23 (citing, e.g., Ex. 1003 ¶ 123; Ex. 2007, 407:7-408:20, 416:5- 15; Ex. 1047 ¶ 52). As to the “other complications” that Patent Owner alleges would result from the proposed modification, Petitioner asserts “such minor issues are ‘part of what [a person of ordinary skill in the art] would bring . . . to the problem and would know how to make the changes needed.’” Pet. Reply 29-30 (quoting Ex. 2007, 384:8-388:12; Ex. 1047 ¶ 53). IPR2020-01539 Patent 10,588,554 B2 29 Patent Owner’s Sur-reply Concerning Petitioner’s first motivation, Patent Owner argues that Inokawa’s disclosure is just as sparse as Aizawa’s disclosure regarding how to use optical data to measure body movement. Sur-reply 23 (citing Ex. 1008 ¶ 59). Patent Owner also asserts that “Petitioner cites nothing in Inokawa that suggests” that Inokawa’s two emitter data gathering is more reliable or otherwise superior to Aizawa’s single emitter data gathering. Id. at 23-24. Concerning Petitioner’s second motivation, Patent Owner argues that the proposed modification eliminates Aizawa’s ability to conduct “real-time collection and display of physiological measurements-a key goal of Aizawa’s system.” Id. at 24. Patent Owner also faults Petitioner for not specifying how a person of ordinary skill in the art would have solved the alleged “additional cost, energy use, and thermal problems” that would ensue from using two emitters in the Aizawa device. Id. at 24-25. Analysis Upon review of the foregoing, we conclude a preponderance of the evidence supports Petitioner’s contention that it would have been obvious to replace Aizawa’s single near infrared LED 21 with an infrared LED and a green LED, in light of Inokawa. First, a person of ordinary skill in the art would have been motivated to make this replacement to improve the pulse measurements recorded by Aizawa’s detector 1. Inokawa teaches that the infrared LED’s signal can be used “to detect vital signs” such as “body motion,” and the green LED’s IPR2020-01539 Patent 10,588,554 B2 30 signal can be “used to detect pulse.” Ex. 1008, Fig. 2, ¶¶ 14, 58-59; Ex. 1003 ¶¶ 64, 72, 75-78; Ex. 1047 ¶¶ 50-51. Patent Owner correctly points out that Aizawa describes its single-emitter detector 1 as transmitting its pulse data to “a device for computing the amount of motion load from the pulse rate.” Ex. 1006 ¶¶ 15, 28, 35. But, this description is the only disclosure in Aizawa cited by Patent Owner as relating to computing a motion characteristic of the user. Further, we are unable to discern any other disclosure in Aizawa relating to motion computation, or what Aizawa proposes to do with its motion computation. See id. Based on the sparse nature of Aizawa’s disclosure concerning motion load, it is not clear exactly what Aizawa proposes to do with the computed motion load, after it is computed. See, e.g., Ex. 1047 ¶ 51 (“Aizawa is silent on whether it uses the computed motion load to improve the detection signal.”). Aizawa does, however, describe the motion load as being computed “from the pulse rate,” rather than being an input to the pulse rate calculation. Ex. 1006 ¶¶ 15, 35. Dr. Kenny, when asked whether it was his understanding that “Aizawa’s sensor could not account for motion load?”, answered that “Aizawa’s sensor attempts to prevent motion load rather than account for it.” Ex. 2007, 400:7-11. He explained that, because Aizawa uses only a single emitter with a single wavelength, “what [Aizawa] sees as a signal would be some mixture of pulse rate and motion load if there was no effort to prevent motion load,” so Aizawa seeks to solve the problem of “prevent[ing] motion load from corrupting the pulse rate signal.” Id. at 400:12-401:10. Dr. Kenny did not further explain this distinction between preventing and accounting for motion load in his deposition testimony cited by the parties as IPR2020-01539 Patent 10,588,554 B2 31 relating to this issue. Id. at 400:7-402:4. We do not rely on this distinction as a basis for our present decision, because we find no express support for it in Aizawa’s disclosure (see Ex. 1006 ¶¶ 15, 28, 35), and it is not explained in persuasive detail by Dr. Kenny. We nonetheless credit Dr. Kenny’s declaration testimony that a person of ordinary skill in the art, upon reviewing Inokawa’s disclosure of using two emitters of different wavelengths to calculate a user’s pulse and motion separately, would understand that these two separate measurements would enable the device to calculate a “more reliable” pulse rate because it “allows noise arising from body motion to be better isolated and accounted for.” Ex. 1047 ¶ 51; see also Ex. 1003 ¶¶ 73, 77, 122-123, 126. Aizawa does not disclose using the computed motion load in this fashion, so it appears that this would improve upon the accuracy of Aizawa’s pulse measurements, by using the computed motion load to isolate and account for noise. See Ex. 1006 ¶¶ 15, 28, 35. Dr. Madisetti also offers no meaningful opposing testimony in this regard. See, e.g., Ex. 2004 ¶ 104. Instead, Dr. Madisetti incorrectly reads Dr. Kenny’s motivation testimony as being limited to the desirability of adding the bare ability to measure body movement to Aizawa. See id. In fact, Dr. Kenny further testified that it would have been beneficial to use the measured body movement to improve the pulse measurement of the device. See Ex. 1003 ¶ 120; Ex. 1047 ¶ 51. Dr. Madisetti does not address that testimony. See Ex. 2004 ¶ 104. Thus, because Dr. Madisetti’s testimony sets up a straw man to attack, rather than directly addressing the entirety of Dr. Kenny’s testimony in this regard, Dr. Kenny’s testimony stands unrebutted in the record before us. IPR2020-01539 Patent 10,588,554 B2 32 Dr. Kenny’s testimony also makes intuitive sense that measuring the user’s motion separately from the user’s pulse, for example by using two interrogating emitters of two different wavelengths, would provide a reliable means of correcting the pulse data for motion artifacts by using the separately measured motion data, rather than by trying to segregate these two components in the single data stream provided by Aizawa’s single emitter device. See, e.g., Ex. 1047 ¶ 51. We, therefore, are persuaded by Dr. Kenny’s unrebutted testimony that using two emitters of different wavelengths would improve Aizawa’s device in this way. Independently, we are also persuaded that a person of ordinary skill in the art would have been motivated to replace Aizawa’s single near infrared LED 21 with an infrared LED and a green LED, to provide a reliable method of uploading pulse data stored by Aizawa’s wrist-worn pulse rate detector 1 to another device for display to the user. Inokawa expressly touts such optically-based uploading of data from Inokawa’s wrist-worn sensor 1 to Inokawa’s base device 17 as a benefit of incorporating two emitters in sensor 1. See Ex. 1008, Figs. 3, 19, ¶¶ 3-7, 14, 76-77, 109-111. Inokawa identifies two specific benefits of this optically-based data communication means. First, the infrared LED can transmit the pulse data, and the green LED can separately transmit “checksum” information to increase the accuracy of data transmission. Id. at Fig. 19, ¶¶ 14, 109-111. Second, using light emitters in this fashion to perform two functions (data collection by emitting light into the user’s wrist, and data transmission by emitting light to photodetectors in a base device) obviates the need for providing “a special wireless communication circuit [in the wrist-worn sensor 1] or a communication cable.” Id. ¶¶ 3-7, 76-77. IPR2020-01539 Patent 10,588,554 B2 33 Patent Owner correctly points out that Aizawa already has a “transmitter” 4 for uploading pulse data stored by Aizawa’s wrist-worn pulse rate detector 1 to another device for processing and for display to the user. Ex. 1006, Fig. 1(b), ¶¶ 15, 23, 28, 35. However, Aizawa’s Figure 1(b) illustrates transmitter 4 only as an empty box contained within outer casing 5, and Aizawa’s written description does not provide further structural details concerning transmitter 4. See id. In particular, Aizawa does not describe exactly how transmitter 4 transmits its data to the other device. See id. Patent Owner contends, and Dr. Madisetti and Dr. Kenny both testify, that Aizawa’s transmitter 4 is a “wireless” transmitter. See, e.g., PO Resp. 54; Ex. 2004 ¶¶ 49, 105-106, 112; Ex. 2007, 403:17-22, 414:19- 21. They all appear to equate “wireless” communication to radio frequency communication, and not to include optical communication, even though both radio frequency and optical communication do not use a wire. Petitioner disagrees that Aizawa discloses any specific form of data transmission, including wireless transmission. See Tr. 71:5-72:3 (“[T]he transmitter disclosure in Aizawa, they don’t say its’s a wireless transmitter. That was a conjuration by [Patent Owner]. They don’t specify whether it’s a wired or wireless.”). We assume, for this decision, that Aizawa expressly contemplates radio frequency communication as one embodiment by which transmitter 4 may transmit data to devices other than detector 1. Patent Owner argues, and Dr. Madisetti testifies, that Aizawa’s express disclosure goes even further. They assert Aizawa’s “goal” is to measure and display pulse data in real time during exercise, using the wireless transmitter. See, e.g., Ex. 2004 ¶¶ 106-108, 111. We find that IPR2020-01539 Patent 10,588,554 B2 34 Aizawa does not support this assertion. Instead, Aizawa discusses prior art devices that “estimat[e] a burden on the heart of a person who takes exercise by real-time measuring his/her heart rate at the time of exercise” (Ex. 1006 ¶ 4 (emphasis added)), and then describes Aizawa’s detector 1 as having a transmitter for transmitting the measured pulse rate data to another device for display (id. ¶ 15). Aizawa does not indicate when this transmission occurs. Aizawa also refers to “noise caused by the shaking of the body of the subject” as a problem to be addressed (id. ¶ 6), but this problem occurs regardless of whether the shaking results from exercise or the normal movement of the user’s wrist over the course of the day. Thus, Aizawa does not tout, as an important feature of Aizawa’s invention, the real time display of pulse rate data during exercise, regardless of whether the data gathered by Aizawa’s wrist-worn detector 1 is transmitted wirelessly or otherwise. Id. ¶¶ 4, 6, 15. No doubt, a person of ordinary skill in the art would have viewed the capability of a wrist-worn pulse detector to transmit its pulse data to another device for display in real time while the user is exercising to be a desirable feature in some cases, even if this is not one of Aizawa’s specifically stated goals. See, e.g., Ex. 1003 ¶ 67 (Dr. Kenny stating: “By wirelessly transmitting the collected data wirelessly, Mendelson 2006’s system provides ‘numerous advantages,’ . . . .”); Ex. 2009, 393:6-14 (Dr. Kenny agreeing that a person of ordinary skill in the art “would have seen the ability to wirelessly transmit collected data as an advantage”). Nonetheless, Inokawa expressly discloses that, in other cases, the benefits achieved by wireless transmission can be outweighed by obviating the need for the wrist-worn sensor to include a special wireless communication circuit. See IPR2020-01539 Patent 10,588,554 B2 35 Ex. 1008 ¶¶ 3-7 (discussing problems associated with wireless transmission, such as the need for a dedicated circuit, which is avoided by Inokawa’s system that risks “few malfunctions” and has a “simple structure”), 76-77 (“As a result, there is no need to use a special wireless communication circuit . . ., which makes it possible to transmit vital sign information to the base device 17 accurately, easily, and without malfunction.”). We therefore conclude that Petitioner’s case for obviousness in this regard is supported by a preponderance of the evidence. See, e.g., In re Urbanski, 809 F.3d 1237, 1243-44 (Fed. Cir. 2016) (persons of ordinary skill in the art may be motivated to pursue desirable properties of one prior art reference, even at the expense of foregoing a benefit taught by another prior art reference). We