Ex Parte Gunday et alDownload PDFPatent Trial and Appeal BoardJul 16, 201813837970 (P.T.A.B. Jul. 16, 2018) Copy Citation UNITED STA TES p A TENT AND TRADEMARK OFFICE APPLICATION NO. 13/837,970 151971 7590 Forge IP, PLLC 1077 Bridgeport Ave Suite 301 Shelton, CT 06484 FILING DATE FIRST NAMED INVENTOR 03/15/2013 Erhan H. Gunday 07/18/2018 UNITED STATES DEPARTMENT OF COMMERCE United States Patent and Trademark Office Address: COMMISSIONER FOR PATENTS P.O. Box 1450 Alexandria, Virginia 22313-1450 www .uspto.gov ATTORNEY DOCKET NO. CONFIRMATION NO. 04530-P0053A 6030 EXAMINER NEAL, TIMOTHY JAY ART UNIT PAPER NUMBER 3779 NOTIFICATION DATE DELIVERY MODE 07/18/2018 ELECTRONIC Please find below and/or attached an Office communication concerning this application or proceeding. The time period for reply, if any, is set in the attached communication. Notice of the Office communication was sent electronically on above-indicated "Notification Date" to the following e-mail address(es): patent@forge-ip.com PTOL-90A (Rev. 04/07) UNITED STATES PATENT AND TRADEMARK OFFICE BEFORE THE PATENT TRIAL AND APPEAL BOARD Ex parte ERHAN H. GUNDA Y, LAWRENCE J. GERRANS, ALEX HSIA, DEVIN J. S. SCHEIFELE, and JESSIE TUNG Appeal2017-007255 Application 13/837,970 1 Technology Center 3700 Before RICHARD M. LEBOVITZ, FRANCISCO C. PRATS, and JOHN E. SCHNEIDER, Administrative Patent Judges. PRATS, Administrative Patent Judge. DECISION ON APPEAL This appeal under 35 U.S.C. § 134(a) involves claims that recite catheter assemblies. The Examiner entered rejections for anticipation and for obviousness. We have jurisdiction under 35 U.S.C. § 6(b ). We affirm-in-part. STATEMENT OF THE CASE The following rejections are before us for review: 1 Appellants state that the "real party in interest for the present application is Sanovas, Inc., the assignee of the patent application." Appeal Br. 2. Appeal2017-007255 Application 13/837,970 (1) Claims 1, 3-5, and 12, under 35 U.S.C. § I02(b) as being anticipated by Perkins2 (Ans. 2--4); (2) Claims 2, 6, 7, 10, 16, 17, 22, 23, and 25, under 35 U.S.C. § I03(a) as being unpatentable over Perkins and Gunday3 (Ans. 4--7); (3) Claims 13-15, 21, and 26-28, under 35 U.S.C. § I03(a) as being unpatentable over Perkins and Rosenbluth4 (Ans. 7-1 O); (4) Claims 8, 9, and 11, under 35 U.S.C. § I03(a) as being unpatentable over Perkins, Gunday, and Powell5 (Ans. 10-11); (5) Claim 18, under 35 U.S.C. § I03(a) as being unpatentable over Perkins and Frye 6 (Ans. 12); (6) Claims 19 and 20, under 35 U.S.C. § I03(a) as being unpatentable over Perkins and Grice 7 (Ans. 12-13); (7) Claim 24, under 35 U.S.C. § I03(a) as being unpatentable over Perkins and Eidenschink8 (Ans. 13); (8) Claim 24, under 35 U.S.C. § I03(a) as being unpatentable over Perkins and Lafontaine9 (Ans. 14); and (9) Claims 26-28, under 35 U.S.C. § I03(a) as being unpatentable over Perkins and Bales 10 (Ans. 14--15). 2 US 2002/0077593 Al (published June 20, 2002). 3 US 2010/0121270 Al (published May 13, 2010). 4 US 5,007,898 (issued Apr. 16, 1991). 5 US 2009/0264866 Al (published Oct. 22, 2009). 6 US 2005/0004504 Al (published Jan. 6, 2005). 7 US 2006/0293559 Al (published Dec. 28, 2006). 8 US 2008/0077085 Al (published Mar. 27, 2008). 9 US 2008/0161890 Al (published July 3, 2008). 10 US 4,726,374 (issued Feb. 23, 1988). 2 Appeal2017-007255 Application 13/837,970 Claim 1 is representative and reads as follows: 1. A catheter assembly, comprising: a hub having a proximal end and a distal end; a catheter with a proximal portion and a distal portion, wherein the proximal portion of said catheter is located at the distal end of said hub; and at least one balloon at the distal portion of said catheter, said balloon at least partially enclosing an inflation chamber and having a resecting surface for resecting biological material; wherein said catheter has a first lumen in fluid communication with the inflation chamber of said balloon, and a second lumen; and wherein said hub has an inflation port in fluid communication with said first lumen for supplying fluid to inflate said balloon, a delivery port in fluid communication with said second lumen for delivering an agent to the distal portion of said catheter, and an aperture for inserting a device into said second lumen. Appeal Br. 1 7. ANTICIPATION-PERKINS The Examiner's Position In rejecting claims 1, 3-5, and 12 as anticipated by Perkins, the Examiner finds that the reference discloses a catheter assembly having all of the features recited in the rejected claims. Ans. 2-3. In particular, the Examiner finds that, as required by Appellants' claim 1, the balloon of Perkins' s catheter has "a resecting surface for resecting biological material (the outer surface of the balloon can resect biological material)." Id. at 3. Analysis As stated in In re Oetiker, 977 F.2d 1443, 1445 (Fed. Cir. 1992): [T]he examiner bears the initial burden ... of presenting a primafacie case ofunpatentability .... 