Ex Parte Sacherer et alDownload PDFPatent Trial and Appeal BoardJul 26, 201612751178 (P.T.A.B. Jul. 26, 2016) Copy Citation UNITED STA TES p A TENT AND TRADEMARK OFFICE APPLICATION NO. FILING DATE FIRST NAMED INVENTOR 121751,178 03/31/2010 Klaus-Dieter Sacherer 41577 7590 07/28/2016 Woodard, Emhardt, Moriarty, McNett & Henry LLP Roche Diagnostics Operations, Inc. (PD) 111 Monument Circle, Suite 3700 Indianapolis, IN 46204-5137 UNITED STATES DEPARTMENT OF COMMERCE United States Patent and Trademark Office Address: COMMISSIONER FOR PATENTS P.O. Box 1450 Alexandria, Virginia 22313-1450 www .uspto.gov ATTORNEY DOCKET NO. CONFIRMATION NO. 007804-000005 21582us1 1175 EXAMINER STOUT, MICHAEL C ART UNIT PAPER NUMBER 3736 NOTIFICATION DATE DELIVERY MODE 07/28/2016 ELECTRONIC Please find below and/or attached an Office communication concerning this application or proceeding. The time period for reply, if any, is set in the attached communication. Notice of the Office communication was sent electronically on above-indicated "Notification Date" to the following e-mail address( es): DocketDept@uspatent.com pair_roche@firsttofile.com meg.ward@contractors.roche.com PTOL-90A (Rev. 04/07) UNITED STATES PATENT AND TRADEMARK OFFICE BEFORE THE PATENT TRIAL AND APPEAL BOARD Ex parte KLAUS-DIETER SACHERER, RONALD MONCH, MICHAEL SCHABBACH, and JORG SCHERER1 Appeal2013-009264 Application 12/751,178 Technology Center 3700 Before DONALD E. ADAMS, JEFFREY N. FREDMAN, and JACQUELINE T. HARLOW, Administrative Patent Judges. PER CURIAM DECISION ON APPEAL This is an appeal under 35 U.S.C. § 134(a) involving claims to systems and methods for receiving a body fluid for analysis. The claims are rejected as failing to comply with the written description requirement, indefinite, and obvious. We have jurisdiction under 35 U.S.C. § 6(b). We AFFIRM-IN-PART. 1 According to Appellants, the Real Party in Interest is Roche Diagnostics Operations, Inc. (App. Br. 3). Appeal2013-009264 Application 12/751,178 STATEMENT OF THE CASE The Specification describes a system and method for receiving a body fluid for analysis, in which a sample receiving unit is coupled to a drive unit for a forwards and backwards movement between a guide chamber and a receiving site (Spec. 1:11; 2:7-8). Claims 1-9, 21-27, 29, 30, 34--36, and 38--43 are on appeal. Claim 1 is illustrative and reads as follows: 1. Device for receiving a body fluid for analytical purposes comprising a container and at least one sample receiving unit- preferably for a single-use test that can be pushed out of a guide chamber of the container by means of a drive unit and to which the body fluid can be applied at a receiving site, wherein the container is designed as a magazine for storing a plurality of sample receiving units and a coupling device to couple the sample receiving unit to the drive unit for a forwards and backwards movement between the guide chamber and the receiving site. The claims stand rejected as follows: I. Claims 21-27, 29, 30, 34--36, and 41--43 stand rejected under 35 U.S.C. § 112, first paragraph, for failing to comply with the written description requirement. II. Claims 1-9, 38, and 39 stand rejected under 35 U.S.C. § 112, second paragraph, as indefinite. III. Claims 1-9 stand rejected under 35 U.S.C. § 103(a) as obvious based on Fritz2 and Mauze. 3 IV. Claim 1 stands rejected under 35 U.S.C. § 103(a) as obvious based on Perez4 and Fritz. 2 Fritz et al., WO 2002/36010 Al, published May 10, 2002. 3 Mauze et al., US 6,375,627 Bl, issued Apr. 23, 2002. 4 Perez et al., US 2002/0188223 Al, published Dec. 12, 2002. 2 Appeal2013-009264 Application 12/751,178 V. Claims 1, 21, 22, 24, 25, 29, 30, 34, 36, 38, 39, and 40-43 stand rejected under 35 U.S.C. § 103(a) as obvious based on DeNuzzio5 and Fritz. VI. Claim 23 stands rejected under 35 U.S.C. § 103(a) as obvious based on DeNuzzio, Fritz, and Kloepfer. 6 VII. Claim 26 stands rejected under 35 U.S.C. § 103(a) as obvious based on DeNuzzio, Fritz, and Sacherer.7 VIII. Claims 27 and 35 stand rejected under 35 U.S.C. § 103(a) as obvious based on DeNuzzio, Fritz, and Bonner. 