disagree with Patent Owner’s argument that Petitioner’s case for obviousness is deficient on the basis that neither Aizawa nor Inokawa expressly discloses a wrist-worn sensor device that has both a plurality of emitters and at least four detectors, as claim 1 recites. Obviousness does not require “‘some motivation or suggestion to combine the prior art teachings’ [to] be found in the prior art.” KSR, 550 U.S. at 407, 415-418. Nor does it require the bodily incorporation of Inokawa’s device into Aizawa’s device. See, e.g., In re Keller, 642 F.2d 413, 425 (CCPA 1981) (test for obviousness is not whether the features of one reference may be bodily incorporated into the structure of the other reference, but rather is “what the combined teachings of the references would have suggested to those of ordinary skill in the art”); see also In re Merck & Co., 800 F.2d 1091, 1097 (Fed. Cir. 1986) (nonobviousness is not established by attacking references individually when unpatentability is predicated upon a combination of prior art disclosures). Instead, “[a] person of ordinary skill is also a person of IPR2020-01539 Patent 10,588,554 B2 36 ordinary creativity, not an automaton,” and “in many cases a person of ordinary skill will be able to fit the teachings of multiple patents together like pieces of a puzzle.” KSR, 550 U.S. at 420-421. In this case, we are persuaded that a person of ordinary skill in the art would have been motivated to modify Aizawa’s wrist-worn detector 1 to replace its single near infrared LED 21 with an infrared LED and a green LED, based on Inokawa, for all the reasons provided above. A person of ordinary skill in the art would additionally have known to keep all four detectors 22 that are already present in Aizawa’s detector 1, so that “[e]ven when the attachment position of the sensor is dislocated, a pulse wave can be detected accurately,” as disclosed by Aizawa. Ex. 1006 ¶¶ 9, 27. In short, the combination of Aizawa and Inokawa teaches that having multiple emitters is beneficial, and having multiple detectors is beneficial, for different and not inconsistent reasons. Finally, we agree with Petitioner’s position that any thermal interference and power consumption issues that may arise in Aizawa’s wrist-worn pulse detector, by using two emitters instead of one emitter, are well within the capabilities of POSITA to solve. We credit Dr. Kenny’s testimony in this regard. See Ex. 1003 ¶¶ 78-79, 121; Ex. 1047 ¶ 53. For example, Dr. Kenny acknowledges that Aizawa already discloses adding additional emitters. Ex. 1003 ¶ 71 (citing Ex. 1006 ¶¶ 32-33). Dr. Kenny further testifies that this modification “amount[s] to nothing more than the use of a known technique [i.e., Inokawa’s use of two emitters in a wrist-worn pulse detector] to improve similar devices [i.e., Aizawa’s wrist-worn pulse detector] in the same way, and combining prior art elements according to known methods to yield predictable results.” Id. IPR2020-01539 Patent 10,588,554 B2 37 ¶¶ 78 (“Furthermore, one of ordinary skill would have readily understood how to select different photodiodes with different sensitivities to detect the different wavelengths of light emitted by the two LEDs.”), 121. Patent Owner cites several portions of Dr. Kenny’s deposition testimony that, in Patent Owner’s view, indicate Dr. Kenny fails to appreciate the significance of the thermal effects, optical interference complications, and power consumption needs, that are posed by adding a second emitter to Aizawa’s device, and fails to explain how these issues would have been overcome. See PO Resp. 55, 57-58 (citing Ex. 2007, 379:17-21, 384:8-388:16, 394:11-395:22, 405:2-11, 409:13-410:2; Ex. 2009, 381:18-382:8, 383:22-385:9, 390:5-392:3). We have reviewed this deposition testimony, and we conclude Patent Owner overstates its significance. It establishes, at most, that Dr. Kenny did not expressly address these issues in his declaration (Exhibit 1003), but Dr. Kenny’s opinion is that these issues would have been within the capability of a person of ordinary skill in the art to resolve. Based on the evidentiary record presented to us, we agree with Dr. Kenny. For example, Inokawa discloses a wrist-worn pulse sensor 1 having two emitters 21 and 23 in close proximity to each other. See Ex. 1008, Figs. 1-2. An artisan must be presumed to know something about the art apart from what the relied-upon references disclose. See In re Jacoby, 309 F.2d 513, 516 (CCPA 1962). Dr. Madisetti’s testimony opposing Dr. Kenny’s foregoing opinion is premised solely on Dr. Kenny’s alleged failure to explain how the issues that arise from adding a second emitter to Aizawa would have been solved; Dr. Madisetti does not provide any affirmative reason why these issues would have been difficult for a person of ordinary skill in the art to solve, in IPR2020-01539 Patent 10,588,554 B2 38 the context of Aizawa’s device or wrist-worn pulse sensing devices in general. See Ex. 2004 ¶ 109. Thus, we conclude a person of ordinary skill in the art would have been motivated to replace Aizawa’s single near infrared LED 21 with an infrared LED and a green LED, and would have had a reasonable expectation of success in doing so. iv. “[c] at least four detectors, wherein each of the at least four detectors has a corresponding window that allows light to pass through to the detector” The cited evidence supports Petitioner’s undisputed contention that Aizawa discloses at least four detectors, each stored in a separate cavity 23c, which would have been understood to be “openings or windows that mirror specific detector placement layouts.” Pet. 45, 49-51; see, e.g., Ex. 1006 ¶¶ 23 (“four phototransistors 22”), 24 (“stored in cavities” and “set back from . . . detection face 23a”), Figs. 1(a)-1(b); Ex. 1003 ¶¶ 138-143. v. “[d] a wall that surrounds at least the at least four detectors” The cited evidence supports Petitioner’s undisputed contention that Aizawa discloses holder 23, which is a wall that surrounds detectors 22, as well as other elements. Pet. 51-52; see, e.g., Ex. 1006 ¶ 23 (“holder 23 for storing . . . light emitting diode 21 and the photodetectors 22”), Fig. 1(b). IPR2020-01539 Patent 10,588,554 B2 39 vi. “[e-g] a cover that operably connects to the wall and that is configured to be located between tissue of the user and the at least four detectors when the physiological sensor device is worn by the user, wherein: the cover comprises a single protruding convex surface, and at least a portion of the cover is sufficiently rigid to cause tissue of the user to conform to at least a portion of a shape of the single protruding convex surface when the physiological sensor device is worn by the user” Petitioner’s Undisputed Contentions Petitioner contends that Aizawa discloses a cover, i.e., an “acrylic transparent plate positioned between the photodetectors and the wrist,” to improve adhesion between the sensor and the subject’s wrist. Pet. 15. Patent Owner does not dispute this contention, and we agree with Petitioner. Aizawa discloses that “acrylic transparent plate 6 is provided on the detection face 23a of the holder 23 to improve adhesion to the wrist 10.” Ex. 1006 ¶ 34, Fig. 1(b) (depicting transparent plate 6 between sensor 2 and wrist 10). Petitioner also contends that Ohsaki teaches a wrist-worn sensor that includes a “translucent board” having a convex surface that contacts the user’s skin. Pet. 19, 34. Patent Owner does not dispute this contention, and we agree with Petitioner. Ohsaki discloses that sensor 1 includes detecting element 2 and sensor body 3, and is “worn on the back side of the user’s wrist.” Ex. 1009 ¶ 16. Ohsaki discloses that detecting element 2 includes package 5 and “translucent board 8[,which] is a glass board which is transparent to light, [and is] attached to the opening of the package 5. A convex surface is formed on the top of the translucent board 8.” Id. ¶ 17. As seen in Ohsaki’s Figure 2, translucent board 8 has a single protruding convex surface, which is placed between a user’s tissue and a light receiving element (e.g., photodetector) 7 when the sensor is worn. Id. at Fig. 2. As IPR2020-01539 Patent 10,588,554 B2 40 also seen in Figure 2, the board 8 is operably connected to the walls of sensor package 5. Id. ¶ 17 (“The translucent board 8 is . . . attached to the opening of the package 5.”), Fig. 2. Petitioner also contends that Ohsaki’s Figure 2 depicts the user’s tissue conforming to the shape of the convex surface of the cover, such that the convex surface would have been understood to be “sufficiently rigid.” Pet. 56. Patent Owner does not dispute this contention, and we agree with Petitioner. Ohsaki’s Figure 2 depicts the user’s tissue 4 conforming to the shape of the protruding convex surface when the sensor is worn by the user. Ex. 1009 ¶ 17 (“The translucent board 8 is a glass board.”), Fig. 2; see, e.g., Ex. 1003 ¶¶ 160 (testifying as to the convex surface’s rigidity), 164. Petitioner’s Disputed Contentions Petitioner further contends that a person of ordinary skill in the art “would have found it obvious to modify the sensor’s flat cover [in Aizawa] . . . to include a lens/protrusion . . . similar to Ohsaki’s translucent board 8, so as to [1] improve adhesion between the user’s wrist and the sensor’s surface, [2] improve detection efficiency, and [3] protect the elements within sensor housing.” Pet. 36-37 (citing, e.g., Ex. 1003 ¶¶ 94-97; Ex. 1009 ¶¶ 25), 53-54 (citing, e.g., Ex. 1003 ¶¶ 71-102). Petitioner contends that Ohsaki’s convex surface is in “intimate contact” with the user’s tissue, which prevents slippage of the sensor and increases signal strength because “variation of the amount of the reflected light . . . that reaches the light receiving element 7 is suppressed” and because “disturbance light from the outside” is prevented from penetrating board 8, as compared to a sensor with a flat surface. Id. at 34-36 (citing, e.g., Ex. 1003 ¶ 95; quoting Ex. 1009 ¶ 25). IPR2020-01539 Patent 10,588,554 B2 41 Accordingly, Petitioner contends that a person of ordinary skill in the art would have modified Aizawa’s sensor to include a cover with a single convex protrusion, as taught by Ohsaki, “that operably connects to the wall and that is configured to be located between tissue of the user and the at least four detectors when the sensor is worn by the user.” Pet. 36-37 (citing, e.g., Ex. 1003 ¶¶ 94-97), 53-54 (citing, e.g., Ex. 1003 ¶¶ 147-153). Petitioner also contends that an ordinarily skilled artisan would have configured the cover to be “sufficiently rigid to cause tissue of the user to conform to the shape of the surface when worn by the user.” Id. at 56 (citing, e.g., Ex. 1003 ¶¶ 61-69, 71-161). Petitioner contends this modification would have been “nothing more than the use of a known technique to improve similar devices in the same way,” i.e., “simply improving Aizawa-Inokawa’s transparent plate 6 that has a flat surface to improve adhesion to a subject’s skin and reduce variation in the signals detected by the sensor.” Pet. 37 (citing Ex. 1003 ¶ 98). Further according to Petitioner, “the elements of the combined system would each perform functions they had been known to perform prior to the combination-Aizawa-Inokawa’s transparent plate 6 would remain in the same position, performing the same function, but with a convex surface as taught by Ohsaki.” Id. at 37-38 (citing, e.g., Ex. 1003 ¶¶ 94-99), 53-54 (citing Ex. 1003 ¶¶ 71-102). To illustrate its proposed modification, Petitioner includes two annotated versions of Aizawa’s Figure 1(b), both of which are reproduced below. Pet. 36-37 (citing Ex. 1003 ¶¶ 94-97), 53-54. IPR2020-01539 Patent 10,588,554 B2 42 Petitioner’s annotated figure on the left depicts Aizawa’s sensor, modified to include LED B (see supra Section II.D.5.iii) and with a flat “light permeable cover” (illustrated with blue outline); Petitioner’s annotated figure on the right depicts Aizawa’s sensor, again modified to include LED B (see supra Section II.D.5.iii) and with a convex “light permeable cover” (illustrated with yellow shading and green outline). Patent Owner’s Arguments