Ex Parte Mongeon et alDownload PDFBoard of Patent Appeals and InterferencesJul 23, 201211312874 (B.P.A.I. Jul. 23, 2012) Copy Citation UNITED STATES PATENT AND TRADEMARK OFFICE ____________________ BEFORE THE BOARD OF PATENT APPEALS AND INTERFERENCES ____________________ Ex parte LUC R. MONGEON, KAREN J. KLECKNER, and JOHN C. RUETER ____________________ Appeal 2010-004669 Application 11/312,874 Technology Center 3700 ____________________ Before: JENNIFER D. BAHR, WILLIAM V. SAINDON, and MICHAEL J. FITZPATRICK, Administrative Patent Judges. BAHR, Administrative Patent Judge. DECISION ON APPEAL Appeal 2010-004669 Application 11/312,874 2 STATEMENT OF THE CASE Luc R. Mongeon et al. (Appellants) originally appealed under 35 U.S.C. § 134 from the Examiner’s decision rejecting claims 1, 3-13, and 15- 20. Notice of Appeal, filed July 28, 2009. Appellants subsequently withdrew claims 1 and 3-12 from the appeal.1 App. Br. 3. Thus, only claims 13 and 15-20 are involved in this appeal. We have jurisdiction under 35 U.S.C. § 6(b). We AFFIRM. The Claimed Subject Matter Claim 13, reproduced below, is illustrative of the claimed subject matter. 13. An apparatus for monitoring, delivering and/or adjusting cardiac therapy delivery, comprising: means for monitoring at least one of an atrial-left ventricular (A-LV) interval, an atrial-right ventricular (A-RV) interval, a left ventricular (LV) to right ventricular (RV) interval; means for storing in a memory structure said at least one of the A-LV interval, the A-RV interval, the LV to RV interval; means for performing a temporal trend analysis upon said at least one of the foregoing stored intervals over a discrete period of time, utilizing a plurality of discrete values of said stored intervals; and means for performing an automated ventricular chamber capture management pacing threshold test for at least one of the LV and the RV, wherein said automated threshold test is triggered by one of: a clinician, a relatively longer duration QRS complex, a morphologically relatively different QRS complex. 1 We suggest that the Examiner cancel these claims upon return of jurisdiction of this application to the Examiner. See Ex Parte Ghuman, 88 USPQ2d 1478, 1480 (BPAI 2008) (precedential). Appeal 2010-004669 Application 11/312,874 3 Evidence The Examiner relied on the following evidence in rejecting the claims on appeal: Florio US 6,512,953 B2 Jan. 28, 2003 Rejections Claims 13 and 16-20 are rejected under 35 U.S.C. 102(b) as anticipated by Florio. Ans. 4. Claim 15 is rejected under 35 U.S.C. § 102(b) as anticipated by or, in the alternative, under 35 U.S.C. § 103(a) as unpatentable over Florio. Ans. 5. OPINION Appellants group claims 13, 16, and 17 together in contesting the anticipation rejection. Thus, in accordance with 37 C.F.R. § 41.37(c)(1)(vii) (2011), we select claim 13 as representative of the group, with claims 16 and 17 standing or falling with claim 13. Appellants present a separate argument for claims 18-20. Appellants argue that “Florio does not disclose an apparatus comprising means for storing and trending measured intervals as required by all pending claims.” App. Br. 11. The Examiner found that “Florio discloses a means for monitoring the delay interval between heart chambers at column 4, lines 45-49.” Ans. 3. Appellants do not contest this finding. The Examiner found that Florio discloses means for storing these intervals in memory at column 5, lines 35-47 and column 16, lines 13-26. Ans. 3; see Florio, col. 5, ll. 38-44 (disclosing that the interval between the evoked response of the captured ventricle and the conducted ventricle as an estimation of the inter-ventricular conduction time “can be stored in Appeal 2010-004669 Application 11/312,874 4 memory”); col. 16, ll. 31-33 (disclosing “[i]nter-ventricular conduction time, measured as the time between two negative deflecting depolarizations 146, 148, is stored in memory 94”). We agree with and adopt this finding. Florio’s disclosure of storing these intervals in memory satisfies the “means for storing in a memory structure said at least one of . . . the LV to RV interval” limitation of claim 13. The Examiner found that Florio discloses the functionality of performing a temporal analysis as claimed at column 4, line 64 to column 5, line 2; column 5, lines 35-47; and column 16, lines 13-44. Ans. 3, 6. Appellants contest this finding. App. Br. 12; Reply Br. 2-5. Florio discloses a “means to monitor CHF [(congestive heart failure)] progression by monitoring inter-ventricular conduction time during bi- ventricular threshold testing and IEGM [(a.k.a. a “pacemaker”)] template acquisition,” storing the inter-ventricular conduction time in memory, and using inter-ventricular conduction time “as a closed-loop feedback parameter within the stimulation device 10 to cause automatic adjustments of pacing parameters such that pacing therapy is optimized as CHF worsens or improves.” Col. 16, ll. 28-39. Florio also discloses that “[i]f, during single-chamber capture, the time elapsed between a