8 I. The Examiner has rejected claims 21-27, 29, 30, 34--36, and 41--43 under 35 U.S.C. § 112, first paragraph, for failing to comply with the written description requirement. In response, Appellants submit that amendments, which "were believed to place the application in better condition for appeal," were filed on September 24, 2012 (App. Br. 11). Appellants acknowledge, however, that "the amendments were not entered in the October 19, 2012 Advisory Action" (id.). The Examiner explains that "the amendments required further consideration and/ or search," and, therefore, were not entered (Ans. 24). Accordingly, because Appellants do not contest the merits of this rejection, we summarily affirm. II. The Examiner has rejected claims 1-9, 38, and 39 under 35 U.S.C. § 112, second paragraph, as indefinite. As with the written description 5 DeNuzzio et al., US 7,959,583 B2, issued June 14, 2011. 6 Kloepfer et al., US 6,840,912 B2, issued Jan. 11, 2005. 7 Sacherer, US 6,497 ,845 B 1, issued Dec. 24, 2002. 8 Bonner et al., US 5,510,266, issued Apr. 23, 1996. 3 Appeal2013-009264 Application 12/751,178 rejection described above, Appellants' filed amendments to remedy this rejection on September 24, 2012, however, those amendments were not entered into record because they required further consideration and/ or search (see App. Br. 11; Ans. 24). Because Appellants do not contest the merits of this rejection, we summarily affirm. III. The Examiner has rejected claims 1-9 under 35 U.S.C. § 103(a) as obvious based on Fritz and Mauze. Because the claims are not separately argued, we focus our discussion on independent claim 1, which is representative. Findings of Fact FF 1. Fritz9 discloses a system for withdrawing body fluids [that] includes a drive unit having a plunger which is moved from a resting position into a lancing position in order to carry out a lancing process and a lancing unit containing a lancet with a needle. The plunger and lancet are coupled together by a form fit in order to carry out a lancing process. (Fritz Abstract; see also Ans. 6-7, 11-16, 18-19.) FF 2. Fritz discloses that the system, "on the one hand, avoids contamination or infection by used lancets and, on the other hand, allows a substantially pain-reduced lancing for the user," that the system "operates with lancets in magazines and allows a user to change a lancet that has not yet been used without having to carry out complicated handling steps," and 9 The Examiner refers to the English equivalent of Fritz, US 2004/003438 Al, published Feb. 19, 2004 (Ans. 6). For purposes of this Appeal, we will refer to this English equivalent. 4 Appeal2013-009264 Application I2/75I,I 78 that the lancing unit is designed "so that the shock caused by the impact has no influence on the lancing process in the tissue which thus avoids incision pain caused by such a vibration" (Fritz i-fi-16, 42; see also Ans. 6, I I-I6, I8- I9, 27-28). FF 3. Fritz depicts in Figures I A-IC cross-sectional views of a lancing unit having a holding device on the lancet at three different positions (Fritz i124). Figures IA-IC are reproduced below. Fig.1 (Id. at Figs. IA-IC.) Figures IA-IC show "how the form-fitting coupling between the driving plunger 10 and the lancet 30 occurs during the actual lancing process," that "as the plunger 10 penetrates the lancet 30 further it pushes the lancet 30 within the sleeve 40 towards the exit opening 41 such that finally the tip of the needle 31 protrudes beyond the exit opening 41 and pierces a tissue lying underneath," and "that a form-fitting connection is formed which not only enables a forward movement of the lancet 30 in order 5 Appeal2013-009264 Application 12/751,178 to carry out an incision, but also an active retraction of the lancet 30 controlled by the drive unit which is essentially without play" (id. at i-fi-138, 39; see also Ans. 6-7, 11-15, 18-19). FF 4. Fritz depicts in Figures 1 OA-1 OB "cross-sectional views of a barrel-shaped magazine" (Fritz i136). Figure lOA-lOB are reproduced below. A B Fig.10 (Id. at Figs. lOA-lOB.) Figures lOA and lOB show a cylindrical magazine 500 that "enables new lancets 502 to be coupled to the drive 503" (id. at i160; see also Ans. 6-7, 11-16, 18-19). As depicted in Figures lOA and lOB, driving plunger 504 can "be fixed relative to a lancet 502 and the barrel-shaped magazine 500" and "rotated like a revolver barrel such that unused