Patent Owner argues that a person of ordinary skill in the art would not have been motivated to modify Aizawa’s sensor to include Ohsaki’s convex cover. PO Resp. 20-51; PO Sur-Reply 3-22. First, Patent Owner argues that the proposed modification “changes Ohsaki’s structure and eliminates the longitudinal shape that gives Ohsaki’s translucent board the ability to prevent slipping.” PO Resp. 21. This argument is premised on Patent Owner’s contention that Ohsaki’s convex cover must be rectangular, with the cover’s long direction aligned with the length of the user’s forearm, to avoid interacting with bones in the wrist and forearm. Id. at 22-24 (citing, e.g., Ex. 2004 ¶¶ 52-54; Ex. 1009 ¶¶ 6, 19, 23, 24); see also PO Sur-reply 3-10. According to Patent Owner, Ohsaki teaches that “aligning the sensor’s longitudinal direction with the IPR2020-01539 Patent 10,588,554 B2 43 circumferential direction of the user’s arm undesirably results in ‘a tendency [for Ohsaki’s sensor] to slip off.’” PO Resp. 23-24 (citing Ex. 1009 ¶ 19). Thus, Patent Owner contends that Petitioner’s proposed modification would “chang[e] Ohsaki’s rectangular board into a circular shape,” which “would eliminate the advantages discussed above” because it “cannot be placed in any longitudinal direction and thus cannot coincide with the longitudinal direction of the user’s wrist.” Id. at 24 (citing Ex. 2004 ¶¶ 55- 57). Patent Owner presents annotated Figures depicting what it contends is Ohsaki’s disclosed sensor placement as compared to that of the proposed modification, reproduced below. Patent Owner’s annotated Figure on the left depicts a rectangular sensor placed between a user’s radius and ulna, while Patent Owner’s annotated Figure on the right depicts a circular sensor placed across a user’s radius and ulna. Based on these annotations, Patent Owner argues that the proposed “circular shape would press on the user’s arm in all directions and thus cannot avoid the undesirable interaction with the user’s bone structure,” such that a skilled artisan “would have understood such a change would eliminate IPR2020-01539 Patent 10,588,554 B2 44 Ohsaki’s benefit of preventing slipping.” Id. at 25 (citing, e.g., Ex. 2004 ¶¶ 54-61). Second, Patent Owner argues that Ohsaki requires its sensor be placed on the back of the user’s wrist to achieve any benefits, but that such a location would have been unsuitable for Aizawa’s sensor. PO Resp. 30. Specifically, Patent Owner argues that Aizawa’s sensor must be worn on the palm side of the wrist, close to radial and ulnar arteries, which is the side opposite from where Ohsaki’s sensor is worn. Id. at 30-35 (citing, e.g., Ex. 1006 ¶¶ 2, 7, 9, 26, 27, 36; Ex. 2004 ¶¶ 66-70). According to Patent Owner, Ohsaki teaches that the sensor’s convex surface has a tendency to slip when placed on the palm side of the wrist, i.e., in the location taught by Aizawa. Id. at 36-38 (citing, e.g., Ex. 1009 ¶¶ 19, 23, 24; Ex. 2004 ¶¶ 74- 80). Thus, Patent Owner argues that a person of ordinary skill in the art “would not have been motivated to use Ohsaki’s longitudinal board- designed to be worn on the back side of a user’s wrist-with Aizawa’s palm-side sensor.” Id. at 39. Similarly, Patent Owner argues that Aizawa teaches away from the proposed modification because Aizawa teaches that its flat acrylic plate improves adhesion on the palm side of the wrist, while Ohsaki teaches that its convex board “has a tendency to slip” on the palm side of the wrist. Id. at 39-41 (citing, e.g., Ex. 2004 ¶¶ 82-84). Third, Patent Owner argues that a person of ordinary skill in the art would not have placed Ohsaki’s convex cover over Aizawa’s peripheral detectors because the convex cover would condense light toward the center and away from Aizawa’s detectors, which would decrease signal strength. PO Resp. 42-48 (citing, e.g., Ex. 2004 ¶¶ 85-97). Patent Owner also contends that Petitioner and Dr. Kenny admitted as much in a related IPR2020-01539 Patent 10,588,554 B2 45 proceeding. Id. at 43-44 (citing, e.g., Ex. 2019, 45; Ex. 2020, 69-70). Patent Owner also relies on Figure 14B of the ’554 patent to support its position. Id. at 44-45 (citing Ex. 1001, 36:3-6, 36:13-15). Additionally, Patent Owner argues that its position is also supported by Inokawa, which also uses a convex lens to direct light toward the center but, in Inokawa’s structure, the light is directed from peripheral emitters toward a central detector. Id. at 48-50 (citing, e.g., Ex. 1008 ¶¶ 15, 58). In light of the foregoing, Patent Owner argues that a person of ordinary skill in the art would have understood that the proposed modification would have decreased signal strength by directing light away from Aizawa’s peripheral detectors. Id. at 45-48. Fourth and finally, Patent Owner argues that a person of ordinary skill in the art “would have understood that Aizawa’s flat plate would provide better protection than a convex surface” because it “would be less prone to scratches.” Id. at 50-51 (citing Ex. 1008 ¶ 106). Petitioner’s Reply Concerning Patent Owner’s first and second arguments, Petitioner responds that Ohsaki does not disclose the shape of its protrusion, other than its convexity as shown in Figures 1 and 2, nor does Ohsaki require a rectangular shape or placement on the back of the wrist in order to achieve the disclosed benefits. Pet. Reply 13-19 (citing, e.g., Ex. 1047 ¶¶ 16-27). Moreover, Petitioner asserts that “even if Ohsaki’s translucent board 8 were somehow understood to be rectangular, obviousness does not require ‘bodily incorporation’ of features from one reference into another”; rather, a person of ordinary skill in the art “would have been fully capable of modifying Aizawa to feature a light permeable protruding convex cover to obtain the IPR2020-01539 Patent 10,588,554 B2 46 benefits” taught by Ohsaki. Id. at 15-16 (citing, e.g., Ex. 1047 ¶ 21). Similarly, regarding the location of the sensor, Petitioner asserts, [E]ven assuming for the sake of argument that a POSITA would have understood Aizawa’s sensor as being limited to placement on the palm side of the wrist, and would have understood Ohsaki’s sensor’s “tendency to slip” when arranged on the front side as informing consideration of Ohsaki’s teachings with respect to Aizawa, that would have further motivated the [person of ordinary skill in the art] to implement a light permeable convex cover in Aizawa’s sensor, to improve detection efficiency of that sensor when placed on the palm side. Id. at 17 (citing, e.g., Ex. 1047 ¶ 25). In other words, Ohsaki’s disclosure that a convex surface suppresses variation in reflected light would have motivated an artisan to add such a surface to Aizawa to improve detection efficiency of that sensor when placed on the palm side. Id. at 18. Concerning Patent Owner’s third argument, Petitioner responds that adding a convex cover to Aizawa’s sensor would not decrease signal strength but, instead, “would improve Aizawa’s signal-to-noise ratio by causing more light backscattered from tissue to strike Aizawa’s photodetectors than would have with a flat cover” because such a cover improves light concentration across the entire lens and does not direct it only towards the center. Id. at 20-28 (citing, e.g., Ex. 1047 ¶¶ 29-45). Petitioner asserts that Patent Owner and Dr. Madisetti “ignore[] the well-known principle of reversibility,” by which “a ray going from P to S will trace the same route as one from S to P.” Pet. Reply 20-22 (quoting Ex. 1040, 92; citing, e.g., Ex. 1040, 87-92; Ex. 1049, 106-111; Ex. 1047 ¶ 31). When applied to Aizawa’s sensor, Petitioner contends that any condensing benefit achieved by a convex cover would thus direct emitted light toward Aizawa’s peripheral detectors. Id. at 21-22 (citing, e.g., IPR2020-01539 Patent 10,588,554 B2 47 Ex. 1047 ¶¶ 31-35). Although Dr. Madisetti “refused to acknowledge this basic principle of reversibility during deposition,” Petitioner contends it is applied in Aizawa. Id. at 22 (citing, e.g., Ex. 1041, 89:12-19; Ex. 1003 ¶ 127 (citing Ex. 1006 ¶ 33); Ex. 1047 ¶ 34). Petitioner also asserts that Patent Owner and Dr. Madisetti overlook the fact that light rays reflected by body tissue will be scattered and diffuse and will approach the detectors “from various random directions and angles.” Pet. Reply 22-24 (citing, e.g., Ex. 1019, 52, 86, 90; Ex. 1042, 803; Ex. 1047 ¶¶ 36-41; Ex. 2006, 163:12-164:2). This scattered and diffuse light, according to Petitioner, means that Ohsaki’s convex cover cannot “focus all light at the center of the sensor device,” as Patent Owner argues. Id. at 23. Instead, due to the random nature of this scattered light, Petitioner asserts that a person of ordinary skill in the art would have understood that “Ohsaki’s convex cover provides at best a slight refracting effect, such that light rays that otherwise would have missed the detection area are instead directed toward that area as they pass through the interface provided by the cover.” Id. at 24 (citing, e.g., Ex. 1047 ¶ 42). Petitioner applies this understanding to Aizawa, and asserts that using a cover with a convex protrusion in Aizawa would “enable backscattered light to be detected within a circular active detection area surrounding” a central light source, thereby “allow[ing] a larger fraction of light randomly backscattered from tissue to be detected within the active detection area surrounding [the light] source.” Id. at 24-27 (citing, e.g., Ex. 1019, 86, 90; Ex. 1047 ¶¶ 42-48). Petitioner relies upon the following illustration of this alleged effect. Pet. Reply 27 (citing Ex. 1047 ¶ 47). IPR2020-01539 Patent 10,588,554 B2 48 The above illustration depicts backscattered light reflecting off user tissue in various directions, such that it impinges upon the peripheral detectors from various random angles and directions. Id. According to Petitioner, this “allow[s] the detector to capture light that otherwise would have been missed by the detectors, regardless of their location within the sensor device.” Id. Finally, Petitioner dismisses Patent Owner’s reliance on Figure 14B of the ’554 patent because it “is not an accurate representation of light that has been reflected from a tissue measurement site. For example, the light rays (1420) shown in FIG. 14B are collimated (i.e., travelling paths parallel to one another), and each light ray’s path is perpendicular to the detecting surface.” Pet. Reply 26 (citing, e.g., Ex. 1047 ¶ 45). Concerning Patent Owner’s fourth argument, Petitioner responds that even if a flat surface might be less prone to scratching, that possible disadvantage would have been weighed against the “known advantages of applying Ohsaki’s teachings,” and would not negate a motivation to combine. Id. at 28 (citing, e.g., Ex. 1047 ¶ 49). Patent Owner’s Sur-reply Concerning Patent Owner’s first and second arguments, Patent Owner reiterates its position that Ohsaki’s purported benefits attach only to a sensor with a rectangular convex surface that is located on the back of the wrist, IPR2020-01539 Patent 10,588,554 B2 49 and that “even small changes in sensor orientation or measurement location result in slippage.” PO Sur-reply 3-14, 7. Concerning Patent Owner’s third argument, Patent Owner asserts that Petitioner’s Reply improperly presents several new arguments, relying on new evidence, as compared with the Petition. Id. at 16 (regarding reversibility), 18-22. Patent Owner argues that Dr. Kenny and Petitioner have not overcome their admissions that a convex lens directs light toward the center. Id. at 14- 16, 19. Moreover, Patent Owner argues that Petitioner’s discussion of the principle of reversibility is “irrelevant” because it “assumes conditions that are not present when tissue scatters and absorbs light.” Id. at 16. The random nature of backscattered light, in Patent Owner’s view, “hardly supports Petitioner’s argument that light will necessarily travel the same paths regardless of whether the LEDs and detectors are reversed,” and is irrelevant to the central issue presented here of “whether changing Aizawa’s flat surface to a convex surface results in more light on Aizawa’s