negative peak detection 146 . . . to the next negative peak detection 148 increases, [that] indicates that interventricular conduction delay has worsened such that hemodynamic performance has been deteriorating.” Col. 16, ll. 21-26. Thus, Florio describes means for monitoring inter-ventricular conduction time intervals, means for storing these time intervals in memory, and means for performing a temporal trend analysis upon the stored intervals over a discrete period of time utilizing a plurality of the stored intervals (i.e., comparing the intervals over time to determine if they are increasing or Appeal 2010-004669 Application 11/312,874 5 decreasing, for use as a closed-loop feedback parameter within the stimulation device to cause automatic adjustments of pacing parameters to optimize pacing therapy as CHF worsens or improves). The Examiner found that Florio discloses performing an automated ventricular chamber capture as required in claim 13, at column 15, lines 5- 15. Ans. 3. Appellants do not contest this finding. In light of the above, the Examiner did not err in finding that Florio anticipates the subject matter of claim 13. We sustain the rejection of claim 13 and of claims 16 and 17, which fall with claim 13, as anticipated by Florio. Appellants rely on the same unpersuasive arguments asserted for claim 13 in contesting the rejection of claim 15 under 35 U.S.C. § 102(b) as anticipated by or, in the alternative, under 35 U.S.C. § 103(a) as unpatentable over Florio. We sustain the rejection of claim 15. Appellants point out limitations recited in each of claims 18-20, but present the same argument for all three claims. Accordingly, we select claim 19 as representative, with claims 18 and 20 standing or falling with claim 18. See 37 C.F.R. § 41.37(c)(1)(vii) (stating that “[a] statement which merely points out what a claim recites will not be considered an argument for separate patentability of the claim.”). In contesting the rejection of claims 18-20 as anticipated by Florio, Appellants concede that Florio discloses means for indicating loss of capture in a single ventricle, but argue that “there is no disclosed mechanism [in Florio] that identifies which ventricle has lost capture.” App. Br. 13. As correctly pointed out by the Examiner (Ans. 7), claims 18-20 do not recite identifying which ventricle has lost capture. Appeal 2010-004669 Application 11/312,874 6 The Examiner found that Florio discloses declaring loss of capture at column 15, lines 27-35. Ans. 4. The Examiner also pointed out that Florio discloses determining that both the left ventricle and the right ventricle have lost capture at step 195 of figure 7, and found that this step corresponds to declaring loss of capture in each of the chambers. Ans. 7. Florio also discloses independent pulse generators and independent sensing circuits and amplifiers. Col. 8, ll. 22-27, 47-54; Ans. 4. In column 15, lines 27-35 relied on by the Examiner, Florio discloses that if a negative peak is detected within a short period of time, such as 16 to 40 msec, after the stimulation pulse, the depolarization is interpreted as an evoked response, but if the depolarization occurs at a later time but still within the sampling window, it is interpreted as an intrinsic response indicating failure to capture either ventricle, and the system flags the failure in memory. Florio also discloses that, in the event that a non-capture in both chambers is flagged (i.e., step 195), the loss of capture counter is increased, the stimulation energy is increased, and the pacing routine continues. Col. 14, ll. 60-65. Florio describes the capture failure determination process in more detail at column 13, lines 10-52. See figs. 5A-5C (showing failure to capture both ventricles, single-chamber capture, and successful capture, respectively). In the particular scenario depicted in figure 5A (failure to capture both ventricles), Florio describes detection of a stimulation pulse 134 and, after the time delay window 136, detection of a subsequent depolarization complex 138 (e.g., an intrinsic R-wave) that represents the intrinsic responses of the left and right ventricles. Col. 13, ll. 16-22. For the above reasons, Florio provides a sound basis for the Examiner’s finding that Florio describes means for declaring loss of capture in each of the left ventricle and right ventricle that recorded the latent Appeal 2010-004669 Application 11/312,874 7 (intrinsic) ventricular event after the time delay window (i.e., 40 msec) but within the sampling period. Thus, Appellants’ argument does not apprise us of error in the Examiner’s rejection of claim 19. We therefore sustain the rejection of claim 19 and of claims 18 and 20, which fall with claim 19, as anticipated by Florio. DECISION The Examiner’s decision rejecting claims 13 and 15-20 is affirmed. No time period for taking any subsequent action in connection with this appeal may be extended under 37 C.F.R. § 1.136(a). See 37 C.F.R. § 1.136(a)(1)(iv). AFFIRMED mls Copy with citationCopy as parenthetical citation