3 Appeal2017-007255 Application 13/837,970 After evidence or argument is submitted by the applicant in response, patentability is determined on the totality of the record, by a preponderance of evidence with due consideration to persuasiveness of argument. In the present appeal, Appellants persuade us that the preponderance of the evidence does not support the Examiner's finding that Perkins describes a catheter assembly having all of the features required by the rejected claims. In particular, we are not persuaded that the Examiner has shown sufficiently that the balloon of Perkins' s catheter has a resecting surface encompassed by claim 1. The language at issue in claim 1 requires the claimed catheter to include "at least one balloon ... having a resecting surface for resecting biological material." Appeal Br. 17. Although we do not import limitations from the Specification into the claims, Appellants' Specification discloses that examples of resecting surfaces are surfaces textured with a mesh material, or bumps, or other surface protrusions that resect biological materials. Spec. ,r,r 45--46. The Examiner contends that "resecting biological material" as recited in claim 1 encompasses removal of "material such as blood, saliva, mucous, and soft-tissue such as fascia or even individual cells." Ans. 15. The Examiner, however, cites to no evidence of record suggesting that the posited interpretation is reasonable. Appellants, in contrast, point to a dictionary definition to support their position that "[t]he term 'resecting' has a particular meaning in the art, namely something that is specifically designed to perform a resection or 'the surgical removal of part of an organ or structure."' Appeal Br. 7 ( citing http://www.merriam-webster.com/dictionary/resection). 4 Appeal2017-007255 Application 13/837,970 We find that Appellants' argument is supported by persuasive evidence. We acknowledge, as the Examiner contends, that claim 1 recites "resecting biological material." Ans. 15. (emphasis added). As noted above, however, the term "resecting" undisputedly has a dictionary definition that requires surgical removal of part of an organ or structure. Accordingly, the Examiner does not persuade us that, reading the claim as a whole, a skilled artisan would have considered it reasonable to interpret claim 1 as encompassing a balloon with a surface capable only of removing fluid materials like blood or mucus. See In re Cortright, 165 F .3d 1353, 1358 (Fed. Cir. 1999) ("Although the PTO must give claims their broadest reasonable interpretation, this interpretation must be consistent with the one that those skilled in the art would reach."). Thus, that the surface of Perkins's balloon might remove mucus or blood does not demonstrate that Perkins' s balloon has a resecting surface encompassed by claim 1. Alternatively, the Examiner contends as follows: [E]ven if the balloon [of Perkins] is only intended to be used to isolate regions of the lung, the balloon can be used in other ways. The balloon can be moved back and forth within any number of hollow structures ( esophagus, large intestine, blood vessels, etc.) creating friction to remove material from the surface of the structure. The claim does not, as the Examiner previously indicated, include any particular structure related to the resecting surface. The surface is referenced solely by what it does, namely, resect biological material. Perkins discloses a balloon with a surface that can perform this function. Ans. 16. We are not persuaded. We agree with the Examiner that claim 1 is not limited to any particular structure constituting the resecting material. That fact, however, does not relieve the Examiner