peripherally located detectors.” Id. at 16-17. Patent Owner also asserts that Petitioner mischaracterizes Patent Owner’s position, which is not that Ohsaki’s cover with a convex protrusion “focuses all light to a single point” at the center of the sensor as Petitioner characterizes it. PO Sur-reply 18-19. Patent Owner’s position, rather, is that Petitioner has not shown that a person of ordinary skill in the art “would have been motivated to change Aizawa’s flat surface to a convex surface to improve signal strength.” Id. at 19. In Patent Owner’s view, by arguing that the convex cover provides only a “slight refracting effect,” Petitioner IPR2020-01539 Patent 10,588,554 B2 50 undermines its contention that providing such a cover would have improved detection efficiency. Id. Patent Owner also argues that Petitioner’s contention that a convex cover allows more light collection generally is a new theory not supported by Dr. Kenny’s original declaration. Id. at 20. Moreover, Patent Owner argues that “Petitioner’s theory is unavailing because it fails to consider the greater decrease in light at the detectors due to light redirection to a more central location.” Id. at 20-21. According to Patent Owner, any light redirected from the sensor’s edge could not make up for the loss of signal strength from light redirected away from the detectors and toward the center. Id. at 21. Concerning Patent Owner’s fourth argument, Patent Owner argues that Petitioner does not dispute Patent Owner’s position that a flat cover would be less prone to scratches and offers “no plausible advantages for its asserted combination.” Id. at 22. Moreover, Patent Owner argues that “the risk of scratches is not merely a disadvantage-it directly undermines Petitioner’s motivation to add a convex cover to ‘protect the elements within the sensor housing.’” Id. Analysis As noted above, Petitioner provides three rationales to support its contention that a person of ordinary skill in the art would have provided “a light permeable cover with a protruding convex surface,” such as that taught by Ohsaki, to Aizawa’s sensor: (1) to improve adhesion between the sensor and the user’s tissue, (2) to improve detection efficiency, and (3) to protect the elements within the sensor housing. Pet. 36-37 (citing, e.g., Ex. 1003 IPR2020-01539 Patent 10,588,554 B2 51 ¶¶ 94-97; Ex. 1009 ¶¶ 25). We conclude all three rationales are supported by the evidence, as follows. Rationales 1 and 2 The evidence of record persuades us that adding a convex cover, such as that taught by Ohsaki, would have improved adhesion between the sensor and the user’s skin, which would have increased the signal strength of the sensor. Ohsaki teaches as much: [T]he convex surface of the translucent board 8 is in intimate contact with the surface of the user’s skin. Thereby it is prevented that the detecting element 2 slips off the detecting position of the user’s wrist 4. If the translucent board 8 has a flat surface, the detected pulse wave is adversely affected by the movement of the user’s wrist 4 as shown in Fig. 4B. However, in the case that the translucent board 8 has a convex surface like the present embodiment, the variation of the amount of the reflected light which is emitted from the light emitting element 6 and reaches the light receiving element 7 by being reflected by the surface of the user’s skin is suppressed. It is also prevented that noise such as disturbance light from the outside penetrates the translucent board 8. Therefore the pulse wave can be detected without being affected by the movement of the user’s wrist 4 as shown in FIG. 4A. Ex. 1009 ¶ 25 (emphases added); see also id. ¶ 27 (“stably fixed”). We credit Dr. Kenny’s testimony that a person of ordinary skill in the art would have been motivated by such teachings to apply a cover with a convex surface to Aizawa to improve that similar device in the same way and to yield predictable results, i.e., to resist movement of the sensor on the user’s wrist. See, e.g., Ex. 1003 ¶¶ 95 (“[T]his contact between the convex surface and the user’s skin is said to prevent slippage, which increases the strength of the signals obtainable by Ohsaki’s sensor.”), 97. We also credit Dr. Kenny’s testimony that, in light of these teachings, a person of ordinary IPR2020-01539 Patent 10,588,554 B2 52 skill in the art would have made such a modification to improve the pulse sensor’s ability to emit light into, and detect light reflected from, the user’s wrist, to generate an improved pulse signal. Ex. 1003 ¶¶ 95, 97, 155; Ex. 1047 ¶¶ 12-13. Indeed, Ohsaki expressly compares the performance of a wrist-worn pulse wave sensor depending on whether translucent board 8 is convex or flat, and concludes the convex surface results in improved performance over the flat surface, especially when the user is moving. Ex. 1009, Figs. 4A-4B, ¶¶ 15, 25 (stating that with “a flat surface, the detected pulse wave is adversely affected by the movement of the user’s wrist 4,” and with “a convex surface like the present embodiment, the variation of the amount of the reflected light” collected by the sensor “is suppressed”). Ohsaki also states that, with a convex surface, “[i]t is also prevented that noise such as disturbance light from the outside penetrates the translucent board 8.” Id. ¶ 25. We also credit Dr. Kenny’s testimony that the proposed modification would have been within the skill level of an ordinary artisan. For example, Dr. Kenny testifies: One of ordinary skill would have combined the teachings of Aizawa-Inokawa and Ohsaki as doing so would have amounted to nothing more than the use of a known technique to improve similar devices in the same way. One of ordinary skill would have recognized that incorporating Ohsaki’s convex surface is simply improving Aizawa-Inokawa’s transparent plate 6 that has a flat surface to improve adhesion to a subject’s skin and reduce variation in the signals detected by the sensor. Furthermore, the elements of the combined system would each perform similar functions they had been known to perform prior to the combination-Aizawa-Inokawa’s transparent plate 6 IPR2020-01539 Patent 10,588,554 B2 53 would remain in the same position, performing the same function, but with a convex surface as taught by Ohsaki. Ex. 1003 ¶ 98; see also id. ¶¶ 84-99, 155, 160. In light of Ohsaki’s express disclosure of the benefits of a convex cover, we credit Dr. Kenny’s testimony that a person of ordinary skill in the art would have been motivated to modify Aizawa as proposed, and would have had a reasonable expectation of success in doing so. We next address Patent Owner’s first through third arguments, each of which implicates Petitioner’s first and second asserted rationales of improved adhesion and detection efficiency. Patent Owner’s first argument is premised on the notion that Ohsaki’s benefits only can be realized with a rectangular convex surface, because such a shape is required to avoid interacting with bones on the back of the user’s forearm. PO Resp. 20-27. We disagree. Ohsaki does not disclose the shape of its convex cover, much less require it be rectangular. In fact, Ohsaki is silent as to the shape of the convex surface. Ohsaki discloses that sensor 1 includes detecting element 2, which includes package 5 within which the sensor components are located. Ex. 1009 ¶ 17. Ohsaki’s convex surface is located on board 8, which is “attached to the opening of the package 5.” Id. Ohsaki provides no further discussion regarding the shape of board 8 or its convex protrusion. We disagree with Patent Owner’s suggestion that the shape of the convex surface can be inferred to be rectangular from Ohsaki’s Figures 1 and 2. PO Resp. 16-17. Ohsaki does not indicate that these figures are drawn to scale, or reflect precise dimensions or shapes of the convex surface. See, e.g., Ex. 1009 ¶ 13 (“schematic diagram”); see also Pet. IPR2020-01539 Patent 10,588,554 B2 54 Reply 14-15; Hockerson-Halberstadt, Inc. v. Avia Group Int’l, 222 F.3d 951, 956 (Fed. Cir. 2000) (“[I]t is well established that patent drawings do not define the precise proportions of the elements and may not be relied on to show particular sizes if the specification is completely silent on the issue.”). To be clear, Ohsaki describes the shape of detecting element 2 as rectangular: “[T]he length of the detecting element from the right side to the left side in FIG. 2 is longer than the length from the upper side to the lower side.” Ex. 1009 ¶ 19. Ohsaki also describes that detecting element 2 is aligned longitudinally with the user’s forearm: “[I]t is desirable that the detecting element 2 is arranged so that its longitudinal direction agrees with the longitudinal direction of the user’s arm,” to avoid slipping off. Id.; see also id. ¶ 9 (“The light emitting element and the light receiving element are arranged in the longitudinal direction of the user’s arm.”). In light of this disclosed rectangular shape of detecting element 2, it is certainly possible that Ohsaki’s convex surface may be similarly shaped. But, it may not be. Contrary to Patent Owner’s argument, Ohsaki neither describes nor requires detecting element 2 to have the same shape as the convex surface of board 8. Accord Pet. Reply. 13-16 (noting also that Ohsaki’s board 8 “is not coextensive with the entire tissue-facing side of detecting element 2”). We have considered the testimony of both Dr. Kenny and Dr. Madisetti on this point. Ex. 1047 ¶¶ 10-11, 14, 16-21; Ex. 2004 ¶¶ 36-41 (relying on Ohsaki’s Figures 1-2 to support his opinion that the convex surface is rectangular). Dr. Madisetti’s reliance on the dimensions of Ohsaki’s figures is unpersuasive. Hockerson-Halberstadt, 222 F.3d at 956. We credit Dr. Kenny’s testimony that Ohsaki does not describe its convex IPR2020-01539 Patent 10,588,554 B2 55 surface as rectangular, because this testimony is most consistent with Ohsaki’s disclosure. Further, Patent Owner suggests that the convex surface must be rectangular, in order to avoid interacting with bones in the user’s forearm. PO Resp. 22-23; PO Sur-Reply 9 (“[A] POSITA would have understood Ohsaki’s convex board must also have a longitudinal shape oriented up-and- down the watch-side of the user’s wrist/forearm.”). Although Ohsaki recognizes that interaction with these bones can cause problems, (see Ex. 1009 ¶¶ 6, 19), we do not agree that the only way to avoid these bones is by aligning a rectangular cover with the longitudinal direction of the user’s forearm. For example, in the annotated Figures provided by Patent Owner, see PO Resp. 23, we discern that the circular sensor that purports to depict the proposed modification would also avoid the bones in the forearm if it were slightly smaller. Patent Owner provides no persuasive explanation to justify the dimensions it provides in this annotated figure, or to demonstrate that such a large sensor would have been required. Indeed, we discern that it would have been within the level of skill of an ordinary artisan to appropriately size a modified sensor to avoid these well-known anatomical obstacles. “A person of ordinary skill is also a person of ordinary creativity, not an automaton.” KSR, 550 U.S. at 421. After all, an artisan must be presumed to know something about the art apart from what the references disclose. See In re Jacoby, 309 F.2d 513, 516 (CCPA 1962). Finally, we do not agree with Patent Owner’s position that Ohsaki’s advantages apply only to rectangular convex surfaces. As discussed, Patent Owner has not shown that Ohsaki’s convex surface is rectangular at all. Moreover, even if Ohsaki’s convex surface is rectangular, when discussing IPR2020-01539 Patent 10,588,554 B2 56 the benefits associated with a convex cover, Ohsaki does not limit those benefits to a cover of any particular shape. Instead, Ohsaki explains that “detecting element 2 is arranged on the user’s wrist 4 so that the convex surface of the translucent board 8 is in intimate contact with the surface of the user’s skin. Thereby it is prevented that the detecting element 2 slips off the detecting position of the user’s wrist 4.” Ex. 1009 ¶ 