Ex Parte FukushimaDownload PDFPatent Trial and Appeal BoardJul 21, 201411656310 (P.T.A.B. Jul. 21, 2014) Copy Citation UNITED STATES PATENT AND TRADEMARK OFFICE ____________ BEFORE THE PATENT TRIAL AND APPEAL BOARD ____________ Ex parte NAOTO FUKUSHIMA ____________ Appeal 2012-002151 Application 11/656,310 Technology Center 2600 ____________ Before LORA M. GREEN, JEFFREY N. FREDMAN, and ULRIKE W. JENKS, Administrative Patent Judges. JENKS, Administrative Patent Judge. DECISION ON APPEAL This is an appeal1 under 35 U.S.C. § 134 involving claims directed to an image processing system. The Examiner has rejected the claims as anticipated. We have jurisdiction under 35 U.S.C. § 6(b). We affirm. STATEMENT OF THE CASE The Specification provides that “the e-Document Law, etc., specify reading condition standard values for adding a digital signature or time stamp to a document digitized by scanning.” (Spec. 20: ¶ 40.) 1 Appellant identifies Sharp Kabushiki Kaisha as the Real Party in Interest. (App. Br. 1.) Appeal 2012-002151 Application 11/656,310 2 Claims 1-27 are on appeal, and can be found in the Claims Appendix of the Appeal Brief. Claim 1 is representative of the claims on appeal, and reads as follows (emphasis added): Claim 1: An image processing apparatus comprising: a reading portion that optically reads an original document and inputs image data; and a digital signature adding portion that adds a digital signature to the read image data, wherein the image processing apparatus comprises: a reading condition value setting portion that sets, at the time of reading of the original document by the reading portion, reading condition values by a kind or an address for submission of the original document; and a reading condition matching determining portion that determines whether the set reading condition values are greater than or equal to predetermined reading condition standard values, the predetermined reading condition standard values corresponding to minimum values of at least one of resolution measured in dots per inch or equivalent units, color, and gradation of the reading portion that are stored in a reading condition table, wherein the digital signature adding portion does not add a digital signature to the image data read at the reading portion when, as a result of determination by the reading condition matching determining portion, the reading condition values do not match the predetermined reading condition standard values. The following rejection is before us for review: Claims 1-27 under 35 U.S.C. § 102(b) as being anticipated by Randle.2 2 Randle et al., US 2005/0071283 Al, published Mar. 31, 2005. Appeal 2012-002151 Application 11/656,310 3 The Issue Does the evidence of record support the Examiner’s finding that the claims are anticipated? Findings of Fact FF1. We adopt the Examiner’s findings and analysis concerning the scope and content of Randle. The following facts are repeated for reference convenience. FF2. Randle disclosed that image quality is measured “in terms of capture resolution (optical and pixel or color depth), capture device quality (specifications such as the Modulation Transfer Function of the device), image format, compression techniques, associated meta data, and the like are significant.” (Randle 3: ¶ 31; see also Ans. 5.) FF3. Randle disclosed To perform the collection of data by a capture device 310, the paper check is exposed to the device to image the check. A machine scan of the paper check is performed to collect and determine data such as MICR data (RTN, account number, check number, amount, etc.), OCR/ICR derived data, image processing derived variables or metrics, including IQI and ISI and the file size and the check size. The captured image may be segmented by a grid and QA [Quality Assurance] performed on one or more than one element and or subset of elements identified in one or more segment of the grid. Through this process, critical data may be identified and non- critical segments of the image eliminated from additional QA processing[.] The operator, which can be a human or can be a computer applied method, optionally inputs meta data such as the Image Quality Value (IQV), the amount, the payee, the payer, the check number, and the like. IQV may be automated and in itself may be part of the QA process and may be machine derived or human based depending on the capture needs. Derived data (data calculated based on known and information Appeal 2012-002151 Application 11/656,310 4 collected) and configuration data (known data about the device, process, check type, etc.) are also determined. Optionally, calibration data, such as established data, resulting from the calibration process is used in the QA process. (Randle 9: ¶ 107; see also Ans. 19.) FF4. Randle disclosed that “[a]fter imaging, the image file is compared to information, such as expected image data 320, including but not limited to expected file size. Any image that is outside of an expected range is reimaged and or flagged for exception processing.” (Randle 9: ¶ 108; see also Ans. 20.) FF5. The Examiner finds that Randle discloses a method of determining if image file resolution falls within a set range. A range has two boundaries. The lower boundary is the minimum value of resolution and it is determined if the image file information is greater than or equal to. And the same is done for the upper boundary in other [sic] to determine if the value is within the