lancets 502 are moved into the position for coupling to the driving plunger 504" (id. at i160; see also Ans. 6-7, 11-16, 18-19). FF 5. Mauze discloses [a] device for sampling and analyzing a physiological fluid from the skin of a patient by puncture. The device includes a body 6 Appeal2013-009264 Application 12/751,178 [that] has a needle with a point for puncturing a physiological tissue and a channel in the body for conducting the physiological fluid away from the point. . . . The device can be used to lance the skin and obtain a representative sample of the physiological fluid[]. (Mauze Abstract; see also Ans. 7.) FF 6. Mauze discloses that "in the finger-stick methods, the process of milking and drawing blood sample to the surface leads to changes in the analyte concentration, ... due to exposure of the sample to the ambient environment (e.g., exposure to air)," and therefore, "the current methods of finger-stick sampling cannot be used to measure parameters such as blood gas contents" (Mauze 1 :35--42; see also Reply Br. 3). FF 7. Mauze discloses that [ s ]ince pain is related to the amount of blood extracted, venipuncture inflicts significant amount of unnecessary pain and trauma to the patient. Thus, there is a need for an apparatus and method for sampling and quickly analyzing blood from a patient with minimal pain using a small volume of sample blood. (Mauze 1 :46-51.) FF 8. The Specification teaches that "the sample receiving unit is formed by a hollow needle to suck in the sample fluid which is preferably at the same time designed as a guide for the lancing unit" (Spec. 5: 20-21 ). Analysis We have considered, but do not find persuasive Appellants' arguments that the Examiner erred in concluding that the claimed device for receiving a body fluid is obvious over the combination of Fritz and Mauze. We address Appellants' arguments below. 7 Appeal2013-009264 Application I2/75I,I 78 Appellants argue that "even together[,] the cited references do not disclose or even suggest a magazine system for sample receiving units that has a coupling device configured to facilitate forward and, more importantly, backward movement of the sample receiving unit" (App. Br. I3). Appellants assert that neither Fritz nor Mauze teaches a container that is "designed as a magazine for storing a plurality of sample receiving units and a coupling device to couple the sample receiving unit to the drive unit for a forwards and backwards movement between the guide chamber and the receiving site" (id.). Appellants also contend that "the Mauze reference never discloses any type of magazine let alone any type of coupling structure between the driver as well as other devices" (id. at I4). We are not persuaded. We agree with the Examiner that Fritz teaches a device for receiving a body fluid for analytical purposes [] comprising a container (cylindrical magazine 500[]) and at least one sample unit (lancet 502 having a needle 3 I) for a single-use test that can be pushed out of a guide chamber of the container by means of a drive unit (503) and to which the body fluid can be applied at a receiving site[], wherein the container[] is designed as a magazine for storing a plurality of sample units [] and a coupling device to couple the sample receiving unit to the drive unit [] for a forwards and backwards movement between the guide chamber and the receiving site[]. (Ans. 6-7 (citing Fritz i-fi-13, 60, Abstract, Figs. IA-IC, IO); see also FF I- 4.) We also agree with the Examiner that "Mauze teaches a needle device for sampling and analyzing fluid comprising a needle I 0 having a tip I 8 and a channel 20 for fluid to enter via capillary action" (Ans. 7 (citing Mauze 3:7-38), see also FF 5-7). As the Examiner explains, the Specification does not define "sample receiving unit" in such a way as to exclude the fluid collecting lancet of Mauze (Ans. 26; see also FF 8). 