25; Ex. 1047 ¶ 10. Thus, we agree with Petitioner that Ohsaki’s teaching of a convex surface would have motivated a person of ordinary skill in the art to add such a surface to Aizawa’s circular-shaped sensor, to improve adhesion as taught by Ohsaki. See, e.g., Pet. 34-35. Nothing in Ohsaki’s disclosure limits such a benefit to a specific shape of the convex surface. Ex. 1047 ¶¶ 10, 14, 16- 21 Moreover, Ohsaki contrasts the ability to properly receive reflected light with a convex surface as compared to a flat surface and notes that, in the case that the translucent board 8 has a convex surface . . . the variation of the amount of the reflected light which is emitted from the light emitting element 6 and reaches the light receiving element 7 by being reflected by the surface of the user’s skin is suppressed. It is also prevented that noise such as disturbance light from the outside penetrates the translucent board 8. Therefore the pulse wave can be detected without being affected by the movement of the user’s wrist 4 as shown in FIG. 4A. Ex. 1009 ¶ 25; Ex. 1047 ¶ 11. Again, we agree with Petitioner that Ohsaki’s teaching of a convex surface would have motivated a person of ordinary skill in the art to add such a surface to Aizawa’s sensor, to improve signal strength, as taught by Ohsaki. See, e.g., Pet. 34-37. Again, nothing in Ohsaki’s disclosure limits such a benefit to the shape of the convex surface. Ex. 1047 ¶¶ 10, 12, 13, 16-21. IPR2020-01539 Patent 10,588,554 B2 57 Accordingly, we do not agree that Ohsaki’s disclosed advantages attach only to a rectangular convex surface, or would have been inapplicable to the proposed combination of Aizawa and Ohsaki.6 We have considered Patent Owner’s second argument, that Ohsaki’s benefits are realized only when the sensor and convex surface are placed on the back of the user’s wrist, which is the opposite side of the wrist taught by Aizawa. PO Resp. 30-39. We do not agree. As an initial matter, Petitioner does not propose bodily incorporating the references; Petitioner simply proposes adding a convex cover to Aizawa’s sensor, without discussing where Aizawa’s sensor is used. See, e.g., Pet. 34. In other words, Petitioner’s proposed modification does not dictate any particular placement, whether on the palm side or back side of the wrist. To be sure, Ohsaki’s Figures 3A-3B compare the performance of detecting element 2, including its translucent board 8 having a convex protrusion, and show better performance when the element is attached to the back side of the wrist versus the front side of the wrist, when the user is in motion. See Ex. 1009 ¶¶ 23-24, Figs. 3A-3B. However, we do not agree that these figures support Dr. Madisetti’s conclusion that “Ohsaki indicates a convex surface only prevents slipping on the back (i.e., watch) side of the wrist in a specific orientation, but tends to slip when used in different locations or orientations” such as the palm side of the wrist-particularly in 6 Patent Owner also argues that, to the extent contended by Petitioner, it would not have been obvious to place a rectangular cover on top of Aizawa’s sensor. PO Resp. 27-30. We do not understand Petitioner to have made any such contention and, accordingly, do not address this argument. See, e.g., Pet. 37 (depicting Aizawa’s circular sensor with an added convex cover). IPR2020-01539 Patent 10,588,554 B2 58 comparison to a flat surface such as Aizawa’s. Ex. 2004 ¶¶ 66, 75. Instead, Ohsaki acknowledges that, even when the detecting element is located “on the front [palm] side of the user’s wrist 4, the pulse wave can be detected well if the user is at rest.” Ex. 1009 ¶ 23 (emphasis added). Thus, Ohsaki discloses that, in at least some circumstances, a convex surface located on the front of the user’s wrist achieves benefits. Id. Notably, the claims are not limited to detection during movement or exercise. We credit, instead, Dr. Kenny’s testimony that a person of ordinary skill in the art would have understood from Ohsaki that a convex protrusion will help prevent slippage, even in the context of Aizawa’s sensor. See Ex. 1047 ¶¶ 15, 22-28. This is because the convex protrusion “promot[es] ‘intimate contact with the surface of the user’s skin,’” which “would have increased adhesion and reduced slippage of Aizawa’s sensor when placed on the palm side of a user’s wrist, with associated improvements in signal quality.” Id. ¶¶ 27, 28 (“additional adhesive effect”). Dr. Madisetti testifies that “[b]ased on Aizawa’s teaching that a flat acrylic plate improves adhesion on the palm side of the wrist, and Ohsaki’s teaching that a convex surface tends to slip on the palm side of the wrist, a [person of ordinary skill in the art] would have come to the opposite conclusion from Dr. Kenny: that modifying Aizawa’s flat adhesive plate ‘to include a lens/protrusion . . . similar to Ohsaki’s translucent board’ would not ‘improve adhesion.’” Ex. 2004 ¶ 84; see also id. ¶ 82. We disagree with this reading of Aizawa. It is true that Aizawa’s plate 6 is illustrated as having a flat surface (Ex. 1006, Fig. 1(b)), and that Aizawa states the plate “improve[s] adhesion” (id. ¶ 13). Aizawa further states: “the above belt 7 is fastened such that the acrylic transparent plate 6 becomes close to the IPR2020-01539 Patent 10,588,554 B2 59 artery 11 of the wrist 10,” and “[t]hereby, adhesion between the wrist 10 and the pulse rate detector 1 is improved.” Id. ¶ 26. These disclosures, however, indicate the improved adhesion is provided by the acrylic material of plate 6, not the shape of the surface of the plate, which is never specifically addressed. See also id. ¶¶ 30, 34 (“Since the acrylic transparent plate 6 is provided . . . adhesion between the pulse rate detector 1 and the wrist 10 can be improved . . . .”). Aizawa does not associate this benefit of improved adhesion with the surface shape of the plate, but rather, with the existence of an acrylic plate to begin with. Thus, there is no teaching away from using a convex surface to improve the adhesion of Aizawa’s detector to the user’s wrist. We have considered Patent Owner’s third argument that a convex cover would condense light away from Aizawa’s peripheral detectors, which Patent Owner alleges would decrease signal strength. PO Resp. 42-50. We disagree. There appears to be no dispute that when emitted light passes through user tissue, the light diffuses and scatters as it travels. See, e.g., Pet. Reply 22-26; Tr. 27:18-28:4 (Petitioner’s counsel agreeing that “the incoming light from a detection standpoint is going to be coming from all sorts of different directions because of the randomness caused by the back scattering”), 65:23-66:16 (Patent Owner’s counsel agreeing that light does not simply enter tissue and come back out “like it came out on a mirror”); Ex. 1041, 35:19-37:18 (Patent Owner’s declarant describing light scattering as it travels through tissue, e.g., reflecting off blood, tissue, or other material); Ex. 1043, 28:2-10 (Patent Owner’s declarant agreeing that reflecting light can be a signal for the ’554 patent’s sensor), 61:20-62:4 IPR2020-01539 Patent 10,588,554 B2 60 (explaining that “a light in this context, light emitted from the LEDs is diffused through the skin in that particular context, whatever that is”); Ex. 1047 ¶ 36. The light thus travels at random angles and directions, and no longer travels in a collimated and perpendicular manner. Exhibit 1040,7 Figure 4.12, illustrates the difference between diffuse and collimated light, and is reproduced below: This figure provides at left a photograph and an illustration showing incoming collimated light reflecting from a smooth surface, and at right a photograph and an illustration of incoming collimated light reflecting from a rough surface. See Ex. 1040, 87-88 (original page numbers). The smooth surface provides specular reflection, in which the reflected light rays are collimated like the incoming light rays. See id. The rough surface provides 7 Eugene Hecht, Optics (2nd ed. 1990). IPR2020-01539 Patent 10,588,554 B2 61 diffuse reflection, in which the reflected light rays travel in random directions. See id.; see also Ex. 1047 ¶ 36 (“A [person of ordinary skill in the art] would have understood that light that backscatters from the measurement site (after diffusing through tissue) reaches the active detection area from many random directions and angles.”). Dr. Kenny testifies that Aizawa “detect[s] light that has been ‘partially reflected, transmitted, absorbed, and scattered by the skin and other tissues and the blood before it reaches the detector.’” Ex. 1047 ¶ 36 (quoting Ex. 1019, 36). Dr. Kenny further opines that a convex cover, when added to Aizawa’s sensor with multiple detectors symmetrically arranged about a central light source, “allows light rays that otherwise would have missed the detection area to instead be directed toward that area as they pass through the interface provided by the cover,” thus increasing the light-gathering ability of Aizawa’s sensor. Id. ¶ 42; see also id. ¶ 46. By contrast Dr. Madisetti testifies that “a convex surface condenses light passing through it towards the center of the sensor and away from the periphery.” Ex. 2004 ¶ 86; see also id. ¶¶ 85, 89. We have considered this testimony, however, Dr. Madisetti’s opinions largely are premised upon the behavior of collimated and perpendicular light as depicted in Figure 14B of the challenged patent. See id. ¶ 88. Dr. Madisetti does not explain how light would behave when approaching the sensor from various angles, as it would after being reflected by tissue. Id. ¶¶ 86-89; see also id. ¶¶ 90-97 (addressing motivation and also failing to discuss diffuse, scattered light). In other words, even if Patent Owner is correct that the ’554 patent’s Figure 14B depicts light condensing toward the center, this is not dispositive to the proposed modification, because light reflected by a user’s tissue is IPR2020-01539 Patent 10,588,554 B2 62 scattered and random, and is not collimated and perpendicular as shown in Figure 14B. Ex. 1001, Fig. 14B. Patent Owner and Dr. Madisetti argue that “Petitioner and Dr. Kenny both [previously admitted] that a convex cover condenses light towards the center of the sensor and away from the periphery,” in a different petition filed against a related patent, i.e., in IPR2020-01520. PO Resp. 43-45; Ex. 2004 ¶ 86. The cited portions of the Petition and Dr. Kenny’s declaration from IPR2020-01520 discuss a decrease in the “mean path length” of a ray of light when it travels through a convex lens rather than through a flat surface. See, e.g., Ex. 2020 ¶¶ 118-120. We do not agree that this discussion is inconsistent with Dr. Kenny’s testimony here that, where light is reflected to the detectors at various random angles and directions, more light will reach Aizawa’s symmetrically disposed detectors when travelling through the convex surface than would be reached without such a surface, because light that might have otherwise missed the detectors now will be captured. See, e.g., Ex. 1047 ¶¶ 29-30, 36. We do not discern that the convergence of a single ray of light toward the center, as discussed in IPR2020-01520, speaks to the aggregate effect on all light that travels through the convex surface. We additionally do not agree with Patent Owner’s argument that Petitioner’s Reply presents new arguments and evidence that should have been first presented in the Petition, to afford Patent Owner an adequate opportunity to respond. The Petition proposed a specific modification of Aizawa to include a convex protrusion in the cover, for the purpose of increasing the light gathering ability of Aizawa’s device. See Pet. 34-38. The Patent Owner Response then challenged that contention, with several IPR2020-01539 Patent 10,588,554 B2 63 arguments that Petitioner’s proposed convex protrusion would not operate in the way the Petition alleges it would operate. See PO Resp. 42-50. This opened the door for Petitioner to provide, in the Reply, arguments and evidence attempting to rebut the contentions in the Patent Owner Response. See PTAB Consolidated Trial