set range. (Ans. 20.) FF6. Randle disclosed Where the image file falls within a predetermined range of acceptable expected file sizes, the image file is further tested based on established standards 340. Images undergo the QA decision process to produce a result representing the correlation of the image to a previously defined Correlation Index. Indexes explained below are used alone or in combination to determine a Quality Assurance Correlation Value (QACV). QACV in combination with the Correlation Index definition establishes the standard by which all captured images are valued for QA. Images that do not meet the QACV are flagged for exception processing 360, or alternatively, undergo additional imaging and or manual or machine based QA steps or QA completion Appeal 2012-002151 Application 11/656,310 5 350 prior to finally flagging for paper processing or approval of the image. (Randle 9: ¶ 110.) FF7. Randle disclosed Capture Device Quality Index (CDQI) is a value or range of values that represent the capture devices ability to accurately capture an image. The index may include one or more specifications such as pixel resolution, pixel depth, color depth, pixel noise, device system signal to noise ratio, pixel correlation noise, the device’s Modulation Transfer Function (MTF) and the like. (Randle 9: ¶ 127.) FF8. Randle disclosed that the “Image Quality Index (IQI): is a value or range of values that are generated from processing the image in terms of the expected resolution and quality.” (Randle 11: ¶ 122; see also Ans. 5.) FF9. Randle disclosed IQI is a derived value used as a QA qualifier for decision making. IQI establishes a value for the captured image that correlates back to the check type and the capture device’s ability to capture data consistently and accurately. Correlation to expected results or a range of results is the basis for decision making using only the IQI. (Randle 11: ¶ 123.) Principle of Law “A single prior art reference that discloses, either expressly or inherently, each limitation of a claim invalidates that claim by anticipation.” Perricone v. Medicis Pharmaceutical Corp., 432 F.3d 1368, 1375 (Fed. Cir. 2005). Appeal 2012-002151 Application 11/656,310 6 Analysis Appellant contends that “‘IQI’ is not equivalent to actual resolution values, and thus the expected range of IQI values is not equivalent to a range of ‘resolution’ values.” (Reply Br. 2; see also Ans. 11.) “[T]he Randle reference does not teach or suggest that resolution values are being compared, but only that the derived value of IQI, which is used for decision making purposes, must fall within a particular range.” (App. Br. 11.) We are not persuaded by Appellant’s contention that Randle does not compare actual resolution values. We agree with the Examiner’s position that the claims are not limited to measuring resolution in a particular manner, such as dots per inch. The use of the “‘or equivalent’ [language in the claim] makes it an open statement” that allows other measures for determining the resolution (Ans. 20). Appellant provides no evidence rebutting the Examiner’s reasonable interpretation of “or equivalent” in claim 1. Here, Randle provides that capture resolution is based on pixel or color depth (FF 2 & 7). According to Randle the image quality is a measurement in terms of the capture resolution based on the device quality (FF 2-9). The image quality index (IQI) is a table that is generated from scanning images with a particular device and provides a measure in terms of expected resolution and quality for a particular device (FF 2-8). It is the correlation of captured values of a particular image to those provided in the IQI index that forms the basis of the decision making in Randle (FF 9). We agree with the Examiner’s finding that the disclosure of a range indicates that there is an upper and lower limit for acceptable image quality for each listed device (FF5). Even if the IQI index is comprised of derived Appeal 2012-002151 Application 11/656,310 7 values, these values are based on actual measurements of resolution data for each particular device and identifies an acceptable resolution range that is listed in the index (FF 2-9). As explained by the Examiner: The IQI is one of the parameter that the QA process looks at and correlate it to what is expected, here the range. Of course, when a document is read, parameters that the QA refers to are extracted from the reading of the document (see Randle [0107]) and compared to the values that are set (see Randle [0108]). So there is an IQI that is obtained from the document read and an IQI range that is set in table 1 of Randle [0113] for correlation. (Ans. 19.) We find no error with Examiner’s finding that Randle discloses an image processing apparatus disclosing “a reading condition matching determining portion that determines whether the set reading condition values are greater than or equal to predetermined reading condition standard values” (Ans. 4-5; FF 2-9). Therefore, we are not persuaded by Appellant’s contention that Randle fails to “teach or suggest that resolution values are being compared” (App. Br. 11). SUMMARY We affirm the rejection of claims 1-27 under 35 U.S.C. § 102(b) by Randle. No time period for taking any subsequent action in connection with this appeal may be extended under 37 C.F.R. § 1.136(a). AFFIRMED Appeal 2012-002151 Application 11/656,310 8 cdc Copy with citationCopy as parenthetical citation