Ex Parte Kim et alDownload PDFBoard of Patent Appeals and InterferencesMar 30, 201211668627 (B.P.A.I. Mar. 30, 2012) Copy Citation UNITED STATES PATENT AND TRADEMARK OFFICE __________ BEFORE THE BOARD OF PATENT APPEALS AND INTERFERENCES __________ Ex parte JAEHO KIM, JOSEPH M. BOCEK, and DAN LI __________ Appeal 2011-000581 Application 11/668,627 Technology Center 3700 __________ Before TONI R. SCHEINER, DONALD E. ADAMS, and JEFFREY N. FREDMAN, Administrative Patent Judges. SCHEINER, Administrative Patent Judge. DECISION ON APPEAL This is an appeal under 35 U.S.C. § 134 from the final rejection of claims 1-27, directed to a cardiac rhythm management device (CRM), specifically, an implantable cardioverter/defibrillator (ICD). The claims have been rejected as anticipated. We have jurisdiction under 35 U.S.C. § 6(b). Appeal 2011-000581 Application 11/668,627 2 BACKGROUND Tachyarrhythmias are abnormal heart rhythms characterized by a rapid heart rate, and generally include ventricular tachyarrhythmia and supraventricular tachyarrhythmia (Spec. 1: 11-13). Supraventricular tachyarrhythmias (SVTs) “are arrhythmias which either originate from the upper chambers of the heart (atria) or involve electrical pathways that connect the upper and lower chambers of the heart.” 1 In other words, SVTs are non-ventricular in origin. “Ventricular cardioversion and defibrillation are used to terminate most ventricular tachyarrhythmias” (id. at 1: 30-31), but “ventricular anti- tachycardia pacing therapy is generally ineffective in terminating an atrial tachyarrhythmia” (id. at 2:11-13). “Additionally, the delivery of each cardioversion/defibrillation pulse consumes a considerable amount of power and results in patient discomfort owing to the high voltage of the shock pulses” (id. at. 2: 13-15). The Specification discloses: [A]n ICD that delivers atrial pacing prior to delivering ventricular anti- tachyarrhythmia if atrial arrhythmia is indicated after a tachyarrhythmia is detected when a ventricular electrogram indicates a fast ventricular rate. A tachyarrhythmia is detected, for example, when the ventricular rate falls within a predetermined rate zone defined by one or more tachyarrhythmia detection thresholds [e.g., the VT zone or the VF zone]. During and/or following the detection, the tachyarrhythmia is classified to determine the necessity and type of a therapy. Because ventricular tachyarrhythmia can be fatal without immediate treatment, the classification process helps ensure that a necessary ventricular therapy is not 1 http://www.heartbeatmd.com/svt.htm (accessed March 26, 2012). Appeal 2011-000581 Application 11/668,627 3 inappropriately delayed or withheld. On the other hand, because a ventricular therapy such as ventricular defibrillation causes significant pain in the patient and shortens the life expectancy of the ICD, it is also desirable to avoid ineffective and/or unnecessary delivery of the ventricular therapy. For example, it has been observed that after a tachyarrhythmia is detected based on ventricular rate and classified as an atrial tachyarrhythmia such as atrial fibrillation (AF) or atrial flutter (AFL), rhythm and waveform characteristics known to be associated with VT may be temporarily present in the ventricular electrocardiogram, leading to an inaccurate VT classification. A subsequent ventricular therapy would be neither necessary nor effective in treating the atrial tachyarrhythmia. The present implantable CRM device delivers atrial pacing after a tachyarrhythmia is detected to avoid such unnecessary, ineffective, and painful ventricular therapy if one or more indications for an atrial tachyarrhythmia are present. (Id. at 5: 29 - 6: l. 19.) STATEMENT OF THE CASE Claims 1-27 are pending and on appeal. Appellants do not present separate arguments for the claims, therefore, we select claim 1 as representative of the subject matter on appeal, as provided by 37 C.F.R. § 41.37(c)(1)(vii). Claim 1 reads as follows (emphases added): 1. An implantable cardioverter/defibrillator (ICD), comprising: a pacing circuit to deliver atrial and ventricular pacing pulses; a defibrillation circuit to deliver ventricular defibrillation pulses; a cardiac sensing circuit to sense cardiac signals; a rate detector to detect an atrial rate and a ventricular rate using the cardiac signals; a tachyarrhythmia detection and classification circuit coupled to the rate detector, the tachyarrhythmia detection and classification circuit including: Appeal 2011-000581 Application 11/668,627 4 