Practice Guide (Nov. 2019) (“Consolidated Guide”),8 73 (“A party also may submit rebuttal evidence in support of its reply.”). This is what Petitioner did here. The Reply does not change Petitioner’s theory for obviousness; rather, the Reply presents more argument and evidence in support of the same theory for obviousness presented in the Petition. Compare Pet. 34-38, with Reply 20-27. Rationale 3 Petitioner further contends that a person of ordinary skill in the art would have recognized that a cover with a protruding convex surface, such as that taught by Ohsaki, would “protect the elements within the sensor housing” of Aizawa. Pet. 36-37. We are persuaded that adding a convex cover, such as that taught by Ohsaki, would also protect the sensor’s internal components in a manner similar to Aizawa’s flat acrylic plate. Ex. 1003 ¶ 97; see also Ex. 1008 ¶ 15 (noting that a cover “protect[s] the LED or PD”). We disagree with Patent Owner’s fourth argument that a person of ordinary skill in the art would not have modified Aizawa as proposed because a convex cover would be prone to scratches and because other alternatives existed. Patent Owner’s counsel did not dispute, during the oral hearing, that a convex cover would indeed serve to protect the internal 8 Available at https://www.uspto.gov/TrialPracticeGuideConsolidated. IPR2020-01539 Patent 10,588,554 B2 64 sensor components in Aizawa, as Petitioner proposes. Tr. 64:6-65:5 (but noting that a flat cover would also protect and would be less prone to scratches). That a convex cover may be more prone to scratches than Aizawa’s flat cover is one of numerous tradeoffs that a person of ordinary skill in the art would consider in determining whether the benefits of increased adhesion, signal strength, and protection outweigh the potential for a scratched cover. Medichem, S.A. v. Rolabo, S.L., 437 F.3d 1157, 1165 (Fed. Cir. 2006). The record does not support that the possibility of scratches alone would have dissuaded a person of ordinary skill in the art from the proposed modification, to achieve the benefits identified by Petitioner. For the foregoing reasons, we are persuaded by Petitioner’s contentions. vii. “[h] a handheld computing device in wireless communication with the physiological sensor device, wherein the handheld computing device comprises” Petitioner’s Contentions Petitioner contends that the combination of, inter alia, Aizawa and Inokawa teaches a sensor device that is in wireless communication with a base device through its LEDs, wherein that base device is connected further to a PC. Pet. 27, 56-57; see, e.g., Ex. 1006 ¶ 15 (“a transmitter for transmitting the measured pulse rate data to a display for displaying the pulse rate data”); Ex. 1008 ¶¶ 75 (explaining that sensed physiological information is transmitted from the sensor to the base device, when mounted, and further that “[t]he base device 17 . . . transmits this information to the PC 59”), 76, 77. IPR2020-01539 Patent 10,588,554 B2 65 Petitioner also contends that Mendelson-2006 discloses a body-worn pulse oximetry system including a sensor module, a receiver module, and a PDA. Pet. 29, 58-59; see, e.g., Ex. 1010, 1-2 (describing system), Fig. 1 (sensor attached to skin), Fig. 3 (PDA). Petitioner contends that signals acquired by the sensor are received and processed by the receiver module, and then wirelessly transmitted to the PDA. Pet. 30; Ex. 1010, 2. Petitioner contends that wireless transmission to a PDA, as discussed in Mendelson- 2006, provides advantages such as offering “a low-cost touch screen interface,” and “more effective medical care.” Pet. 30-31; Ex. 1010, 3-4. Petitioner contends that a person of ordinary skill in the art “would have also found it obvious to implement the physiological sensor device resulting from the combined teachings of Aizawa, Inokawa, and Ohsaki as part of a physiological measurement system including a handheld computing device, and to enable the physiological sensor device to communicate wirelessly with the handheld computing device,” to obtain the advantages taught by Mendelson-2006. Pet. 28, 31-32, 33, 58; Ex. 1003 ¶¶ 80-83, 89- 91; see also Pet. 43 n.4. Patent Owner’s Arguments Patent Owner disputes Petitioner’s contentions. Patent Owner argues that Petitioner’s proposed combination is rooted in hindsight and results in a more complicated system. PO Resp. 59, 60. Specifically, Patent Owner characterizes Petitioner’s combination as (1) eliminat[ing] Aizawa’s existing transmitter so the resulting device will not require “a separate RF circuit”; (2) chang[ing] Aizawa’s structure to add a second LED to transmit data using a base station, which would also require that a user remove the sensor before any data transfer can occur and thus eliminate the IPR2020-01539 Patent 10,588,554 B2 66 ability to display data in real-time; and then (3) add[ing] back in a separate communications circuit to the base station based on Mendelson 2006 so that the base station can send data to a PDA with a touch screen display. Id. at 59-60 (citing, e.g., Ex. 2004 ¶ 114), 61. Patent Owner further argues that such a modification eliminates the desired real-time monitoring employed by Mendelson-2006. Id. (citing Ex. 2004 ¶¶ 113, 115).9 Analysis As discussed above in Section II.D.5.iii, we determine that Petitioner demonstrated sufficiently that a person of ordinary skill in the art would have found it obvious to modify Aizawa to include an additional LED to, inter alia, allow for wireless transmission of sensed pulse rate and motion data to a base device. We further noted that although Aizawa discloses transmission of data for display (Ex. 1006 ¶¶ 15, 35), Aizawa is silent as to how the data is transmitted or displayed. In light of the combination with Inokawa, therefore, Aizawa’s multiple LEDs would have allowed wireless transmission of data to a base device. See, e.g., Ex. 1008 ¶ 76 (“[V]ital sign information stored in the memory 63, such as pulse and body motion, is transmitted to the base device 17 using the S-side infrared LED 23 of the pulse sensor 1 and the B-side PD 45 of the base device 17.”). Inokawa further discloses that the base device, once it receives information from the sensor, “transmits this information to the PC 59.” Id. ¶ 75; see also id. ¶¶ 67, 77. As described by Dr. Kenny, “the physiological 9 We do not address Patent Owner’s argument that Mendelson-2006 does not disclose a “multi-emitter/multi-detector sensor” because Mendelson-2006 is not relied upon for such limitations. See supra § II.D.iii, iv. IPR2020-01539 Patent 10,588,554 B2 67 sensor device’s sensor component transmits physiological measurement data to an included base station via an optical communications interface, and the physiological sensor device’s base station transmits signals responsive to a physiological parameter to a computer, via a network interface.” Ex. 1003 ¶ 81. With this backdrop, we are persuaded by Petitioner’s contention that it would have been obvious to implement the physiological sensor device resulting from the combined teachings of Aizawa, Inokawa, and Ohsaki as part of a physiological measurement system that includes a handheld computing device. Indeed, Aizawa and Inokawa already teach the desirability of transmitting sensed data to, e.g., a computer or a display, although neither discloses further detail. See, e.g., Ex. 1006, 15; Ex. 1008 ¶ 75; see also Ex. 1047 ¶ 56 (Aizawa is silent). In light of these teachings, we credit Dr. Kenny’s testimony that transmitting sensed data wirelessly to a handheld computing device, as taught by Mendelson-2006, would have achieved the identified benefits of, e.g., providing a low-cost display with a simple user interface and easy activation of functions (Ex. 1003 ¶ 86) and the ability to provide more effective medical care when the handheld device is carried by first responders (id. ¶ 90). See, e.g., id. ¶¶ 82-87, 91-93; Ex. 1047 ¶¶ 59-60. We are also persuaded that this would have been within the skill level of an ordinary artisan and would have achieved predictable results. Id. ¶ 92. We do not agree with Patent Owner’s characterization of the proposed combination. Petitioner does not propose “(1) eliminat[ing] Aizawa’s existing transmitter . . . (2) chang[ing] Aizawa’s structure to add a second LED to transmit data using a base station . . .; and then (3) add[ing] back in a IPR2020-01539 Patent 10,588,554 B2 68 separate communications circuit to the base station.” Contra PO Resp. 59- 60; Ex. 1047 ¶ 58. As discussed above, Petitioner proposes that the system suggested by, inter alia, Aizawa and Inokawa-which includes a sensor in communication with a base device, and which contemplates additional communication from the base device to a PC-further includes a handheld computing device in wireless communication with that system. In other words, Petitioner’s proposed combination effectively replaces or supplements Inokawa’s PC 59 with a PDA, such as that taught by Mendelson-2006. Thus, in Petitioner’s proposed combination, physiological data is sensed by Aizawa’s sensor, transmitted to a base device through an additional LED, as taught by Inokawa, and further transmitted to, inter alia, a PDA, as taught by Mendelson-2006. See, e.g., Pet. 27-34; see also id. at 43 n.4 (describing the proposed combination as, inter alia, adding “Inokawa’s base station to Aizawa’s physiological sensor device such that the sensor device includes a sensor and a base station with which the sensor communicates and through which the sensor communicates with a handheld device”). Indeed, both Aizawa and Inokawa expressly contemplate transmission to an additional computing device (see, e.g., Ex. 1006, 15; Ex. 1008 ¶ 75); Petitioner’s proposed modification merely states that such transmission occurs wirelessly to a handheld device. The record supports this contention. We have considered Dr. Madisetti’s testimony, but it is based on the same mischaracterization put forth by Patent Owner. Ex. 2004 ¶¶ 112 (mischaracterizing the combination), 114 (same). Notwithstanding this misrepresentation of the proposed modification, Dr. Madisetti does not dispute Dr. Kenny’s testimony that wireless transmission to a handheld IPR2020-01539 Patent 10,588,554 B2 69 computing device would have achieved the identified benefits, such as a low-cost device that improves medical care. See id. ¶¶ 112-118. As such, we credit Dr. Kenny’s unrebutted testimony. Patent Owner and Dr. Madisetti further criticize the combination, asserting that Mendelson-2006’s wireless transmission exists to allow real- time monitoring, which is impossible where a sensor must be mounted on a base device to transfer information through LEDs. Id. ¶ 113; see also PO Resp. 59. However, as discussed in Section II.D.iii above, the lack of real- time measurement and transmission is simply one consideration among many. As noted in Inokawa, real-time wireless communication has its drawbacks. Ex. 1008 ¶ 5. We discern that a skilled artisan would have weighed these competing interests. “[A] given course of action often has simultaneous advantages and disadvantages, and this does not necessarily obviate motivation to combine.” Medichem, 437 F.3d at 1165 (citation omitted). viii. “[i] one or more processors configured to wirelessly receive one or more signals from the physiological sensor device, the one or more signals responsive to at least a physiological parameter of the user” The cited evidence supports Petitioner’s contention that Mendelson- 2006 describes wirelessly transmitting vital physiological information acquired from a sensor to a PDA, which receives it. Pet. 59-61; see, e.g., Ex. 1010, 1, 2 (“The PDA can monitor multiple wearable pulse oximeters simultaneously and allows medics to collect vital physiological information to enhance their ability to extend more effective care to those with the most urgent needs.”), 3 (explaining that the PDA “has sufficient computational resources for the intended application” and “can also serve to temporarily store vital medical information received from the wearable unit”), 3 (“The IPR2020-01539 Patent 10,588,554 B2 70 [PDA’s graphical user interface] also displays the subject’s vital signs, activity level, body orientation, and a scrollable PPG waveform that is transmitted by the wearable device.”), Fig. 3 (displaying SpO2 and HR data); Ex. 1003 ¶¶ 179-180. As discussed above, Petitioner’s proposed combination involves transmission of sensed data from Aizawa’s physiological sensor to a base device, as taught by Inokawa, and further wireless transmission of that data from the base device to a handheld computing device, such as a PDA. See supra §§ II.D.5.iii (transmission to base device accomplished with an additional LED, as taught by Inokawa), II.D.5.vii (further transmission from base device to, e.g., a PC and/or PDA, as taught by Mendelson-2006, and contemplated by Aizawa and Inokawa). In light of these teachings, we are persuaded by Petitioner’s contention that a person of ordinary skill in the art “would have found it obvious to configure a processor of the PDA to wirelessly receive signals from the physiological sensor device” taught by the combination of, inter alia, Aizawa, Inokawa, and Mendelson-2006, wherein “the signals [are] responsive to physiological parameters of the user.” Pet. 60-61; see, e.g., Ex. 1003 ¶ 185. Petitioner’s stated reasoning for the proposed modification is sufficiently supported, including by the unrebutted testimony of Dr. Kenny. Ex. 1003 ¶¶ 181-186. Patent Owner does not present any argument against this limitation, apart from the arguments already addressed in Section II.D.5.vii. IPR2020-01539 Patent 10,588,554 B2 71 ix. “[j]-[l] a touch-screen display configured to provide a user interface, wherein: the user interface is configured to display indicia responsive to measurements of the physiological parameter, and an orientation of the user interface is configurable responsive to a user input” The cited evidence supports Petitioner’s contention that Mendelson- 2006 describes a PDA with a touchscreen display configured to display indicia responsive to measurements of, e.g., SpO2 and HR. Pet. 61-62; see, e.g., Ex. 1010, 3 (“The use of a PDA . . . also provides a low-cost touch screen interface.”). Petitioner acknowledges that “Mendelson-2006 does not explicitly state that an orientation of the GUI provided by the PDA is configurable responsive to a user input.” Pet. 63. However, Petitioner contends that a person of ordinary skill in the art would have understood that “the LabVIEW software that was used ‘to control all interactions between the PDA and the wearable unit via [t]he graphical user interface’ included the option to configure an orientation of a user interface,” e.g., by setting the report orientation to portrait or landscape view. Id. (alteration in original); see, e.g., Ex. 1003 ¶¶ 191-192; Ex. 1027, 186 (“Set the report orientation- portrait or landscape.”). Petitioner further contends that, in light of these teachings, a person of ordinary skill in the art “would have found it obvious to make an orientation of the PDA’s user interface configurable responsive to a user input, for the sake of user convenience.” Pet. 64; see, e.g., Ex. 1003 ¶ 193. Petitioner’s stated reasoning for the proposed modification is sufficiently supported, including by the unrebutted testimony of Dr. Kenny. See, e.g., Ex. 1003 ¶¶ 187-195. Patent Owner does not present any IPR2020-01539 Patent 10,588,554 B2 72 argument against this limitation, apart from the arguments already addressed in Section II.D.5.vii. x. “[m] a storage device configured to at least temporarily store at least the measurements of the physiological parameter.” The cited evidence supports Petitioner’s contention that Mendelson- 2006 teaches that the PDA is configured to store vital medical information received from the wearable pulse oximeter, and that an ordinarily skilled artisan “would have understood that the vital medical information would have included measurements of the physiological parameters obtained by the physiological sensor device (e.g., SpO2 and HR).” Pet. 65; Ex. 1010, 3 (“The PDA can also serve to temporarily store vital medical information received from the wearable unit.”); Ex. 1003 ¶ 198. Thus, Petitioner contends that a person of ordinary skill in the art “would have configured a storage device of the PDA to at least temporarily store measurements of physiological parameters (e.g., SpO2 and HR).” Pet. 64-65; see, e.g., Ex. 1003 ¶ 197. Petitioner’s stated reasoning for the proposed modification is sufficiently supported, including by the unrebutted testimony of Dr. Kenny. See, e.g., Ex. 1003 ¶¶ 196-199. Patent Owner does not present any argument against this limitation, apart from the arguments already addressed in Section II.D.5.vii. xi. Reasonable Expectation of Success Patent Owner argues that Petitioner has failed to demonstrate a reasonable expectation of success because Dr. Kenny did not perform a design analysis to create a functional sensor. PO Resp. 62-63. We disagree. As discussed in detail above, each of Petitioner’s proposed modifications to IPR2020-01539 Patent 10,588,554 B2 73 Aizawa-whether to include a second emitter, as taught by Inokawa; or to include a cover with a convex surface, as taught by Ohsaki; or to communicate with a handheld computing device, as taught by Mendelson- 2006-is rooted in explicit teachings of the prior art, and is supported by persuasive declarant testimony. We credit Dr. Kenny’s testimony that, for each proposed modification, the combined prior art teachings would have been applied as known, to achieve predictable results. See, e.g., Ex. 1003 ¶¶ 78 (applying Inokawa’s teachings would have been “nothing more than improving Aizawa’s pulse wave sensor that uses a single LED with the use of a known technique disclosed by Inokawa to detect and record body motion in addition to blood flow”), 160 (applying Ohsaki’s teachings would have been “nothing more than adjusting transparent plate 6’s shape to include a convex protrusion/lens surface similar to that disclosed by Ohsaki, and one of ordinary skill would have understood that this adjustment would improve adhesion to the user’s skin and reduce variation in the signals detected by the sensor”), 178 (“applying Mendelson[-]2006’s teachings . . . would have led to predictable results without altering or hindering the functions performed by the physiological sensor device. In fact, one of ordinary skill would have been motivated to implement the well-known technique of wirelessly transmitting data . . . to a handheld computing device”). For similar reasons discussed above with respect to each proposed modification, we conclude that that a skilled artisan would have had a reasonable expectation of success. See supra § II.D.5.iii, vi, vii-x; see also Ex. 1003 ¶¶ 71-199. IPR2020-01539 Patent 10,588,554 B2 74 xii. Summary For the foregoing reasons, we determine that Petitioner has met its burden of demonstrating by a preponderance of the evidence that claim 1 would have been obvious over the cited combination of references. Independent Claim 20 Independent claim 20 consists of limitations that are substantially similar to elements [a]-[h] of claim 1. Compare Ex. 1001, 44:51-45:21, with id. at 46:31-52 (reciting that the “convex surface,” as opposed to “the cover,” is “sufficiently rigid”; omitting details of the “handheld computing device”). In asserting that claim 20 also would have been obvious over the combined teachings of Aizawa, Inokawa, Ohsaki, and Mendelson-2006, Petitioner refers to the same arguments presented as to claim 1. See Pet. 76- 78. Patent Owner relies on the same arguments discussed above regarding claim 1. PO Resp. 11-63. For the same reasons discussed above, we determine that Petitioner has met its burden of demonstrating by a preponderance of the evidence that claim 20 would have been obvious over the cited combination of references. See supra § II.D.5. Dependent Claim 28 Dependent claim 28 ultimately depends from independent claim 20 and further recites “the single protruding convex surface protrudes a height greater than 2 millimeters and less than 3 millimeters.” Ex. 1001, 48:16-18. Petitioner reiterates that the sensor rendered obvious by the combined teachings of Aizawa, Inokawa, Ohsaki, and Mendelson-2006 would have included a cover with a single protruding convex surface, see supra § II.D.5.vi, and further contends that a person of ordinary skill in the art IPR2020-01539 Patent 10,588,554 B2 75 “would have found it obvious that a device designed to fit on a user’s wrist would be on the order of millimeters,” consistent with Ohsaki’s disclosure that the device is in “intimate contact” with the user’s skin. Pet. 85-86 (citing, e.g., Ex. 1003 ¶ 275). Petitioner also contends that an ordinarily skilled artisan would have taken user comfort into account when establishing the dimensions of the device’s convex cover. Id. at 86-87. With these considerations in mind, Petitioner contends that, “in order to provide a comfortable cover featuring a protruding convex surface that prevents slippage, the surface should protrude a height greater than 2 millimeters and less than 3 millimeters,” because “there would have been a finite range of possible protruding heights, and it would have been obvious to select a protruding height that would have been comfortable to the user.” Id. (citing, e.g., Ex. 1003 ¶¶ 273-277). Patent Owner argues that none of the cited references disclose the claimed height range and that Petitioner relies on hindsight reconstruction. PO Resp. 64-66 (citing, e.g., Ex. 2004 ¶¶ 121-124). Patent Owner also characterizes Dr. Kenny’s testimony as conclusory and unsupported. Id. at 66-67. Petitioner is correct that, “[w]hen there are a finite number of identified, predictable solutions, a person of ordinary skill in the art has good reason to pursue the known options within his or her technical grasp. If this leads to the anticipated success, it is likely the product . . . of ordinary skill and common sense.” KSR, 550 U.S. at 398. Petitioner has shown sufficiently that only a finite number of solutions existed with respect to the height of a convex protrusion on a tissue-facing sensor, which would have met the art-recognized goals of both (1) intimate contact between the IPR2020-01539 Patent 10,588,554 B2 76 sensor’s surface and the user and (2) user comfort. See, e.g., Ex. 1009 ¶¶ 6, 25. Bearing in mind these considerations, we credit Dr. Kenny’s testimony that it would have been obvious, “in order to provide a comfortable cover featuring a protruding convex surface that prevents slippage, [that] the surface should protrude a height greater than 2 millimeters and less than 3 millimeters.” Ex. 1003 ¶ 276. We have considered Patent Owner’s argument, and Dr. Madisetti’s cited testimony. However, it is not dispositive that none of Aizawa, Inokawa, Ohsaki, or Mendelson-2006 teach the claimed range. PO Resp. 64-66; Ex. 2004 ¶¶ 105-107. Petitioner relies upon the knowledge, ability, and creativity of a person of ordinary skill in the art, not the teachings of a specific reference. Notably, Dr. Madisetti does not dispute Dr. Kenny’s position that there were a finite number of options available for the height of the convex surface. Ex. 2004 ¶¶ 121-124. Therefore, we do not agree that Petitioner’s contentions are rooted in impermissible hindsight. See, e.g., In re McLaughlin, 443 F.2d 1392, 1395 (CCPA 1971) (“Any judgment on obviousness is in a sense necessarily a reconstruction based upon hindsight reasoning, but so long as it takes into account only knowledge which was within the level of ordinary skill at the time the claimed invention was made and does not include knowledge gleaned only from applicant’s disclosure, such a reconstruction is proper.”). Accordingly, for the foregoing reasons, we determine that Petitioner has met its burden of demonstrating by a preponderance of the evidence that claim 28 would have been obvious over the cited combination of references. IPR2020-01539 Patent 10,588,554 B2 77 Dependent Claims 2-7 and 21-27 Petitioner also contends that claims 2-7 and 21-27 would have been obvious based on the same combination of prior art addressed