8 Appeal2013-009264 Application 12/751,178 Accordingly, we agree with the Examiner that the combination of Fritz and Mauze discloses each element of claim 1, and furthermore would yield the predictable result of "draw[ing] fluid to sensors for determining the parameters of the sampled physiological fluid" (Ans. 7 (citing Mauze 3:7- 38)). See In re Merck & Co., 800 F.2d 1091, 1097 (Fed. Cir. 1986) ("Non- obviousness cannot be established by attacking references individually where the rejection is based upon the teachings of a combination of references. . . . [The reference] must be read, not in isolation, but for what it fairly teaches in combination with the prior art as a whole."). Appellants contend that "Fritz focuses on the use of lancets" instead of a test strip, and that "[t]he operational needs for lancets are quite different from sample receiving units, like test strips" (App. Br. 13). Appellants assert that the lancet would need to be extended and retracted quickly to avoid significant pain while test strips just need to be extended (id. at 13- 14). Appellants further argue that "the approaches in Fritz and Mauze are incompatible to one another such that one skilled in the art would not have considered combining the references" (Reply Br. 3). In particular, Appellants contend that modifying Mauze' s sampling needle, which is silicon and brittle, to achieve the desired flexibility for the holding elements of Fritz, would require significant and dramatic modifications (id.). We do not find these arguments persuasive. As an initial matter, we observe that the claims subject to this rejection do not require that the sample receiving unit is a test strip (see also Ans. 26). The absence of such a requirement is consistent with the Specification, which teaches that the sample receiving unit can be a hollow needle designed as a guide for the lancing unit (FF 8). 9 Appeal2013-009264 Application 12/751,178 Appellants' arguments concerning avoidance of pain and the requirement for sampling needle flexibility are similarly unavailing. Claim 1 does not require any degree of pain avoidance or lancet flexibility, and Appellants have not provided evidentiary support for the assertion that the combined teachings of Fritz and Mauze would result in an inoperable device. See In re Pearson, 494 F.2d 1399, 1405 (CCPA 1974) ("Attorney's argument in a brief cannot take the place of evidence."). Furthermore, as the Examiner notes, "lancet systems which cause more pain than the system disclosed by Fritz are still suitable for the intended purpose of sampling and have been used in the past and are still used today, see background section of Fritz" (Ans. 27). In addition, we observe that Fritz and Mauze each teach techniques for reducing the pain associated with sample acquisition (FF 2, 7). Appellants also assert that Mauze teaches away from the finger-stick type devices of Fritz because these devices expose the body fluid to air (Reply Br. 3). We note, however, that the concerns expressed by Mauze regarding exposure of a sample to ambient air pertains only to those instances in which blood gas or a similar parameter is being measured (FF 6). Accordingly, because claim 1 does not require the measurement of blood gas or similar parameters, and because exposure to ambient air would not necessarily interfere with other blood measurements, we do not find Appellants' position persuasive. See DyStar Textilfarben GmbH & Co. Deutschland KG v. C.H. Patrick Co., 464 F.3d 1356, 1364 (Fed. Cir. 2006) ("[ w ]e will not read into a reference a teaching away from a process where no such language exists."). 10 Appeal2013-009264 Application 12/751,178 IV. The Examiner has rejected claim 1 under 35 U.S.C. § 103(a) as obvious based on Perez and Fritz. Findings of Fact FF 9. Perez discloses "incising, sampling and/or testing of the bodily fluid" (Perez Abstract; see also Ans. 10). FF 10. Perez depicts in Figure 16 "a side, elevational view, partially in cross-section" including movable, rigid constricting elements, "with the elements in the constricting position" (Perez i-fi-133, 34). Figure 16 is reproduced below. Fig .. 16 (Id. at Fig. 16.) Figure 16 shows housing 42, capillary 45, and lancet 46, in which "[t]he lancet 46 is extended beyond the capillary and incises the skin[], and the resulting fluid contacts and enters the capillary passageway" (id. i-fi-179, 82; see also Ans. 10-11). 