of providing evidentiary 5 Appeal2017-007255 Application 13/837,970 support for the assertion that the surface of Perkins' s balloon is capable of performing the claimed function of resecting biological material. In the present case, the Examiner does not identify any specific disclosure in Perkins that discusses the structure of the surface of its balloon. Nor does the Examiner identify any evidence of record, in Perkins or elsewhere, demonstrating that Perkins' s balloon necessarily has a surface structure capable of resecting biological material. In the absence of such an evidentiary basis for finding that Perkins' s balloon necessarily has a resecting surface encompassed by Appellants' claim 1, we must reverse the Examiner's rejection of that claim, and its dependent claims 3-5, and 12, as anticipated by Perkins. See In re Oelrich, 666 F.2d 578, 581 (CCPA 1981) ("Inherency ... may not be established by probabilities or possibilities. The mere fact that a certain thing may result from a given set of circumstances is not sufficient."). OBVIOUSNESS-PERKINS AND GUNDA Y The Examiner's Position In rejecting Appellants' claims 2, 6, 7, 10, 16, 17, 22, 23, and 25 for obviousness over Perkins and Gunday, the Examiner cited Perkins for the teachings discussed above, but acknowledged that Perkins differs from the rejected claims in that, contrary to the requirements of claims 2 and 25, Perkins does not disclose its catheter as "necessarily hav[ing] any imaging capabilities (the possibility is not foreclosed; see Paragraph 0021 )." Ans. 4, 7. To address that deficiency, the Examiner cited Gunday as describing "an imaging device disposed in the aperture of said hub and the second lumen of said catheter (Paragraph O 101 ). " Id. The Examiner asserted, 6 Appeal2017-007255 Application 13/837,970 moreover, that "[p ]lacing an inner imaging device through an outer sheath- like device for internal visualization of the body is old and well-known in the art." Id. Based on the references' combined teachings, the Examiner concluded that it would have been obvious to "include an inner catheter with imaging capabilities in Perkins' assembly. Such a modification permits the internal area of a surgical site to be viewed by the surgeon." Id. The Examiner also acknowledged that, contrary to the requirements of claims 16 and 17, Perkins "does not explicitly disclose wherein the resecting surface comprises a mesh affixed to the outer surface of the balloon and wherein the mesh comprises elastane." Id. at 5. To address that deficiency, the Examiner cited Gunday as describing a catheter having a balloon with a resecting surface, "wherein the resecting surface comprises an elastane mesh affixed to the outer surface of the balloon (34, Fig. 2B; Paragraphs 0078, 0081, 0083) for resecting." Id. Therefore, the Examiner concluded, it would have been obvious to "modify Perkins' balloon to include Gunday's resecting surface. Such a modification helps the balloon resect tissue." Id. In response to Appellants' arguments, the Examiner further reasoned that including Gunday's mesh on the surface of Perkins's balloon would have been obvious "in that it will help Perkins' balloon stay in place. Increasing the friction between Perkins' balloon and an anatomical wall is an improvement over Perkins' device in that the balloon is more likely to stay in place and keep the region of the lung isolated." Id. at 16-17. 7 Appeal2017-007255 Application 13/837,970 Analysis In this instance, Appellants do not persuade us that the preponderance of the evidence fails to support the Examiner's conclusion of obviousness. In particular, Appellants do not persuade us that the Examiner erred in concluding that the combination of Perkins and Gunday suggests a catheter having the mesh resecting surface required by Appellants' claims 16 and 1 7. Nor are we persuaded that the Examiner erred in concluding that the combination of Perkins and Gunday also suggests an imaging device disposed within the catheter as required by Appellants' claim 25. 