Ex Parte Deno et alDownload PDFPatent Trial and Appeal BoardMar 28, 201411342785 (P.T.A.B. Mar. 28, 2014) Copy Citation MOD PTOL-90A (Rev.06/08) APPLICATION NO. FILING DATE FIRST NAMED INVENTOR ATTORNEY DOCKET NO. 11/342,785 01/30/2006 D.Curtis Deno P0022582.00/LG10126 EXAMINER MEDTRONIC, INC. (CRDM) 710 MEDTRONIC PARKWAY NE MS: LC340 LEGAL PATENTS MINNEAPOLIS, MN 55432-9924 PORTER, JR. GARY A ART UNIT PAPER NUMBER 3766 MAIL DATE DELIVERY MODE 03/28/2014 ELECTRONIC Please find below and/or attached an Office communication concerning this application or proceeding. The time period for reply, if any, is set in the attached communication. UNITED STATES DEPARTMENT OF COMMERCE U.S. Patent and Trademark Office Address : COMMISSIONER FOR PATENTS P.O. Box 1450 Alexandria, Virginia 22313-1450 www.uspto.gov UNITED STATES PATENT AND TRADEMARK OFFICE _____________________________________________________________________________________ UNITED STATES PATENT AND TRADEMARK OFFICE __________ BEFORE THE PATENT TRIAL AND APPEAL BOARD __________ Ex parte D. CURTIS DENO and WILLIAM J. HAVEL __________ Appeal 2011-011774 Application 11/342,785 Technology Center 3700 __________ Before LORA M. GREEN, JEFFREY N. FREDMAN, and ULRIKE W. JENKS, Administrative Patent Judges. PER CURIAM DECISION ON APPEAL This is a decision on appeal1 under 35 U.S.C. § 134 from the Examiner’s rejection of claims 13-18, 21, and 23. We have jurisdiction under 35 U.S.C. § 6(b). We affirm. 1 The Real Party in Interest is Medtronic Inc. (App. Br. 3). Appeal 2011-011774 Application 11/342,785 2 STATEMENT OF THE CASE The Specification discloses that “electrical defibrillation restoring normal sinus rhythm (NSR) after a prolonged period of fast ventricular tachyarrhythmia (VT) or ventricular fibrillation (VF) may not . . . restore adequate mechanical function of the heart” (Spec. 1, ¶ 03). “[T]his condition, known as pulseless electrical activity (PEA), is likely due to ischemic stunning” (id.). According to the Specification, “[i]n some cases of mechanical cardiac dysfunction, PESP [post-extrasystolic potentiation] therapy has been found to restore an adequate response of the cardiac muscle to electrical depolarization, thereby increasing cardiac output” (id. at 1, ¶ 02). “PESP is achieved by delivering electrical pulses to a ventricle soon after a refractory period of a previous ventricular depolarization expires” (id.). Claim 13 is representative of the claims on appeal and reads as follows (emphasis added): 13. A medical device system adapted for implantation in a patient having a heart for the treatment of the heart having an arrhythmia having a duration from an onset of the arrhythmia to a termination of the arrhythmia, the system comprising: a pair of electrodes; a set of low polarization electrodes configured to sense electrical activity of the heart; a counter operatively coupled to the set of low polarization electrodes and configured to log the duration of the arrhythmia from the onset of the arrhythmia to the termination of the arrhythmia; and a therapy module operatively coupled to the pair of electrodes and the counter, and configured to deliver over the pair of electrodes: an electrical stimulation to terminate the arrhythmia; and electrical potentiation pulses when the duration of the arrhythmia from the onset of the arrhythmia to the termination of the Appeal 2011-011774 Application 11/342,785 3 arrhythmia is longer than approximately 30 seconds and is shorter than approximately 300 seconds corresponding to a prescribed time window. Issue The Examiner has rejected claims 13, 14, 18, 21, and 23 under 35 U.S.C. § 103(a) as obvious in view of Deno2 (Ans. 3-6). Claims 14, 18, 21, and 23 have not been argued separately from claim 13 and therefore stand or fall with that claim. 37 C.F.R. § 41.37(c)(1)(vii). The Examiner has also rejected, under 35 U.S.C. § 103(a), claim 15 over the combination of Deno and Lovett3 (Ans. 6) and claims 16 and 17 over the combination of Deno and Bennett4 (id. at 6-7). Because Appellants only argue that Lovett and Bennett do not remedy the deficiencies of Deno, we will consider these rejections together (App. Br. 18-20). The issue presented is: Has the Examiner established by a preponderance of the evidence that Deno would have made obvious a medical device system for the treatment of heart arrhythmia comprising a therapy module configured to deliver electrical potentiation pulses when the arrhythmia duration, from onset to termination, “is longer than approximately 30 seconds and is shorter than approximately 300 seconds,” as required by claim 1? 2 Deno et al., US 2004/0049235 A1, published Mar. 11, 2004. 3 Lovett et al., US 2004/0215240 A1, published Oct. 28, 2004. 