above. These challenged claims all depend directly or indirectly from independent claim 1 or 20. Petitioner identifies teachings in the prior art references that teach or suggest the limitations of these claims, and provides persuasive reasoning as to why the claimed subject matter would have been obvious to one of ordinary skill in the art. Pet. 65-75, 78-87. Petitioner also supports its contentions for these claims with the testimony of Dr. Kenny. Ex. 1003 ¶¶ 200-230, 245-277. Patent Owner does not present any arguments for these claims other than those we have already considered with respect to independent claims 1 and 20. PO Resp. 63-64 (“[T]he Petition fails to establish that independent claims 1 and 20 are obvious in view of the cited references of Ground 1 and therefore fails to establish obviousness of any of the challenged dependent claims.”); see supra § II.D.5. We have considered the evidence and arguments of record and determine that Petitioner has demonstrated by a preponderance of the evidence that claims 2-7 and 21-27 would have been obvious over the combined teachings of Aizawa, Inokawa, Ohsaki, and Mendelson-2006, for the reasons discussed in the Petition and as supported by the testimony of Dr. Kenny. Conclusion For the foregoing reasons, we determine that Petitioner has met its burden of demonstrating by a preponderance of the evidence that claims 1-7 IPR2020-01539 Patent 10,588,554 B2 78 and 20-28 would have been obvious over the cited combination of references. E. Obviousness over the Combined Teachings of Aizawa, Inokawa, Ohsaki, Mendelson-2006, and Bergey Petitioner contends that claims 8-19 of the ’554 patent would have been obvious over the combined teachings of Aizawa, Inokawa, Ohsaki, Mendelson-2006, and Bergey. Pet. 87-98; see also Pet. Reply 36-38. Patent Owner disagrees. PO Resp. 67-69; see also PO Sur-reply 27-29. Based on our review of the parties’ arguments and the cited evidence of record, we determine that Petitioner has met its burden of showing by a preponderance of the evidence that claims 8-12 and 14-19 are unpatentable, but has not met its burden with respect to claim 13. Overview of Bergey (Ex. 1016) Bergey is a U.S. patent titled “Solid State Watch with Magnetic Setting,” and discloses a watch in which the electronics are “hermetically sealed in the watch case to be free of dust and moisture.” Ex. 1016, code (57). Moreover, the electronic components are “resiliently mounted for improved shock resistance.” Id. Dependent Claims 8-10, 12, 14-16, 18, and 19 Petitioner contends that it would have been obvious to have modified the sensor of Aizawa-Inokawa-Ohsaki-Mendelson-2006 to hermetically seal the sensor components within the substrate, wall, and cover, so as to obtain advantages disclosed by Bergey, e.g., to protect the electronics and prevent condensation within the case. Pet. 88-89 (citing Ex. 1003 ¶¶ 278-281; Ex. 1016, code (57), 2:56-67, 8:48-9:34). IPR2020-01539 Patent 10,588,554 B2 79 Petitioner identifies teachings in the prior art references that teach or suggest the limitations of each of dependent claims 8-10, 12, 14-16, 18, and 19, and provides persuasive reasoning as to why the claimed subject matter would have been obvious to one of ordinary skill in the art. Pet. 89-97. Petitioner also supports its contentions for these claims with the testimony of Dr. Kenny. Ex. 1003 ¶¶ 282-291, 294-298, 302-312, 315-316. Patent Owner does not present any arguments for these claims other than those we have already considered with respect to independent claims 1 and 20. PO Resp. 63-64 (“[T]he Petition fails to establish that independent claims 1 and 20 are obvious in view of the cited references of Ground 1 and therefore fails to establish obviousness of any of the challenged dependent claims.”); see supra § II.D.5. We have considered the evidence and arguments of record and determine that Petitioner has demonstrated by a preponderance of the evidence that claims 8-10, 12, 14-16, 18, and 19 would have been obvious over the combined teachings of Aizawa, Inokawa, Ohsaki, Mendelson-2006, and Bergey for the reasons discussed in the Petition and as supported by the testimony of Dr. Kenny. Dependent Claims 11 and 17 Dependent claim 11 ultimately depends from independent claim 1 and further recites “the single protruding convex surface protrudes a height between 1 millimeter and 3 millimeters.” Ex. 1001, 45:65-67. Dependent claim 17 also ultimately depends from independent claim 1 and further recites “the single protruding convex surface protrudes a height greater than 2 millimeters and less than 3 millimeters.” Id. at 46:23-25. IPR2020-01539 Patent 10,588,554 B2 80 With respect to claims 11 and 17, both Petitioner and Patent Owner refer to their positions advanced with respect to claim 28, addressed in Section II.D.7 above. Pet. 91 (citing, e.g., Ex. 1003 ¶¶ 61-102, 273-277, 278-293), 96 (citing, e.g., Ex. 1003 ¶¶ 61-199, 273-277, 292-293, 313- 314); PO Resp. 67-68, 69 (citing Ex. 2004 ¶¶ 126-129). For the same reasons discussed above in Section II.D.7, we determine that Petitioner has met its burden of demonstrating by a preponderance of the evidence that claims 11 and 17 would have been obvious over the cited combination of references. Dependent Claim 13 Dependent claim 13 ultimately depends from independent claim 1 and further recites “the displayed indicia are further responsive to temperature.” Ex. 1001, 46:4-6; see also id. 45:14-16 (claim 1 reciting, “the user interface is configured to display indicia responsive to measurements of the physiological parameter”). Petitioner contends that, as discussed above in Section II.D.5, the combined sensor “would have included a touch-screen display that provides a GUI including display indicia that are response [sic] to a user input.” Pet. 92. Petitioner further contends, Consistent with Mendelson-2006’s description of the PDA’s “simple GUI” being configured to allow for “easy activation of various functions,” a [person of ordinary skill in the art] would have found it obvious to make an orientation of the PDA’s user interface configurable responsive to temperature, for the sake of user convenience. A POSITA would have understood that a display indicia that changes when the temperature reaches a threshold value is responsive to temperature. Thus, a POSITA would have found it obvious that a system resulting from the combination of Aizawa, Inokawa, Ohsaki, Mendelson-2006, and IPR2020-01539 Patent 10,588,554 B2 81 Bergey would have allowed an orientation of the user interface to be further responsive to temperature. Id. at 93 (citing Ex. 1003 ¶¶ 299-301; Ex. 1010, 3; Ex. 1020, 2, 6). Patent Owner disputes Petitioner’s contention, arguing that “Petitioner fails to identify any mention of temperature in Ohsaki, Aizawa, Inokawa, Mendelson 2006, or Bergey, let alone display indicia responsive to temperature.” PO Resp. 68. Patent Owner also argues that the cited portion of Exhibit 1020 likewise fails to mention temperature, and while an uncited portion mentions “operating temperature” and “storage temperature,” they are not relied upon or explained. Id. On Reply, Petitioner contends that Dr. Kenny’s cited testimony described the “LabView program, which includes a user interface configured to indicate when a temperature exceeds a threshold. . . . Thus, the Petition provides clear evidence that it was well known to display indicia that are responsive to temperature, and relates the same to Mendelson-2006’s description of the PDA’s user interface.” Pet. Reply 37-38 (citing, inter alia, Ex. 1003 ¶ 300; Ex. 1027, 46; Ex. 1047 ¶¶ 66-67). In its Sur-Reply, Patent Owner argues that reliance on the LabView manual provided as Exhibit 1027 is untimely, as it was not cited in the Petition or in Dr. Kenny’s Declaration. PO Sur-reply 28-29. Moreover, Patent Owner argues that Petitioner still fails to explain why a person of ordinary skill in the art “would have added indicia responsive to temperature to Aizawa’s pulse wave sensor based on LabVIEW’s discussion of ‘monitoring the temperature of an experiment.’” Id. at 29 (citing Ex. 1027, 46). Further, Patent Owner argues that “Petitioner abandons its original argument” regarding the “orientation of the user IPR2020-01539 Patent 10,588,554 B2 82 interface” being “responsive to temperature,” and now on Reply argues that it is the “display indicia that are responsive to temperature.” Id. (citing Pet. 93; Reply 37-38). We agree with Patent Owner. The Petition contends it would have been obvious “to make an orientation of the PDA’s user interface configurable responsive to temperature, for the sake of user convenience.” However, none of the cited evidence supports this contention. Pet. 93; Ex. 1003 ¶¶ 299-301; Ex. 1010, 3; Ex. 1020, 2, 6. Specifically, Dr. Kenny’s declaration discusses the LabVIEW program and contends that it “included features that indicated ‘when the temperature goes above a certain level,’” and concluded that “[o]ne of ordinary skill would have understood that a display indicia that changes when the temperature reaches a threshold value is responsive to temperature.” Ex. 1003 ¶ 300. However, while the citation to Mendelson- 2006 (Ex. 1010, 3) references LabVIEW, it does not discuss temperature, or display indicia responsive to temperature. Likewise, the cited portions of the HP iPAQ Specifications (Ex. 1020, 2, 6) also fail to discuss temperature or display indicia responsive to temperature. As such, Dr. Kenny’s initial testimony lacks any supporting evidence. On Reply, Petitioner cites a LabVIEW User Manual (Ex. 1027), which states that a warning light may activate to indicate when the temperature of an experiment exceeds a certain level. Id. at 46. We agree with Patent Owner that this evidence should have been cited in the Petition. As discussed above, the Petition lacks any supporting evidence to support the contention of obviousness. Petitioner cannot provide such evidence on Reply when it should have been offered in the Petition. IPR2020-01539 Patent 10,588,554 B2 83 Conclusion For the foregoing reasons, we determine that Petitioner has met its burden of demonstrating by a preponderance of the evidence that claims 8- 12 and 14-19 would have been obvious over the cited combination of references, but has not met its burden with respect to claim 13. IPR2020-01539 Patent 10,588,554 B2 84 III. CONCLUSION In summary:10 Claims 35 U.S.C. § Reference(s)/ Basis Claims Shown Unpatentable Claims Not Shown Unpatentable 1-7, 20-28 103 Aizawa, Inokawa, Ohsaki, Mendelson- 2006 1-7, 20-28 8-19 103 Aizawa, Inokawa, Ohsaki, Mendelson- 2006, Bergey 8-12, 14-19 13 Overall Outcome 1-12, 14-28 13 IV. ORDER Upon consideration of the record before us, it is: ORDERED that claims 1-12 and 14-28 of the ’554 patent have been shown to be unpatentable; 10 Should Patent Owner wish to pursue amendment of the challenged claims in a reissue or reexamination proceeding subsequent to the issuance of this decision, we draw Patent Owner’s attention to the April 2019 Notice Regarding Options for Amendments by Patent Owner Through Reissue or Reexamination During a Pending AIA Trial Proceeding. See 84 Fed. Reg. 16654 (Apr. 22, 2019). If Patent Owner chooses to file a reissue application or a request for reexamination of the challenged patent, we remind Patent Owner of its continuing obligation to notify the Board of any such related matters in updated mandatory notices. See 37 C.F.R. § 42.8(a)(3), (b)(2). IPR2020-01539 Patent 10,588,554 B2 85 FURTHER ORDERED that claim 13 of the ’554 patent has not been shown to be unpatentable; FURTHER ORDERED that, because this is a final written decision, parties to the proceeding seeking judicial review of the decision must comply with the notice and service requirements of 37 C.F.R. § 90.2. IPR2020-01539 Patent 10,588,554 B2 86 PETITIONER: Walter Renner Andrew Patrick FISH & RICHARDSON P.C. Axf-ptab@fr.com patrick@fr.com PATENT OWNER: Joseph Re Stephen Larson Jarom Kesler Jacob Peterson KNOBBE, MARTENS, OLSON, & BEAR LLP 2jrr@knobbe.com 2swl@knobbe.com 2jzk@knobbe.com jup@knobbe.com Copy with citationCopy as parenthetical citation