11 We tum first to Appellants' claims 16 and 17. Claim 16 recites "[ t ]he catheter assembly of claim 1, wherein the resecting surface comprises a mesh affixed to the outer surface of the balloon." Appeal Br. 19. Claim 1 7 recites "[ t ]he catheter assembly of 16, wherein the mesh comprises elastane." Id. Appellants do not assert error in the Examiner's finding (Ans. 5 (citing Gunday Fig. 2B and ,r,r 78, 81, and 83)) that Gunday discloses a catheter having a balloon with a resecting surface encompassed by claims 16 and 17. See Appeal Br. 9--10; Reply Br. 5---6. Rather, Appellants contend that it would not have been obvious to modify the balloon of Perkins' s 11 Technically, to present separate argument regarding claims, "[u]nder each heading identifying the ground of rejection being contested, any claim(s) argued separately or as a subgroup shall be argued under a separate subheading that identifies the claim(s) by number." 37 C.F.R. § 4I.37(c)(l)(iv). Although Appellants direct argument to the rejections of claims 16, 17, and claim 25 (see Appeal Br. 9--12), Appellants have not presented separate argument as to any claims by providing a separate subheading, as required. 8 Appeal2017-007255 Application 13/837,970 catheter to have the resecting surface of Gunday's balloon. See id. For the reasons below, we are not persuaded. Perkins describes an "apparatus for isolating sub-bronchial regions of the lung and delivering or retrieving substances from such isolated regions." Perkins ,r 3. As required by claims 16 and 1 7, through their dependence on claim 1, Perkins' s catheter includes a balloon 18 at its distal end. See, e.g., Fig. 5. As Appellants contend (Appeal Br. 9), Perkins discloses that balloon 18 is deployed to a desired location within the lung, and then inflated, thereby isolating the region of the lung distal to the inflated balloon. Perkins ,r 57; see also id. at Fig. 12B (showing inflated balloon 18 and target region of lung "TR" distal to balloon 18). As Appellants contend, Perkins discloses that the region of the lung isolated by the balloon may be treated without treating other portions of the lung, for example by delivering a drug, or delivering a washing fluid for lavage purposes. Id. ,r 58. We, thus, acknowledge that Perkins is directed to using its balloon to allow targeted drug delivery or lavage to an isolated portion of the lung. Appellants do not persuade us, however, that it would have been "undesirable" (Appeal Br. 9; Reply Br. 5) for Perkins's balloon to have a resecting surface encompassed by claims 16 and 1 7. Specifically, as background to the disclosure of the inventive resecting surface of its balloon catheter, Gunday explains that removal from the lungs of undesired or obstructive cancerous materials was known to be desirable. See Gunday ,r,r 2-7. Thus, even if it were true that Perkins viewed alone would not have suggested that a balloon with a resecting surface would be a 9 Appeal2017-007255 Application 13/837,970 desirable feature of a lung-treating catheter, Gunday teaches that, in certain instances, a balloon with a resecting surface is in fact a highly useful feature of a lung-treating catheter, contrary to Appellants' arguments. As to Appellants' contention that modifying Perkins' s catheter to include a balloon with Gunday' s resecting surface would render Perkins' s catheter unsuitable for its intended purpose (Appeal Br. 9--10; Reply Br. 5- 6), we note that Gunday discloses multiple-balloon embodiments in which two outer balloons hold the catheter in place, and an inner balloon performs the resection. See Gunday ,r 91 ( disclosing that "proximal and distal balloon segments (72, 76) ... hold the catheter (24) steady where the tissue to be removed is located while the center balloon (74) is cyclically inflated and deflated to resect the unwanted biological material"); see also id. ("[B]y inflating the proximal and distal balloons (72, 76), one can prevent the resected material from escaping into the bodily cavity, and instead, can capture the loose tissue for easy removal."). Thus, contrary to Appellants' contention that adding Gunday's resecting surface to Perkins' s balloon would render Perkins' s balloon unsuitable for its intended purpose, Gunday discloses that balloons with its resecting surface are not only useful for holding a catheter in place, as the Examiner found (Ans. 16-17), but are also useful for maintaining the isolation of the occluded portion of the lungs, one of Perkins's primary objectives. Consequently, the Examiner's rationale for modifying Perkins based on Gunday's teaching is