a tachyarrhythmia detector to detect tachyarrhythmia using the ventricular rate and one or more tachyarrhythmia threshold rates; and a tachyarrhythmia classifier to classify the detected tachyarrhythmia and detect one or more indications for one or more types of tachyarrhythmia; and a pacing controller, coupled to the pacing circuit, defibrillation circuit, and the tachyarrhythmia detection and classification circuit, to control the delivery of the atrial and ventricular pacing pulses, the pacing controller including an atrial pacing initiator adapted to initiate the delivery of the atrial pacing pulses in a selected atrial pacing mode during the detected tachyarrhythmia if the detected tachyarrhythmia is classified as ventricular tachyarrhythmia and one or more indications for atrial tachyarrhythmia are detected. Claims 1-27 stand rejected under 35 U.S.C. § 102(b) as unpatentable over Kim et al. (US 2005/0149125 A1, published July 7, 2005). We affirm. ISSUE Does the evidence of record support the Examiner‟s finding that Kim‟s ICD is configured to detect a tachyarrhythmia, and to classify the detected tachyarrhythmia, and to administer atrial anti- tachyarrhythmia pacing if the tachyarrhythmia is classified as a ventricular tachyarrhythmia, but indicators of atrial tachyarrhythmia are also present? FINDINGS OF FACT Kim teaches that ventricular tachyarrhythmia (VT) “can be detected when the ventricular rate falls within the VT zone” (Kim ¶ 5), i.e., a zone defined by a predetermined number of beats per minute. Kim further teaches that “[a] rapid ventricular rate in the VT zone, however, is not necessarily due to VT but can also result from a tachycardia that originates from „above‟ the ventricles. Such tachyarrhythmias are Appeal 2011-000581 Application 11/668,627 5 referred to as supraventricular tachycardias (SVTs) and include . . . atrial tachyarrhythmias” (id.). “An abnormal rhythm in the atria . . . can be transmitted ante-gradely to the ventricles . . . [and] is characterized by elevated rates in both the atria and the ventricles” (id.), but “[e]levated rates in both the atria and ventricles can also occur with VT as well, . . . due to retrograde conduction of excitation from the ventricles to the atria” (id.). According to Kim “retrograde conduction is possible in most people and confounds the discrimination between VT and SVT based upon atrial and ventricular rates alone when both rates are similar” (id.). Kim teaches that: Ventricular . . . therapy delivered to treat an SVT will not be effective and potentially could make matters worse by triggering a ventricular arrhythmia . . . It is thus important for an ICD to recognize than an elevated ventricular rate is due to an SVT rather than a VT so that ventricular . . . therapy and specific therapy for the atrial tachyarrhythmia can be delivered if appropriate. Conversely, because VT is generally a more serious condition, the ICD also needs to detect VT with a high degree of sensitivity so that therapy can be delivered promptly. (Id.) The Examiner‟s fact findings regarding Kim are set forth on pages 3-5 of the Answer, and we adopt them as our own, with particular emphasis on the Examiner‟s findings regarding paragraph 28 of Kim, which we reproduce below: When the measured atrial and/or ventricular rates exceed specified threshold values, the device detects an arrhythmia and is programmed to respond with appropriate therapy. For example, if a ventricular rate is measured which is in the VT zone, the device decides whether VT or an SVT is present using the rate and [ECG] morphology criteria . . . If the ventricular rate is greater than the atrial rate, VT is detected, and the device Appeal 2011-000581 Application 11/668,627 6 may be programmed to initiate ventricular anti-tachycardia pacing or a ventricular shock. If the atrial rate is greater than or equal to the ventricular rate and a specified minimum number of normally conducted beats are detected, an SVT is detected, and, if the SVT is classified as an atrial tachyarrhythmia, the device may be programed to deliver atrial anti-tachycardia pacing or an atrial cardioversion shock. Rate stability, suddenness of onset, and/or atrial rate criteria can also be used as additional criteria to classify the arrhythmia as atrial in origin when the ventricular rate is in the VT zone. (Kim ¶ 28.) DISCUSSION Appellants contend that “claim 1 clearly distinguishes between (1) detecting a tachyarrhythmia and (2) then classifying the „detected tachyarrhythmia.