11 Appeal2013-009264 Application 12/751,178 Analysis We have considered, but do not find persuasive Appellants' arguments that the Examiner erred in concluding that the claimed device for receiving a body fluid is obvious over Perez and Fritz. We address Appellants' arguments below. Similar to the arguments presented regarding the combination of Fritz and Mauze, Appellants contend that the combination of Perez and Fritz does not teach "a magazine system for sample receiving units that has a coupling device configured to facilitate forward and, more importantly, backward movement of the sample receiving unit" (App. Br. 15). Appellants assert that Perez does not teach retracting the capillary, and does not teach the capillary being integrated into a magazine-type device (id.; Reply Br. 3). We agree with the Examiner that Perez discloses "a device for receiving a body fluid for analytical purposes" having a container ( 42) and at least one sample receiving unit ( 45) preferably for a single-use test that can be pushed out of a guide chamber (space between the housing 42 and the capillary tube 45) of the container [] and to which the body fluid can be applied at a receiving site[]. (Ans. 10 (citing Perez i182, Abstract, Figs. 15-16); see also FF 9-10.) As previously discussed, we also agree with the Examiner that Fritz teaches a device for sampling body fluids comprising a container that is designed as a magazine for storing a plurality of sample units [] and a coupling device to couple the sample units to the drive unit [] for a forwards and backwards movement between the guide chamber and the receiving site[]. (Ans. 11 (citing Fritz Figs. lA---C, 10), see also FF 1--4). We further note that the Specification does not define "sample receiving unit" to exclude a 12 Appeal2013-009264 Application 12/751,178 capillary, as disclosed by Perez (see also Ans. 26; see also FF 8). Therefore, we conclude that the combination of Perez and Fritz satisfies the recited elements, and would yield predictable results of "provid[ing] a magazine which can store multiple sample receiving units thereby allowing the user to perform a plurality of test prior to reloading the device" (Ans. 11-12). See In re Merck & Co., 800 F.2d at 1097. Appellants also contend that the Examiner's rationale for combining Fritz and Perez to arrive at the claimed invention is improperly based on hindsight (Reply Br. 3). In particular, Appellants assert that "[a]t best, one skilled in the art would consider incorporating the drive 503 of Fritz to actuate the lancet in Perez. There is no hint or any suggestion in either reference or elsewhere of the need for retracting the capillary in Perez" (Reply Br. 3). While we are fully aware that hindsight bias often plagues determinations of obviousness, Graham v. John Deere Co., 383 U.S. 1, 36 (1966), we are also mindful that the Supreme Court has clearly stated that "combination of familiar elements according to known methods is likely to be obvious when it does no more than yield predictable results." KSR, 550 U.S. at 416. In this instance, we agree with the Examiner that it would have been obvious to a skilled artisan to combine the needle devices for sampling fluid taught by Fritz and Perez to arrive at the claimed invention (Ans. 11- 12), absent any impermissible hindsight. Fritz discloses a magazine for storing sample units, as well as a coupling device to couple the sample units to a drive unit to permit forward and backward movement between the guide chamber and the receiving site 13 Appeal2013-009264 Application 12/751,178 (FF 1--4), and Perez discloses a sample receiving unit (FF 9-10). Accordingly, we agree with the Examiner that it would have been obvious to a person having ordinary skill in the art at the time of the invention to modify the device taught by Perez to include a magazine for storing a plurality of sample units with a drive unit as taught by Fritz in order to provide a magazine which can store multiple sample receiving units thereby allowing the user to perform a plurality of test[ s] prior to reloading the device. (Ans. 11-12.) See also In re Merck & Co., 800 F.2d at 1097. V. The Examiner has rejected claims 1, 21, 22, 24, 25, 29, 30, 34, 36, 38, 39, and 40--43 under 35 U.S.C. § 103(a) as obvious based on DeNuzzio and Fritz. We focus our discussion on independent claims 1, 21, and 34. Findings of Fact FF 11. DeNuzzio discloses "[a] lancing device integrated with a sensing strip" in which "[a] lance and analyte sensing strip are combined in a single, preferably disposable, device" and "a slot is provided in a support member so [that] the lance can be engaged by a lance driver and driven into the skin producing a bodily fluid sample from a wound" (DeNuzzio Abstract). FF 12. DeNuzzio depicts in Figure 2a "a perspective view of a cartridge containing lance and test strip assemblies" (id. at 2: 19-20). Figure 2a is reproduced below. 