supported by objective, factual evidence. Accordingly, for the reasons discussed, Appellants do not persuade us that the preponderance of the evidence fails to support the Examiner's conclusion of obviousness as to Appellants' claims 16 and 1 7. We, 10 Appeal2017-007255 Application 13/837,970 therefore, affirm the Examiner's rejection of claims 16 and 17 over Perkins and Gunday. We next tum to Appellants' claim 25. We acknowledge, as Appellants contend (Appeal Br. 11) that claim 25 requires the catheter to be configured such that delivery of an agent may be accomplished through the same lumen in which the imaging device is disposed. See id. at 21 ( claim 25 reciting a "second lumen for delivering an agent to the distal portion of said catheter ... and [an] imaging device is disposed in said second lumen of the catheter"). We acknowledge, as Appellants contend (Appeal Br. 10-11; Reply Br. 6-7), that the outer "isolation catheter" 10 of Perkins' s device includes lumens for the imaging devices ( optical fiber 24 and illuminating fiber 26) that are separate from the main lumen 32. See Perkins ,r,r 47--48; see also id. at Fig. IA (showing lumens 24 and 26 distinct from main lumen 32). Perkins explains, however, that inner catheter 80, which includes the occluding balloon discussed above, is deployed through main lumen 32. See id. ,r 52 ("As illustrated in FIGS. 4 and 5, the inner catheter 80 is sized and adapted to fit within the main lumen 32 of the isolation catheter 10."). Perkins discloses further that "the inner catheter may in some instances incorporate optical and illumination fibers .... " Id. ,r 21. Thus, despite Perkins's disclosure its inner catheter "will usually not have such imaging capabilities" (id.), and despite Perkins's disclosure that imaging devices are generally deployed through lumens that are distinct from the main lumen (id. ,r 47), Perkins nonetheless also expressly discloses that its inner catheter, which is deployed through the main lumen of the 11 Appeal2017-007255 Application 13/837,970 outer isolation catheter, "may in some instances incorporate optical and illumination fibers." Id. ,r 21. In addition to this disclosure that optical and illumination fibers may be delivered through the main lumen 32 of its outer isolation catheter via the inner catheter, Perkins also discloses that, as required by Appellants' claim 25, agents may be delivered through the main lumen. See id. ,r 52 ( disclosing that "materials can be withdrawn or introduced through the isolation catheter 10 as well as through the lumen of the inner catheter 80") ( emphasis added). Thus, given Perkins' s disclosure that, in certain instances, optical and illumination fibers may be delivered through the main lumen 32 of its outer isolation catheter via the inner catheter, and given Perkins's disclosure that therapeutic material may also be delivered through the main lumen of the outer isolation catheter, Appellants do not persuade us that Perkins fails to suggest a catheter in which an imaging device and agents are deployed through the same lumen, as Appellants' claim 25 requires. Accordingly, for the reasons discussed, Appellants do not persuade us that the preponderance of the evidence fails to support the Examiner's conclusion of obviousness as to Appellants' claim 25. We, therefore, affirm the Examiner's rejection of claim 25 over Perkins and Gunday. Claims 2, 6, 7, 10, 22, and 23 fall with claim 25. See 37 C.F.R. § 4I.37(c)(l)(iv). OBVIOUSNESS-PERKINS AND ROSENBLUTH In rejecting Appellants' claims 13-15, 21, and 26-28 over Perkins and Rosenbluth, the Examiner relied on Perkins for the teachings discussed above, including the disclosure of a resecting surface, and cited Rosenbluth as evidence that certain additional elements recited in claims 13-15, 21, and 12 Appeal2017-007255 Application 13/837,970 26-28, such as a silicone plug, would have been obvious features of Perkins's catheter. Ans. 7-10. Each of claims 13-15 and 21 depends directly or ultimately from claim 1, which includes a balloon with a resecting surface, as discussed above. See Appeal Br. 19-20. Independent claim 26, like claim 1, includes "at least one balloon ... having a resecting surface for resecting biological material." Id. at 21. Claims 27 and 28 depend from claim 26. Id. at 22. As discussed