4 Bennett et al., US 5,331,966, issued Jul. 26, 1994. Appeal 2011-011774 Application 11/342,785 4 Findings of Fact FF1. The Examiner finds that “Deno discloses an ICD [implantable cardioverter/defibrillator] system 14 . . . comprising electrodes at the end of endocardial leads 16, 32 and 52 that are configured to pace the heart and/or sense the response of the heart to therapy” (Ans. 3-4). FF2. The Examiner finds that Deno “discloses that these leads are low polarization lead systems” (id. at 4). FF3. The Examiner finds that Deno’s Example 2 “discloses a counter configured to log a duration of any detected tachyarrhythmia from beginning to termination and then subsequently initiates therapy with a therapy module if the duration of the episode exceeds a predetermined threshold value” (id.). FF4. The Examiner finds that “Deno does not explicitly disclose that the predetermined threshold range is between 30 seconds and 300 seconds” (id.). FF5. The Examiner finds that the instant Specification indicates “that a time frame of greater than 30 seconds is needed to result in ischemic stunning of the heart while not attributing any criticality to the upper limit of 300 seconds” (id., citing the Spec. at 6, ¶ 19). FF6. The Examiner finds that, with regard to the lower threshold of 30 seconds, . . . [Deno] states ‘If the duration of the episode exceeded a programmable threshold before being terminated, indicating that the likelihood was high that cardiac mechanical function was severely impaired, the device would initiate a NES/PESP therapy for a fixed duration following the episode (Id. at 11, citing Example 2). Appeal 2011-011774 Application 11/342,785 5 FF7. The Examiner finds that “[r]egarding the upper threshold of 300 seconds claimed by Appellants, the Examiner notes that a general upper limit of 300 seconds exists in the teaching of Deno” (id. at 8). FF8. The Examiner finds that Deno states ‘A patient with a history of HF may not tolerate a tachyarrhythmia well for more than a few minutes . . . Despite prompt and good care, defibrillation after a prolonged several minutes of cardiac ischemia may result in EMD/PEA or asystole and death.’ Deno ‘235 clearly states that allowing an arrhythmia to persist for longer than a ‘few minutes’ would be detrimental to the patient. (Id.) FF9. The Examiner concludes that it would have been obvious to one having ordinary skill in the art . . . to set the detection threshold between 30 seconds and 300 seconds, since it has been held that where the general conditions of a claim are disclosed in the prior art (in this case, the general condition is treating ischemic stunning with PESP), discovering the optimum or workable ranges involves only routine skill in the art. In re Aller, 105 USPQ 233. (Id. at 4-5.) FF10. The Examiner reasons that “finding the optimal time frame for indicating that ischemic stunning is present after the termination of a tachyarrhythmia would only require routine skill in the art in light of the Deno . . . teaching” (id. at 5). FF11. Deno discloses that “[d]elivering pacing during the refractory period is a type of non-excitatory stimulation (NES) that causes the release of catecholamines such as norepinephrine within the tissue of the heart . . . Appeal 2011-011774 Application 11/342,785 6 [that] results in an increased contractility of the cardiac tissue” (Deno 4, ¶ 0027). FF12. Deno discloses that “[a]nother type of electrical stimulation can be provided during the nonrefractory period of the cardiac cycle. This type of stimulation results in an additional electrical depolarization and, when appropriately timed, results in post extrasystolic potentiation (PESP)” (id. at 4, ¶ 0030). FF13. In Example 2, Deno discloses that “[d]uring a tachyarrhythmia, the coronary blood flow perfusing the heart can become severely impaired, leading to ischemia and a temporary loss in cardiac contractility referred to as stunning. If the loss in contractility is severe . . . further ischemia will result” (id. at 25, ¶ 0246). FF14. In Example 2, Deno discloses that [a]n example of a PESP/NES stimulation therapy would include treating impaired cardiac mechanical function following a tachyarrhythmia. . . . If the duration of the episode exceeded a programmable threshold before being terminated, indicating that the likelihood was high that cardiac mechanical function was severely impaired, the device would initiate a NES/PESP therapy . . . to hasten re-perfusion of the heart . . . (Id. at 25, ¶ 0247.) FF15. In Example 4, Deno discloses that a “patient with a history of HF may not tolerate a tachyarrhythmia well for more than a few minutes” (id. at 26, ¶ 0256). Appeal 2011-011774 Application 11/342,785 7 Analysis Appellants argue that Deno “does not show, disclose or suggest a therapy module that triggers electrical potentiation therapy when the duration logged in a counter is longer than approximately 30 seconds and is shorter than approximately 300 seconds” (App. Br. 11). Appellants argue that Deno “does not set a lower limit for a timeframe, i.e., does not show, disclose or suggest that the duration . . . must . . . be greater than a particular time” (id.). That argument is not persuasive. As recognized by the Examiner (Ans. 11), Deno discloses a programmable threshold for an arrhythmia duration that would trigger the initiation of NES/PESP therapy following termination of the arrhythmia episode. Deno also discloses that achievement of the threshold indicates a high likelihood that cardiac mechanical function has occurred. “[W]here general conditions of a claim are disclosed in the prior art, it is not inventive to discover the optimum or workable ranges by routine experimentation.” In re Aller, 220 F.2d 454, 456 (CCPA 1955). Thus, in accord with In re Aller, because Deno discloses both the concept of a threshold and the purpose of the threshold, it would have been obvious for one of skill in the art to optimize the length of the threshold duration to deliver post-arrhythmia therapy to only those patients that require such therapy. Appellants also argue that Deno “does not discuss withholding PESP therapy unless the duration is less than approximately 300 seconds since Deno . . . sets no upper limit whatsoever” (App. Br. 12). Appellants argue that although Deno’s Example 2 discloses PESP therapy after the Appeal 2011-011774 Application 11/342,785 8 termination of the arrhythmia, the Examiner improperly relied Deno’s Example 4, which relates to PESP during supraventricular tachycardias, for the suggestion of an upper limit. Those arguments are not persuasive. Although Deno’s Example 4 relates to PESP during supraventricular tachycardias, Example 4 provides the concept that patients with different heart conditions may tolerate arrhythmias to different degrees and that for some patients, arrhythmia’s that last longer than five minutes may present different considerations. The Supreme Court has emphasized that “the [obviousness] analysis need not seek out precise teachings directed to the specific subject matter of the challenged claim, for a court can take account of the inferences and creative steps that a person of ordinary skill in the art would employ.” KSR Int’l v. Teleflex Inc., 550 U.S. 398, 418 (2007); see also id. at 421 (“A person of ordinary skill is also a person of ordinary creativity, not an automaton.”). “If a person of ordinary skill can implement a predictable variation, § 103 likely bars its patentability.” (Id.) Here, since Deno suggests that different patients may tolerate arrhythmia’s differently, it would have been obvious to set an upper limit on the duration of an arrhythmia that would be followed by PESP therapy so that other interventions could be employed in situations where longer-duration arrhythmia’s would be problematic. Thus, we affirm the rejection of claim 13 as being obvious in view of Deno, and claims 14, 18, and 23 fall with claim 13. 37 C.F.R. § 41.37(c) (1) (vii). Appeal 2011-011774 Application 11/342,785 9 As to claim 21, Appellants reiterate the argument made with respect to claim 13, that is that Deno “does not show, disclose or suggest a therapy module operatively coupled to the pair of electrodes and the counter, and configured to deliver over the pair of electrodes the first electrical stimulation pulse and the second electrical stimulation pulse to terminate the arrhythmia and electrical potentiation pulses when the duration between the first electrical stimulation and the second electrical stimulation is longer than approximately 30 seconds and is shorter than approximately 300 seconds within a prescribed time window” (App. Br. 17). That argument is not found to be convincing for the reasons set forth above. The Examiner has also rejected dependent claims 15, 16, and 17 under 35 U.S.C. § 103(a) in view of Deno and further in view of either Lovett or Bennett. Appellants argue that neither Lovett nor Bennett cures the deficiencies of Deno with respect to the claim 13 limitation of delivering electrical potentiation pulses when the duration of the arrhythmia is longer than approximately 30 seconds and is shorter than approximately 300 seconds (App. Br. 18-20). Those arguments are not persuasive because, as discussed above, we have concluded that claim 13 is obvious in view of Deno. Conclusion of Law The preponderance of the evidence supports the Examiner’s conclusion that Deno would have made obvious a medical device system for the treatment of heart arrhythmia comprising a therapy module configured to deliver electrical potentiation pulses when the arrhythmia duration, from Appeal 2011-011774 Application 11/342,785 10 onset to termination, is longer than approximately 30 seconds and is shorter than approximately 300 seconds. SUMMARY We affirm the rejection of claims 13-18, 21, and 23 under 35 U.S.C. § 103(a). TIME PERIOD FOR RESPONSE No time period for taking any subsequent action in connection with this appeal may be extended under 37 C.F.R. § 1.136(a). AFFIRMED dm Copy with citationCopy as parenthetical citation