14 Appeal2013-009264 Application 12/751,178 (Id. at Fig. 2a.) Figure 2a shows "a plurality of[] devices [that] are packaged in a cartridge 114 []ready for insertion into a handheld, portable glucose meter" in which "the cartridge 114 has an opening 116 adapted to allow access to the sensing[] surface [] of the bottom-most assembly 101 in the stack," that the "[ o ]pening 116 is also used to eject the bottom-most assembly 101 after use," and that "[t]he lance and test strip assemblies[] are preferably oriented within cartridge 114 such that slot 110 is aligned with a slot [] in the bottom of cartridge 114" (id. at 4:56-5: 1; see also Ans. 12-13). FF 13. DeNuzzio depicts in Figure 4a "the test strip [that] is brought into position for contact with the bodily fluid sample after the lancing action" (DeNuzzio 2:30-33). Figure 4a is reproduced below. 15 Appeal2013-009264 Application 12/751,178 JH : F!G.40 (Id. at Fig. 4a.) Figure 4a shows that a "device 300 automatically moves the sensing strip 304 into position after the lancet 306 is withdrawn from the skin," that "[a] lateral displacement of one to two millimeters is created by pressing a push button 312 on the device support structure," that "[t]he push button 312 is used to trigger the lancet 306 when pressed, and also to displace the strip 304 when released," that "the device 300[,] after the push button 312 has been pressed, releas[es] the lancet 306 into the skin 313," that "[a]s push button 312 is released, lancet 306 [is] withdrawn from the skin, and spring 314 causes sensing strip 304 to be displaced to the location of the wound 316," and that "[a] wick 318 is preferably included to aid bodily fluid flow to the sensing area by capillary action" (id. at 5:43-57; see also Ans. 12-13). Analysis We agree with Appellants that the Examiner erred in rejecting claims 1, 21, and 34 as being obvious over DeNuzzio and Fritz. Fundamental to the Examiner's rejection of each of claims 1, 21, and 34 is the conclusion that 16 Appeal2013-009264 Application 12/751,178 [i]t would have been obvious to a person of ordinary skill in the art at the time of the invention to modify ... DeNuzzio to include a driving means and coupling device as taught by Fritz in order to selective extend and retract a sampling device of a magazine. (Ans. 13; see also id. at 16, 19.) Appellants contend that the Examiner erred in determining that a skilled artisan would have found it obvious to combine DeNuzzio and Fritz to arrive at the claimed invention because "[ t ]he design approaches of the magazines used in Fritz and Denuzzio are [] counter to one another such that incorporating the two approaches together would be technically difficult such that one skilled in the art would not have considered making such a combination" (Reply Br. 4; see also id. at 6). In particular, Appellants assert that DeNuzzio never expressly or inherently discloses retracting the sensing strip 102. In fact, by the sheer design of the cassette having the individual devices 200 stacked upon one another in a fashion similar to an old-fashioned razor dispenser, there is no need to retract any type of test strip back into the device because the individual sensors have to be removed for a new one to be used. Fritz only discloses coupling lancets 502 to the drive 503. At best, after reading both references, one of ordinary skill in the art would be motivated to use the coupling structure in Fritz for the lancets and not the sensing strips. (App. Br. 17; see also Reply Br. 4, 6.) We agree with Appellants, and conclude that the structural and operational differences between Fritz and DeNuzzio are such that the combination proposed by the Examiner would change "the basic principles under which the [prior art] was designed to operate," and is therefore insufficient to establish a prima facie case of obviousness. In re Ratti, 270 F.2d 810, 813 (CCPA 1959). 