above, the Examiner does not persuade us that Perkins describes a catheter having a balloon with a resecting surface. Because the Examiner does not identify, nor do we discern, any teaching in Rosenbluth that remedies that deficiency as to claims 13-15, 21, and 26-28, we reverse the Examiner's rejection of claims 13-15, 21, and 26-28 over Perkins and Rosenbluth. OBVIOUSNESS- PERKINS, GUNDA Y, AND POWELL In rejecting claims 8, 9, and 11 over Perkins, Gunday, and Powell, the Examiner relied on the teachings in Perkins and Gunday discussed above, and cited Powell as evidence that the elements recited in claims 8, 9, and 11 would have been obvious features of the device suggested by Perkins and Gunday. Ans. 10-11. Each of claims 8, 9, and 11 depends ultimately from claim 1. See Appeal Br. 18. Appellants argue only that "[t]he purported combinations of these references suffer from the same deficiencies discussed above with respect to claim 1 and also regarding Gunday et al. with respect to claim 25." Id. at 14. 13 Appeal2017-007255 Application 13/837,970 We are not persuaded. As discussed above, we agree with the Examiner that it would have been obvious to include Gunday's resecting surface on Perkins' s balloon. Because Appellants therefore do not persuade us that the combination of Perkins, Gunday, and Powell suggests a catheter assembly having all of the features required by claims 8, 9, and 11, we affirm the Examiner's rejection of claims 8, 9, and 11 over Perkins, Gunday, and Powell. OBVIOUSNESS- REMAINING REJECTIONS In the remaining rejections, the Examiner rejected claims 18-20, 24, and 26-28 over Perkins combined with either Frye, Grice, Eidenschink, Lafontaine, or Bales. Ans. 12-15. In making those rejections, the Examiner relied on Perkins for the teachings discussed above, including the disclosure of a resecting surface, and cited the additional references evidence that the elements recited in claims 18-20, 24, and 26-28 would have been obvious features of Perkins's catheter. See id. Claims 18-20 and 24 depend directly or ultimately from claim 1, and thus include the resecting surface feature recited in claim 1. See Appeal Br. 19-20. As noted above, moreover, claims 26-28 also include the resecting surface feature. As discussed above, the Examiner does not persuade us that Perkins describes a catheter having a balloon with a resecting surface. Because the Examiner does not identify, nor do we discern, any teachings in Frye, Grice, Eidenschink, Lafontaine, or Bales that remedy that deficiency as to claims 14 Appeal2017-007255 Application 13/837,970 18-20, 24, and 26-28, we reverse the Examiner's rejections of claims 18- 20, 24, and 26-28. SUMMARY For the reasons discussed, we reverse the Examiner's rejection of claims 1, 3-5, and 12, under 35 U.S.C. § 102(b) as being anticipated by Perkins; For the reasons discussed, however, we affirm the Examiner's rejection of claims 2, 6, 7, 10, 16, 17, 22, 23, and 25, under 35 U.S.C. § 103(a) for obviousness over Perkins and Gunday; For the reasons discussed, we reverse the Examiner's rejection of claims 13-15, 21, and 26-28, under 35 U.S.C. § 103(a) for obviousness over Perkins and Rosenbluth; For the reasons discussed, however, we affirm the Examiner's rejection of claims 8, 9, and 11, under 35 U.S.C. § 103(a) for obviousness over Perkins, Gunday, and Powell; For the reasons discussed, we reverse the Examiner's rejection of claim 18, under 35 U.S.C. § 103(a) for obviousness over Perkins and Frye; For the reasons discussed, we reverse the Examiner's rejection of claims 19 and 20, under 35 U.S.C. § 103(a) for obviousness over Perkins and Grice; For the reasons discussed, we reverse the Examiner's rejection of claim 24, under 35 U.S.C. § 103(a) for obviousness over Perkins and Eidenschink; For the reasons discussed, we reverse the Examiner's rejection of claim 24, under 35 U.S.C. § 103(a) for obviousness over Perkins and Lafontaine; and 15 Appeal2017-007255 Application 13/837,970 For the reasons discussed, we reverse the Examiner's rejection of claims 26-28, under 35 U.S.C. § 103(a) for obviousness over Perkins and Bales. TIME PERIOD No time period for taking any subsequent action in connection with this appeal may be extended under 37 C.F.R. § 1.136(a). AFFIRMED-IN-PART 16 Copy with citationCopy as parenthetical citation