N.H. Admin. Code § He-P 4044.10

Current through Register No. 36, September 5, 2024
Section He-P 4044.10 - Operating Procedures
(a) Particle accelerators, when not in operation, shall be secured to prevent unauthorized use.
(b) The safety interlock system shall not be used to turn off the accelerator beam except in an emergency.
(c) All safety and warning devices, including interlocks, shall be checked for proper operation at intervals not to exceed 3 months. Results of such tests shall be maintained at the accelerator facility for inspection by DHHS/RHS.
(d) Electrical circuit diagrams of the accelerator, and the associated safety interlock systems, shall be kept current and maintained for inspection by DHHS/RHS and available to the operator at each accelerator facility.
(e) If, for any reason, it is necessary to intentionally bypass a safety interlock or interlocks, such action shall be:
(1) Authorized by the radiation safety committee and the radiation safety officer after having notified the radiation safety committee of such action;
(2) Recorded in a permanent log and a notice posted at the accelerator control console; and
(3) Terminated as soon as possible.
(f) A copy of the current operating and the emergency procedures shall be maintained at the accelerator control panel.

N.H. Admin. Code § He-P 4044.10

#6827, eff 8-6-98; ss by #8692, INTERIM, eff 7-27-06, EXPIRES: 2-2-07; ss by #8808, eff 1-24-07

Amended byVolume XXXV Number 14, Filed April 9, 2015, Proposed by #10804, Effective 3/26/2015, Expires3/26/2025.