17 Appeal2013-009264 Application 12/751,178 In particular, we note that the top-to-bottom magazine housing containing ejectable cartridges having an integrated lance and sensing strip assembly taught by DeNuzzio (FF 11, 12) is fundamentally different from, and incompatible with, the barrel-shaped lancet-only magazine disclosed by Fritz (FF 4). For example, DeNuzzio teaches that the integrated lancet and strip containing cartridges are ejected from the magazine after use (FF 12). In addition, DeNuzzio teaches that operation of a push button causes lateral displacement of the lancet on the integrated cartridge to puncture the skin, and subsequent release of the push button withdraws the lancet and activates a spring mechanism that automatically moves the sensing strip into contact with the sample (FF 13). Fritz, in contrast, does not include, and is not designed to accommodate test strips (FF 1--4). In addition, Fritz utilizes a drive unit having a plunger that is coupled to a lancet alone, and teaches "that a form-fitting connection is formed which not only enables a forward movement of the lancet 30 in order to carry out an incision, but also an active retraction of the lancet 30 controlled by the drive unit" (FF 1--4). In view of these differences in the principles of operation for DeNuzzio and Fritz, and absent of any explanation by the Examiner as to why a skilled artisan would have nevertheless been motivated to combine the references, or how such combination would have been achieved, we reverse the rejection of claims 1, 21, and 34. Because they depend from claim 1, the rejection of claims 3 8 and 3 9 is also reversed. Because they depend from claim 21, the rejection of claims 22, 24, 25, 29, 30, and 40 is also reversed. Because they depend from claim 34, the rejection of claims 36 and 41--43 is also reversed. 18 Appeal2013-009264 Application 12/751,178 VI. The Examiner has rejected claim 23 under 35 U.S.C. § 103(a) as obvious based on DeNuzzio, Fritz, and Kloepfer. Because the Examiner has pointed to no clear or specific teaching in Kloepfer that addresses the deficiencies discussed above of DeNuzzio and Fritz with respect to claim 21, we reverse this rejection as well. VII. The Examiner has rejected claim 26 under 35 U.S.C. § 103(a) as obvious based on DeNuzzio, Fritz, and Sacherer. Because the Examiner has pointed to no clear or specific teaching in Sacherer that addresses the deficiencies discussed above of DeNuzzio and Fritz with respect to claim 21, we reverse this rejection as well. VIII. The Examiner has rejected claims 27 and 35 under 35 U.S.C. § 103(a) as obvious based on DeNuzzio, Fritz, and Bonner. Because the Examiner has pointed to no clear or specific teaching in Bonner that addresses the deficiencies discussed above of DeNuzzio and Fritz with respect to claims 21 and 34, we reverse this rejection as well. SUMMARY We summarily affirm the rejection of claims 21-27, 29, 30, 34--36, and 41--43 under 35 U.S.C. § 112, first paragraph, for failure to comply with the written description requirement. We summarily affirm the rejection of claims 1-9, 38, and 39 under 35 U.S.C. § 112, second paragraph, as indefinite. We affirm the rejection of claim 1under35 U.S.C. § 103(a) as obvious based on Fritz and Mauze. Claims 2-9 fall with claim 1. 19 Appeal2013-009264 Application 12/751,178 We atlirm the rejection of claim 1under35 U.S.C. § 103(a) as obvious based on Perez and Fritz. We reverse the rejection of claims 1, 21, 22, 24, 25, 29, 30, 34, 36, 38, 39, and 40-43 under 35 U.S.C. § 103(a) as obvious based on DeNuzzio and Fritz. We reverse the rejection of claim 23 under 35 U.S.C. § 103(a) as obvious based on DeNuzzio, Fritz, and Kloepfer. We reverse the rejection of claim 26 under 35 U.S.C. § 103(a) as obvious based on DeNuzzio, Fritz, and Sacherer. We reverse the rejection of claims 27 and 35 under 35 U.S.C. § 103(a) as obvious based on DeNuzzio, Fritz, and Bonner. TIME PERIOD FOR RESPONSE No time period for taking any subsequent action in connection with this appeal may be extended under 37 C.F.R. § 1.136(a)(l )(iv). AFFIRMED-IN-PART 20 Copy with citationCopy as parenthetical citation