‟ . . . [but] “the Examiner‟s . . . interpretation conflates the two” (Reply Br. 3). Appellants contend that “Kim first detects a tachyarrhythmia, by measuring a ventricular rate which is in the VT zone, and only then classifies the tachyarrhythmia, that is „decides whether a VT or SVT is present‟” (id. at 4; Cf. Appellants‟ claim 1, which requires a “tachyarrhythmia detection and classification circuit including: a tachyarrhythmia detector to detect tachyarrhythmia . . .; and a tachyarrhythmia classifer to classify the detected tachyarrhythmia”). Appellants contend that “the Examiner‟s interpretation conflating detecting and classifying impermissibly renders the separately recited „classifying‟ meaningless” (id. at 3). Appellants contend “the express language of claim 1 requires the tachyarrhythmia to be separately detected by the tachyarrhythmia detector and classified by the tachyarrhythmia classifier” (id). “Consequently, it is improper to consider the limitation „classified as a ventricular Appeal 2011-000581 Application 11/668,627 7 tachyarrhythmia‟ . . . to require only „the identification of the ventricular rate within the VT zone‟ at least because the tachyarrhythmia is classified by the tachyarrhythmia classifier rather than the tachyarrhythmia detector alone” (id. at 5). Essentially, Appellants contend that Kim detects a ventricular tachyarrhythmia, and subsequently classifies it as ventricular in origin, in which case ventricular therapy is applied, or classifies it as supraventricular in origin, in which case atrial therapy is applied, whereas the claims require detecting a tachyarrhythmia, then classifying it, and applying atrial therapy if the detected tachyarrhythmia is classified as ventricular, if one or more indications for atrial tachyarrhythmia are also present. Appellants‟ argument is not persuasive. Kim‟s ICD determines whether a detected putative ventricular tachyarrhythmia is ventricular or atrial in origin, based on whether indicators for atrial tachyarrhythmia are also detected (i.e., atrial rate versus ventricular rate, morphology criteria, etc.). For example, if the ICD detects a ventricular tachyarrhythmia, but the atrial rate is faster than the ventricular rate, the tachyarrhythmia is, in Kim‟s words, “classified” as atrial in origin (i.e., supraventricular or non- ventricular in origin) - that is, it is not a true ventricular tachyarrhythmia. In that case, Kim‟s ICD is adapted to deliver atrial anti- tachyarrhythmia pacing or an atrial cardioversion shock, instead of the more extreme and unnecessary ventricular anti- tachyarrhythmia pacing or ventricular cardioversion shock. While Kim classifies a detected ventricular tachyarrhythmia as supraventricular (which triggers atrial therapy), we are persuaded that any difference between the ICD as claimed and Kim‟s ICD is merely semantic. Appeal 2011-000581 Application 11/668,627 8 That is, we see no actual difference between Kim‟s triggering atrial anti- tachyarrhythmia pacing based on detecting a putative ventricular tachyarrhythmia, and subsequently classifying the detected ventricular tachyarrhythmia as an SVT of atrial origin, and triggering atrial anti- tachyarrhythmia pacing based on detecting a tachyarrhythmia and subsequently classifying the tachyarrhythmia as a ventricular tachyarrhythmia accompanied by “indications for atrial tachyarrhythmia,” as required by the claims. That is, Kim‟s classification of the detected putative ventricular tachyarrhythmia as atrial in origin, i.e., as a supraventricular tachyarrhythmia (one that is other than ventricular in origin) is predicated on an underlying detection of a tachyarrhythmia with a ventricular rate in the VT zone, coupled with an indication that the origin of the tachyarrhythmia is not a true ventricular tachyarrhythmia, but is actually atrial in origin (e.g., if the measured atrial rate is greater than the ventricular rate). Thus Kim does disclose separate detection and classification steps as required by the claims, but Kim‟s classification step takes the presence of “indications for atrial tachyarrhythmia” into account in classifying the putative ventricular event as an atrial SVT, i.e., not a true ventricular tachyarrhythmia. SUMMARY The rejection of claims 1-27 as anticipated by Kim is affirmed. TIME PERIOD FOR RESPONSE No time period for taking any subsequent action in connection with this appeal may be extended under 37 C.F.R. § 1.136(a). AFFIRMED Appeal 2011-000581 Application 11/668,627 9 cdc Copy with citationCopy as parenthetical citation